GB/T 5201-2012Test procedures for semiconductor charged particle detectors (English PDF)
带电粒子半导体探测器测量方法
Open the GB/T 5201-2012 preview as PDF
This is a limited preview
Buy now to download the full PDF (14 pages)
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
June 29, 2012
Implementation date
November 1, 2012
Scope
GB/T 5201-2012 is the English-translated version of 带电粒子半导体探测器测量方法.
GB/T 5201-2012 lays down how the electrical characteristics and the nuclear radiation performance of semiconductor charged particle detectors are measured, together with test methods for certain special environments. It is written for detectors with a partially depleted layer, and fully depleted devices may be measured by reference to it. The standard replaces GB/T 5201-1994 and was prepared with reference to IEC 60333:1993, from which it differs, the correspondence being declared non-equivalent. Terms and definitions are taken over wholly from GB/T 4960.6-2008 and the environmental tests clause refers wholly to GB/T 10263-2006. A general clause fixes the reference and standard test conditions, requires complete darkness, forbids exceeding the permitted bias and particle fluence rate, and limits the voltage drop that the leakage current produces across the bias resistor. The clause on electrical characteristics covers the forward and reverse voltage-current characteristic, the capacitance-voltage characteristic measured with an ac bridge, noise measured by the pulse height method with a quasi-Gaussian shaping amplifier and a precision pulse generator, and timing performance: electrical rise time, charge collection time, and time resolution measured with a laser diode pulser and a time-to-amplitude converter.
Document preview — GB/T 5201-2012
National Standard of the People's Republic of China
- ICS
- 27.120
- Classification
- F 88
- Replacing
- GB/T 5201-1994
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 General requirements
- 5 Electrical characteristics
- 5.1 Voltage-current characteristic (V-I characteristic)
- 5.2 Capacitance-voltage characteristic
- 5.3 Noise measurement
- 5.3.1 Measurement method and measuring system
- 5.3.2 Measurement requirements
- 5.3.3 Calibration of the input signal
- 5.3.4 Measurement of amplifier noise and of its variation with the external capacitance
- 5.3.5 Detector noise
- 5.3.6 Variation of detector noise with bias voltage
- 5.3.7 Variation of detector noise with the amplifier time constant
- 5.4 Timing performance
- 5.4.1 Electrical rise time of the detector
- 5.4.2 Charge collection time
- 5.4.3 Time resolution
1 Scope
The standard lays down methods for measuring the electrical characteristics and the nuclear radiation performance of semiconductor charged particle detectors, together with test methods for certain special environments.
It applies to semiconductor detectors for charged particles with a partially depleted layer.
The measurement of fully depleted semiconductor detectors may be carried out by reference to this standard.
2 Normative references
Three documents are cited: GB/T 4960.6-2008 Terminology of nuclear science and technology, Part 6: Nuclear instruments; GB/T 10263-2006 Environmental conditions and test methods for nuclear radiation detectors; and GB/T 13178-2008 Gold-silicon surface barrier detectors.
For dated references only the edition cited applies; for undated references the latest edition, including all amendments, applies.
3 Terms and definitions
The terms and definitions given in GB/T 4960.6-2008 apply to this document; the standard defines no terms of its own.
4 General requirements
4.1 Measurement shall be carried out under the reference conditions or the standard test conditions given in Table 1. That table has three columns, for the item, the reference condition and the standard test condition, and covers thirteen items: ambient temperature, relative humidity, atmospheric pressure, ac supply voltage, ac supply frequency, ac supply waveform, dc supply voltage, ambient gamma radiation as air absorbed dose rate, external magnetic field interference, external magnetic induction and radioactive contamination. The reference ambient temperature is 20 °C and the standard test range 18 °C to 22 °C; the reference relative humidity is 65 % and the standard range 50 % to 75 %; the reference atmospheric pressure is 101.3 kPa and the standard range 86 kPa to 106 kPa. Two footnotes state that the single phase supply is 220 V or the three phase supply 380 V, that when a battery supply is used its voltage variation is plus or minus 1 % of the rated value and ripple is not taken into account, and that in special cases the ac supply frequency is dealt with as the product standard provides.
4.2 Every measurement of the performance of a detector shall be carried out in complete darkness.
4.3 During measurement the limiting conditions of the detector, such as the permitted working bias and the maximum particle fluence rate, shall not be exceeded, and the technical specification of the instruments used shall have no marked influence on the result of the measurement of the detector parameters.
4.4 The reproducibility of the performance parameters measured shall lie within the measurement uncertainty.
4.5 The voltage drop that the leakage current of the detector produces across the bias resistor shall be less than 15 % of the bias applied to the detector.
4.6 For detectors working under low vacuum, the relevant performance may be measured at 0.1 Pa to 5 Pa. For detectors working under high vacuum, the relevant performance shall be measured at a pressure below 0.0001 Pa.
5.1 Voltage-current characteristic (V-I characteristic)
When the reverse V-I characteristic is measured, the circuit shall be connected as in Figure 1; for each value of reverse bias the corresponding reverse current shall be recorded, and the ambient temperature at the time of measurement shall be recorded as well.
When the forward V-I characteristic is measured, the circuit shall be connected as in Figure 2; the result is expressed as the forward voltage drop at a fixed forward current, for instance 5 mA.
Both block diagrams show the detector with an ammeter in series and a voltmeter in parallel, a resistor R, and a bias supply.
5.2 Capacitance-voltage characteristic
The block diagram of the measurement is Figure 3. The working frequency of the bridge shall lie between 1000 Hz and 100 000 Hz. The peak to peak value of the ac signal that the bridge applies to the detector shall be less than one tenth of the detector bias. The value of the dc blocking capacitor, in picofarads, shall be at least ten times the maximum junction capacitance of the detector under test, and its voltage rating shall be twice the maximum bias applied to that detector. The inductor may also be replaced by a high value resistor; the resistance shall be chosen so that the voltage drop across it is negligible compared with the detector bias and so that the output capacitance of the supply does not affect the result.
The steps are: measure the capacitance of the system at a given bias with the detector disconnected; connect the detector and measure, at the same bias, the sum of the detector capacitance and the system capacitance, the detector capacitance in picofarads being the difference of the two, as given by equation 1; then change the bias and repeat, so obtaining the variation of the detector capacitance with bias.
5.3 Noise measurement
5.3.1 Detector noise is measured by the pulse height method, the measuring system being connected as in Figure 4: bias supply, load resistor R, detector, charge sensitive amplifier, amplifier and multichannel pulse height analyser, with a test capacitor and a matching impedance Z fed from a precision pulse generator. If the multichannel pulse height analyser used is of the digital type, it may be connected directly to the charge sensitive amplifier and no amplifier is needed.
5.3.2.1 The amplifier shall have a quasi-Gaussian shaping function with adjustable shaping time. The shaped pulse shape is that of Figure 5, settled by the performance of the shaper and by the time constant. Figure 5 gives the parameters that fix the shape, the rise time to the peak, the tail time and the width at half the peak amplitude; noise shall be measured with the half amplitude width at 1 microsecond.
5.3.2.2 The pulse signal delivered by the precision pulse amplitude generator shall be an exponentially decaying signal whose rise time is not greater than 5 % of the shaping time constant of the amplifier. The decay time of the pulse signal shall be such that, over an interval equal to the shaping time constant of the amplifier, the pulse amplitude falls by less than 2 %. The integral non-linearity of the generator shall be less than 0.1 %.
5.3.2.3 The test coupling capacitor shall in general be less than 10 pF.
5.3.2.4 The gain of the measuring system and the signal amplitude shall work in the linear range, and the width at half height of the peak measured on the multichannel pulse height analyser shall be greater than 12 channels.
5.3.3 For the calibration of the input signal, the detector is given its normal working bias and irradiated with a charged particle source of known energy, an alpha source requiring vacuum conditions, so that a corresponding spectrum is obtained on the multichannel pulse height analyser. The source is then removed, the precision amplitude pulse generator is switched on and its output is adjusted until its peak position on the multichannel analyser coincides with the peak position of the source.
5.3.4 To measure the amplifier noise, the detector is taken off and the input of the preamplifier is closed with a screening cap; the spectral lines corresponding to the generator signals of energy E1 and E2 are recorded separately. With the generator signal set to E1, when the count in the peak channel of the spectrum is greater than 4000, the points on the spectral line whose count is more than half the peak count are taken and the peak channel is calculated by the weighted mean of equation 2. The amplifier noise expressed as the width at half height is settled by interpolation. If a computer curve fitting method is used, a lower count may be used provided the deviation of the width at half height is within plus or minus 5 % at 90 % confidence. The generator signal is then changed to E2 and the peak position determined in the same way. The amplifier noise with zero external capacitance is calculated by equation 3, and the difference between the chosen E1 and E2 shall be greater than three times the width at half height of the spectral line. Capacitors of different values are then connected at the input of the preamplifier and the measurement repeated, so that the curve of amplifier noise against external capacitance is obtained and from it the slope of the preamplifier noise against capacitance.
5.3.5 To measure detector noise, the detector is given the intended bias and the spectral lines corresponding to the generator signals of energy E1 and E2 are obtained, as shown in Figure 6. Again with a count in the peak channel greater than 4000, the peak positions and the width at half height are recorded, and the total noise of the detector and amplifier system is calculated by equation 4. If the pulse spectral line is Gaussian or quasi-Gaussian, the detector noise may be obtained as follows: the detector capacitance at the intended bias is read from the capacitance-voltage curve measured in 5.2, and the amplifier noise corresponding to that capacitance is read from the curve of amplifier noise against external capacitance measured in 5.3.4; this noise value represents the electrical noise other than that of the detector. The detector noise is then calculated by equations 5 and 6, which subtract the two contributions in quadrature. For the result to be valid, the total noise and the electrical noise other than the detector, expressed either in kiloelectronvolts or in channels, shall differ by at least 20 %.
5.3.6 To obtain the variation of detector noise with bias, the working bias of the detector is changed and the measurement steps of 5.3.5 are repeated, so giving the detector noise at different bias values; when required, the curve of detector noise against bias may be drawn.
5.3.7 To obtain the variation of detector noise with the amplifier time constant, the measuring system of Figure 4 is used and the amplifier shaping time is varied over a wide range. Since the amplifier noise differs with the shaping time, and the way that noise varies with the external capacitance differs as well, the corresponding amplifier noise shall be subtracted from the measured noise whenever the shaping time is changed. At the intended detector working bias the shaping time of the amplifier is varied and the steps of 5.3.4 and 5.3.5 are repeated, giving the curve of detector noise against amplifier shaping time at that bias; to obtain that curve at other bias values, only the detector working bias need be changed and the steps repeated.
5.4 Timing performance
5.4.1 The electrical rise time of the detector is measured with the system of Figure 7: bias supply, load resistor, detector, fast preamplifier, fast amplifier and oscilloscope, with a step pulse generator feeding the input through a capacitor and a matching impedance. Capacitors of different values are connected at the input of the fast preamplifier and a step voltage of amplitude less than one tenth of the bias is applied; the oscilloscope gives the curve of the rise time of the fast preamplifier against the external capacitance. The detector is then connected as in Figure 7, given the intended bias, and the same step voltage applied; the oscilloscope gives the rise time of the output waveform. From the curve of preamplifier rise time against external capacitance, the preamplifier rise time corresponding to the detector capacitance is found, and the electrical rise time of the detector is calculated by equation 7, which subtracts the two in quadrature. For the result to be valid, the total system rise time and the preamplifier rise time shall differ by at least 20 %. Changing the bias and repeating the steps gives the electrical rise time of the detector at different bias values.
5.4.2 For detectors in which the observed rise time and the total system rise time differ by more than 20 %, the charge collection time may be measured as follows: the detector is irradiated with a mono-energetic particle source of short range and the rise time of the output signal observed on the oscilloscope is taken; the total system rise time is measured as in 5.4.1; the charge collection time is then calculated by equation 8, again a quadrature subtraction. A note states that the observed rise time and the total system rise time should differ by at least 20 %: if they differ by more than 20 % the detector is regarded as slow, and if by less than 20 % as fast. For detectors in which the two differ by less than 20 %, a fast proton pulse accelerator or a pulsed light source may be used to create a large number of electron-hole pairs in the detector; a sampling oscilloscope then measures the current pulse they form directly, and the charge collection time is measured by its width. When the charge collection time is reported, the type and name of the radiation source used and the rise time of the amplifier shall be stated.
5.4.3 Time resolution is measured with the system of Figure 8. The light pulse produced by a laser diode pulser serves as the start pulse of a time-to-amplitude converter, and the output signal of the detector under test triggers the stop signal of the converter, so that a time distribution spectrum is obtained on the multichannel pulse height analyser. The constant fraction of the constant fraction discriminator is set to 20 % and the delay to 1.1 times the rise time from 10 % to 90 %. The shaping time is adjusted for best timing performance; usually no integration is used, and the best performance is obtained when the differentiation time constant is slightly greater than the rise time of the output signal of the preamplifier. The system contains a shaping amplifier that starts the multichannel pulse height analyser and serves for amplitude selection. Time resolution shall be measured at three or more energies between 1 MeV and 100 MeV, the intervals between the energies being approximately equal. A calibrated precision pulse generator is used and the laser diode pulser adjusted so that the amplitude of its output signal corresponds to the pulse amplitude chosen on the generator; during the timing measurement the precision pulse generator is switched off. With this method the time distribution peak is acquired with a width at half height of at least 12 channels and a count in the peak channel of at least 4000; if a computer curve fitting method is used, the deviation of the width at half height shall be within plus or minus 5 % at 90 % confidence. The time axis shall be calibrated with at least two known delay values before the width at half height and the width at one tenth height, both in nanoseconds, can be measured, the detector's own time resolution being found from the measured system time resolution by quadrature subtraction. Each measurement shall state the spectral characteristics of the pulsed light, the detector bias, the constant fraction and the delay of the constant fraction discriminator, and the differentiation time and rise time of the fast shaping amplifier.
Relationship to previous standards Relationship to previous standards
The cover carries the replacement mark for GB/T 5201-1994. The foreword states that the present standard replaces GB/T 5201-1994, whose Chinese title ended in the words for test methods, and lists the main technical changes: a foreword was added; a clause 2 Normative references was added and the following clauses renumbered; the former clause on terms and symbols became clause 3 Terms and definitions, which now cites GB/T 4960.6-2008 in full instead of restating the terms; the symbols subclause of the former standard was deleted, symbols being explained where they are used; a subclause 4.1 on the reference conditions or standard test conditions was added in place of the former 3.1; the former 3.3 and 3.4 were merged into 4.3 and the former 3.7 deleted; a reverse V-I characteristic test was added to 5.1; the introductory paragraph before the former noise measurement subclause became 5.3.1 Measurement method and measuring system and the following subclauses were renumbered; the former subclause on the variation of detector noise with the amplifier time constant was expanded and renumbered 5.3.7; a criterion for distinguishing fast and slow detectors was added to 5.4.2; and clause 7 Environmental tests now cites GB/T 10263-2006 in full instead of restating it.
The foreword also states that the standard was prepared by the redrafting method with reference to IEC 60333:1993 Nuclear instrumentation, Semiconductor detectors for charged particles, Test procedures, the degree of correspondence with IEC 60333:1993 being non-equivalent.
Note on the printed titles Note on the printed titles
The English title on the cover, kept unchanged above, is Test procedures for semiconductor charged particle detectors. The Chinese title of this edition ends in the words for measurement methods, whereas the Chinese title of the edition it replaces, GB/T 5201-1994, ended in the words for test methods. The printed English follows the wording of the earlier edition rather than the Chinese wording of this one. The English title has been reproduced exactly as printed, since the difference is one of wording and not a misprint.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 14 pages — is available in the English PDF.
Referenced standards
Editions of GB/T 5201
| Edition | Title | Revision | Status |
|---|---|---|---|
| GB/T 5201-2012 | Test procedures for semiconductor charged particle detectors | current edition | Current |
| GB/T 5201-1994 | Test procedures for semiconductor charged particle detectors | previous edition | In force until 2012-11-01 |
This page sells the current edition, GB/T 5201-2012. Earlier editions are listed for reference only.
How to Buy GB/T 5201-2012
- 1Add to cart. Click the "Buy GB/T 5201-2012" button on this page. You can add more standards before checkout.
- 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
- 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 47310-2026 — Determination of total silicon, aluminium, iron, potassium, sodium, calcium, magnesium, manganese, phosphorus, titanium and sulfur in soil - Monochromatic excitation energy dispersive X-ray fluorescence spectrometry
GB/T 47321-2026 — Specification for the warning data exchange of the national emergency early warning dissemination system
GB/T 47293-2026 — Determination of available mercury in soil
Secure payment via Stripe
Payments accepted
GB/T 5201-2012
$250.00