HJ 700-2014Water quality. Determination of 65 elements. Inductively coupled plasma-mass spectrometry (English PDF)
水质 65种元素的测定 电感耦合等离子体质谱法
This is a limited preview
Buy now to download the full PDF (17 pages)
Need quotation for your management?
Get a formal quotation for HJ 700-2014 in 30 minutes.
Issued by
Ministry of Ecology and Environment
Level / Type
National · Mandatory
Issue date
May 16, 2014
Implementation date
July 1, 2014
Scope
HJ 700-2014 is the English-translated version of 水质 65种元素的测定 电感耦合等离子体质谱法.
HJ 700-2014 — This standard specifies inductively coupled plasma mass spectrometry for the determination of 65 elements in water.
This standard applies to surface water, groundwater, domestic sewage, low concentration industrial wastewater, silver, aluminum, arsenic, gold, boron, antimony, , , calcium, cadmium, strontium, cobalt, chromium, bismuth, copper, bismuth, antimony, bismuth, iron, gallium, antimony, bismuth, antimony, bismuth, indium, antimony, Potassium, strontium, lithium, strontium, magnesium, manganese, molybdenum, sodium, strontium,, nickel, phosphorus, lead, palladium, iridium, platinum, rhodium, iridium, osmium, iridium, Determination of antimony, bismuth, selenium, tellurium, tin, antimony, bismuth, antimony, bismuth, titanium, antimony, bismuth, uranium, vanadium, tungsten, antimony, bismuth, zinc and zirconium.
The detection limit of each element of the method is 0.02 µg/L to 19.6 µg/L, and the lower limit of determination is 0.08 µg/L to 78.4 µg/L. each See Appendix A for the method detection limits for the elements.
Found what you need?
Document preview — HJ 700-2014
National Standard of the People's Republic of China
- Classification
- Z 16
Issued by: Ministry of Ecology and Environment
Contents
- Foreword
- 1 Scope of application
- 2 Normative references
- 3 Terms and definitions
- 3.1 Soluble elements
- 3.2 Total amount of elements
- 4 Principle of the method
- 5 Interference and elimination
- 5.1 Mass Spectral Interference
- 5.2 Non-mass speculative interference
- 6 Reagents and materials
- 6.8 Standard solution
- 6.10 Internal standard use solution
- 7 Instruments and equipment
- 8 Samples
- 8.1 Sample collection
- 8.2 Preservation of samples
- 9.1 Instrument Commissioning
- 9.2 Drawing of the calibration curve
- 9.4 Determination
- 10 Calculation and representation of results
- 10.1 Calculation of results
- 10.2 Results are expressed
- 11 Precision and accuracy
- 11.1 Precision
- 11.2 Accuracy
- 12 Quality Assurance and Quality Control
Foreword
To implement the Environmental Protection Law of the People's Republic of China and the Law of the People's Republic of China on Water Pollution Prevention and Control, to protect the environment, this standard is formulated to ensure human health and standardize the monitoring methods for 65 elements in water.
This standard specifies inductively coupled plasma mass spectrometry for the determination of 65 elements in water.
This standard is the first release.
Appendix A of this standard is a normative appendix, and Appendix B to Appendix C are informative appendices.
This standard was formulated by the Science and Technology Standards Department of the Ministry of Environmental Protection.
This standard is mainly drafted by. Jiangsu Environmental Monitoring Center and Suzhou Environmental Monitoring Center Station.
The units that participate in the verification of this standard are. Nanjing Environmental Monitoring Center Station, Wuxi Environmental Monitoring Center Station, Shaoxing City Ring Environmental Monitoring Center Station, Ningbo Environmental Monitoring Center, Suzhou Water Supply Company Water Quality Testing Center, Maanshan Environmental Monitoring Heart station.
This standard was approved by the Ministry of Environmental Protection on May 16, 2014.
This standard has been implemented since July 1, 2014.
This standard is explained by the Ministry of Environmental Protection.
Water quality - Determination of 65 elements - Inductively coupled plasma mass spectrometry WARNING - Avoid direct contact with the skin when preparing and measuring standard solutions for highly toxic carcinogens such as barium, arsenic, and cadmium. Hydrochloric acid, nitric acid They are all highly chemically corrosive and irritating. They should be worn with protective equipment according to the requirements and operated in a fume hood to avoid acid mist absorption.
Into the respiratory tract and contact with skin and clothing.
1 Scope of application
This standard specifies inductively coupled plasma mass spectrometry for the determination of 65 elements in water.
This standard applies to surface water, groundwater, domestic sewage, low concentration industrial wastewater, silver, aluminum, arsenic, gold, boron, antimony, , , calcium, cadmium, strontium, cobalt, chromium, bismuth, copper, bismuth, antimony, bismuth, iron, gallium, antimony, bismuth, antimony, bismuth, indium, antimony, Potassium, strontium, lithium, strontium, magnesium, manganese, molybdenum, sodium, strontium,, nickel, phosphorus, lead, palladium, iridium, platinum, rhodium, iridium, osmium, iridium, Determination of antimony, bismuth, selenium, tellurium, tin, antimony, bismuth, antimony, bismuth, titanium, antimony, bismuth, uranium, vanadium, tungsten, antimony, bismuth, zinc and zirconium.
The detection limit of each element of the method is 0.02 µg/L to 19.6 µg/L, and the lower limit of determination is 0.08 µg/L to 78.4 µg/L. each See Appendix A for the method detection limits for the elements.
2 Normative references
The contents of this standard refer to the following documents or their terms. For undated references, the valid version is appropriate.
Used in this standard.
GB/T 6682 Analytical laboratory water specifications and test methods HJ/T 91 Surface Water and Wastewater Monitoring Technical Specifications HJ/T 164 Technical Specifications for Groundwater Environmental Monitoring HJ 493 Technical Regulations for Preservation and Management of Water Quality Samples HJ 677 Water quality metal total digestion nitric acid digestion method HJ 678 Water quality metal total digestion microwave digestion method
3 Terms and definitions
The following terms and definitions apply to this standard.
3.1 Soluble elements
Refers to the element content measured after filtration through a 0.45 µm filter on a sample that has not been acidified.
3.2 Total amount of elements
Refers to the unmeasured sample, the element content measured after digestion.
4 Principle of the method
After pretreatment, the water sample is detected by inductively coupled plasma mass spectrometry according to the mass spectrum or characteristic ion of the element.
Qualitative, internal standard method quantitative. The sample is carried into the atomization system by the carrier gas for atomization, and then enters the plasma as an aerosol.
The axial channel is fully vaporized, dissociated, atomized and ionized in high temperature and inert gas, and converted into charged positive the ions enter the mass spectrometer through the ion collection system, and the mass spectrometer separates and determines according to the mass-to-charge ratio of the ions, that is, the mass of the elements.
Sexual and quantitative analysis. Within a certain concentration range, the signal response value corresponding to the element mass number is proportional to its concentration.
5 Interference and elimination
5.1 Mass Spectral Interference
Mass spectrometry interference mainly includes polyatomic ion interference, isobaric interference, oxide and double charge interference. Multiple original Sub-ion interference is the most important source of interference for ICP-MS, and can be used for interference correction equations, instrument optimization, and collision reactions.
Pool technology to solve, common polyatomic ion interference see Appendix Table B.1. Isobaric interference can use interference correction the equation is corrected, or the sample is chemically separated before analysis, and the main interference correction equation is attached.
Record B.2. Oxide interference and double charge interference can reduce the effects by adjusting instrument parameters.
5.2 Non-mass speculative interference
Non-mass spectroscopic interference mainly includes matrix suppression interference, space charge effect interference, physical effect interference and the like. Non-mass type the degree of interference is related to the nature of the sample matrix and can be eliminated by internal standard method, instrument condition optimization or standard addition method.
6 Reagents and materials
Reagents used in this standard are analyzed using high-grade purification reagents that meet national standards, unless otherwise stated.
6.1 Experimental water. resistivity >=18Mohm·cm, and other indicators meet the first-class standard in GB/T 6682.
6.2 Nitric acid. rho(HNO3)=1.42 g/mL, superior grade or superior grade pure, if necessary, purified.
6.3 Hydrochloric acid. rho(HCl)=1.19 g/mL, superior grade or superior grade pure, if necessary, purified.
6.4 Nitric acid solution. 1 99.
6.5 Nitric acid solution. 2 98.
6.6 Nitric acid solution. 1 1.
6.7 Hydrochloric acid solution. 1 1.
6.8 Standard solution
6.8.1 Single element standard stock solution. rho = 1.00 mg/mL.
Available in spectrally pure metals (purity greater than 99.99%) or other standard materials to a concentration of 1.00 mg/mL Stock solution, select the appropriate medium according to the nature of each element (see Appendix Table B.3 Mixed Standard Stock Solution Group Recommendation Save media). A certified standard solution can also be purchased.
6.8.2 Mixed standard stock solution
Can purchase certified mixed standard solution, depending on the interference between elements, the nature of the standard solution and the test the content of the elements, the elements are grouped into a mixed standard stock solution (see Appendix B.3 recommended mixed standard stock solution) Liquid grouping and storage medium).
Note 1. Standard stock solutions for all elements should be stored in sealed polyethylene or polypropylene bottles after preparation.
Note 2. The solution containing the element Ag needs to be stored in the dark.
6.8.3 Mixed standard use solution
Can purchase certified mixed standard solution, depending on the interference between elements, the nature of the standard solution and the test the content of the element, the elemental standard stock solution (6.8.1 or 6.8.2) is diluted with a nitric acid solution (6.5), and the elements are grouped and formulated.
Into a mixed standard use solution, potassium, sodium, calcium, magnesium stock solution is the use of its solution, the concentration is 100 mg/L; the rest the element mixing solution concentration was 1 mg/L.
6.9 Internal standard storage solution. rho=100µg/L.
6Li, 45Sc, 74Ge, 89Y, 103Rh, 115In, 185Re, 209Bi should be selected as the internal standard elements (the selection of internal standard elements) refer to Appendix Table B.4) for details. The certified standard solution can be purchased directly and diluted to 100 µg/L with a nitric acid solution (6.4).
6.10 Internal standard use solution
The internal standard stock solution (6.9) was diluted with a nitric acid solution (6.4) to prepare an internal standard use solution. Due to different instruments Different internal diameter peristaltic pump tubes are added to the internal standard online, resulting in different concentrations of internal standard into the sample, so the internal concentration of the liquid is used.
The concentration of the internal standard element in the sample solution should be considered to be about 5 µg/L to 50 µg/L.
6.11 Mass spectrometer tuning solution. rho = 10 µg/L.
A tuning solution containing Li, Y, Be, Mg, Co, In, Tl, Pb, and Bi elements as a mass spectrometer should be used. Directly Purchase a certified standard solution and dilute to 10 µg/L with a nitric acid solution (6.4).
6.12 Argon. The purity is not less than 99.99%.
7 Instruments and equipment
7.1 Inductively coupled plasma mass spectrometer and its corresponding equipment. The working environment of the instrument and the requirements for the power supply are based on the instrument.
The instructions are prescribed. Instrument scanning range. 5-250 amu, resolution. 10% peak height corresponding to the peak width should be better than 1 amu.
7.2 Temperature control electric heating plate.
7.3 Microwave digestion instrument.
7.4 Filter device, 0.45 µm pore water microporous membrane.
7.5 Teflon beaker. 250 mL.
7.6 Polyethylene volumetric flask. 50mL, 100mL.
7.7 Polypropylene or Teflon bottle. 100mL.
7.8 Common instruments and equipment used in general laboratories.
8 Samples
8.1 Sample collection
Sample collection is performed in accordance with the relevant provisions of HJ/T 91 and HJ/T 164. Soluble element samples and total element samples are divided.
Do not collect.
8.2 Preservation of samples
Soluble element samples were collected immediately after collection with a 0.45 µm filter (7.4), and the initial filtrate was discarded.
The filtrate is cleaned of the sampling bottle, and the required volume of the filtrate is collected in a sampling bottle, and an appropriate amount of nitric acid (6.6) is added to adjust the acidity to pH9 Analysis steps
9.1 Instrument Commissioning
9.1.1 Reference operating conditions of the instrument
Different types of instruments have different optimal working conditions. Standard mode, collision/reaction cell mode, etc. should be used according to the instrument.
Instructions for operation.
9.1.2 Instrument Tuning
After igniting the plasma, the instrument needs to be preheated for 30 minutes. First use the mass spectrometer to tune the solution (6.11) to the instrument Sensitivity, oxide and double charge are tuned, under the condition that the sensitivity, oxide and double charge of the instrument meet the requirements.
The relative standard deviation of the signal intensity of the elements contained in the tuning solution is <= 5%. Then proceed within the mass range covering the elements to be tested Quality correction and resolution check, such as the difference between the quality correction result and the true value exceeds ±0.1 amu or the resolution of the tuned element signal the ratio of the peak width corresponding to the peak height of 10% exceeds 0.6-0.8 amu, and the mass spectrometer should be used according to the requirements of the instrument manual.
Make corrections.
9.2 Drawing of the calibration curve
A series of standard solutions of the elements to be tested are sequentially prepared, and the concentration range of the calibration curve can be adjusted according to the measurement needs. In capacity Take a certain volume of the standard use solution (6.8.3) in the bottle, and prepare a series of standard curve using the nitric acid solution (6.4).
As follows. aluminum, boron, antimony, cobalt, copper, iron, manganese, titanium, zinc concentrations of 0µg/L, 10.0µg/L, 50.0µg/L, 100 µg/L, 200 µg/L, 300 µg/L, 400 µg/L, 500 µg/L; silver, arsenic, gold, antimony, bismuth, cadmium, antimony, chromium, antimony, bismuth, antimony, Bismuth, gallium, germanium, antimony, bismuth, antimony, indium, antimony, bismuth, antimony, molybdenum, antimony, bismuth, nickel, phosphorus, lead, palladium, rhodium, platinum, rhodium, 浓度, , , , , selenium, bismuth, tin, antimony, bismuth, antimony, bismuth, antimony, uranium, vanadium, tungsten, antimony, bismuth, zirconium 0µg/L, 0.5µg/L, 1.0µg/L, 5.0µg/L, 10.0µg/L, 20.0 µg/L, 40.0 µg/L, 50.0µg/L; potassium, sodium, Calcium and magnesium concentrations were 0 mg/L, 5.0 mg/L, 10.0 mg/L, 20.0 mg/L, 40.0 mg/L, 60.0 mg/L, 80.0 mg/L, 100mg/L; lithium and strontium concentrations are 0 mg/L, 0.1 mg/L, 0.5 mg/L, 1.0 mg/L, 2.0 mg/L, 3.00 mg/L, 4.0 mg/L, Standard series of 5.0mg/L. The internal standard element standard solution (6.10) can be directly added to the working solution or in the sample.
It is automatically added by a peristaltic pump before atomization. The concentration of the internal standard should be much higher than the concentration of the internal standard element contained in the sample itself.
The internal standard elements and concentration range are given in 6.10.
The standard solution is determined by ICP-MS, and the ratio of the sample signal to the internal standard signal is taken as the abscissa of the standard solution concentration.
A calibration curve is established on the ordinate. The slope was determined by linear regression analysis method for sample content calculation.
9.4 Determination
9.4.1 Determination of samples
Before each sample is measured, flush the system with nitric acid solution (6.5) until the signal is at a minimum, and the signal to be analyzed is stable.
The measurement can only be started afterwards. The same amount of internal standard element standard use solution should be added to the sample when measuring the calibration curve.
(6.10). If the concentration of the element to be tested in the sample exceeds the calibration curve, it should be diluted with nitric acid solution (6.4) and then re-measured.
The dilution factor is f. The sample solution matrix is complex, and when the polyatomic ion interference is serious, the calibration method listed in Appendix B.2 can be used.
The calibration is performed, and can also be corrected by the collision/reaction cell mode technique according to the conditions recommended by each instrument manufacturer.
9.4.2 Determination of laboratory blank samples
Laboratory blank samples were determined according to the same measurement conditions as the samples.
10 Calculation and representation of results
10.1 Calculation of results
The element content in the sample is calculated according to formula (1).
( )1 2 frho rho rho= - × (1) where. rho -- the concentration of the element in the sample, µg/L or mg/L;
1rho - the mass concentration of the elements in the sample after dilution, µg/L or mg/L;
2rho -- mass concentration of elements in laboratory blank samples after dilution, µg/L or mg/L;
f -- dilution factor.
10.2 Results are expressed
The number of decimal places of the measurement is consistent with the method detection limit, and up to three significant digits are reserved.
11 Precision and accuracy
11.1 Precision
Six laboratories performed standard solutions with potassium, sodium, calcium and magnesium concentrations of 5.0 mg/L, 25.0 mg/L and 50.0 mg/L.
Direct measurement.
The relative standard deviations in the experimental room were 2.90% to 12.8%, 1.27% to 7.93%, and 1.46% to 6.85%, respectively.
The relative standard deviations between laboratories were. 2.83% to 7.71%, 3.99% to 5.70%, and 1.41 to 3.53%.
The repeatability limits ranged from 0.99 mg/L to 1.26 mg/L, from 1.78 mg/L to 3.42 mg/L, and from 3.56 mg/L to 5.67 mg/L.
The reproducibility limits range from 1.01 mg/L to 1.53 mg/L, from 3.06 mg/L to 4.60 mg/L, and from 4.85 mg/L to 7.58 mg/L.
6 laboratories for the remaining 61 elements with concentrations of 0.10~20.0µg/L, 20.0~50.0µg/L and 50.0~100µg/L the quasi-solution was directly measured.
The relative standard deviations in the experimental room were. 2.01% to 14.1%, 1.27% to 10.1%, and 1.15% to 7.84%, respectively.
The relative standard deviations between laboratories were 0.86% to 10.4%, 1.00% to 9.81%, and 0.85 to 4.80%, respectively.
The repeatability range is from 0.02 µg/L to 3.31 µg/L, from 0.08 µg/L to 7.38 µg/L, and from 3.47 µg/L to 11.9 µg/L.
The reproducibility limit ranges from 0.02 µg/L to 4.37 µg/L, from 0.08 µg/L to 8.24 µg/L, and from 3.82 µg/L to 16.9 µg/L.
11.2 Accuracy
Six laboratories carried out the spiked recovery test of soluble elements on surface water samples, and added back to the soluble elements of surface water.
The yield ranged from 86.8% to 102%, and the final recovery of the spiked standard was 86.8%±8.6%~102%±7.7%.
6 laboratories carried out the spiked recovery test of soluble elements in wastewater samples, and the recovery rate of soluble elements in wastewater was increased.
The circumference is. 85.7%~109%, and the final recovery rate is 85.7%±7.6%~109%±9.9%.
6 laboratories carried out spiked recovery experiments on surface water samples, and spiked recovery of total surface water elements the range is 91.3%~110%, and the final recovery of spiked standard is 91.3%±14.4%~110%±13.2%.
6 laboratories carried out spiked recovery experiments on total amount of waste water samples, and spiked recovery range of total waste water elements for. 92.1% to 108%, the final recovery of the spiked standard is 92.0% ± 11.4% ~ 108% ± 3.6%.
See Appendix C for statistics on precision and accuracy results.
12 Quality Assurance and Quality Control
12.1 Standard curve. A calibration curve should be drawn for each sample analyzed. Normally, the correlation coefficient of the calibration curve should be up to More than 0.999.
12.2 Internal standard. The intensity of the internal standard must be monitored in each analysis. The response of the internal standard in the sample should be in the response of the calibration curve.
70% ~ 130% of the value, otherwise it indicates that the instrument drifts or has interference, should be re-analyzed after finding the cause. If the matrix interference is now determined after dilution; if the sample contains internal standard elements, the internal standard needs to be replaced or the internal standard needs to be replaced.
Standard element concentration.
12.3 Blank. At least one full program blank and laboratory blank should be made for each batch of samples. Blank values should meet the following conditions One can be considered acceptable. (1) the blank value should be lower than the method detection limit; (2) 10% below the standard limit; (3) it is less than 10% of the lowest measured value of each batch of samples. Otherwise, you must find the reason, re-analyze until you pass the test.
Product.
12.4 Laboratory Control Samples. In each batch of samples, each analyte should be added to the reagent blank with spiked recovery.
Should be between 80% and 120%; can also use certi...
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 17 pages — is available in the English PDF.
Referenced standards
Normative references
Cited by
- GB 4287-2026Discharge standard of water pollutants for textile industry
- GB 47946-2026Discharge standard of main water pollutants (total antimony, total molybdenum and others) in the Danjiangkou reservoir area and its upper reaches (on trial)
- GB 16889-2024Standard for pollution control on the landfill site of municipal solid waste
- GB 18598-2019Standard for pollution control on the hazardous waste landfill
How to Download HJ 700-2014
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 6682-2008 — Water for analytical laboratory use - Specification and test methods
HJ 1409-2025 — Technical guidelines for environmental impact assessment - Marine ecology and environment
HJ 1410.1-2025 — Basic dataset of derivation of water quality criteria for aquatic organisms - Part 1: General principles
Secure payment via Stripe
Payments accepted
HJ 700-2014
$550.00