GB/T 4325.8-2013Methods for chemical analysis of molybdenum - Part 8: Determination of cobalt content - 5-Cl-PADAB spectrophotometry and flame atomic absorption spectrometry (English PDF)
钼化学分析方法 第8部分:钴量的测定 钴试剂分光光度法和火焰原子吸收光谱法
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for GB/T 4325.8-2013 in 30 minutes.
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
May 9, 2013
Implementation date
February 1, 2014
Scope
GB/T 4325.8-2013 is the English-translated version of 钼化学分析方法 第8部分:钴量的测定 钴试剂分光光度法和火焰原子吸收光谱法.
GB/T 4325.8-2013 is Part 8 of GB/T 4325 and specifies the determination method of cobalt in molybdenum.
Clause 1 states that this part is suitable for the determination of cobalt in molybdenum powder, molybdenum bar, molybdenum trioxide and ammonium molybdate, with a measurement range of 0.0006 % to 0.010 % for method 1 and of 0.0010 % to 0.010 % for method 2.
Its six clauses are Scope; Normative references; General; Method 1 - Cobalt reagent spectrophotometry; Method 2 - Flame atomic absorption spectrometry; and Test report.
Clause 3 requires that only confirmed analytical reagents be used, that the water be distilled or deionized or of equivalent purity meeting GB/T 6682, that the instruments be within their verification period and meet the technical parameters required by the verification, and that the glass containers be the Class A products specified in GB/T 12808, GB/T 12809 and GB/T 12806 and be used as required by GB/T 12810.
In clause 4 the sample is decomposed with hydrogen peroxide or sodium hydroxide solution, the main molybdenum is complexed with potassium sodium tartrate, the cobalt reagent forms an orange-red complex with cobalt in a phosphate buffer solution at pH 6 to 7, and the absorbance is measured in a phosphoric acid medium at a wavelength of 550 nm in a 2 cm cuvette against a blank, the cobalt content being read from a working curve plotted from 0 mL to 1.00 mL of cobalt standard solution and reported as the mass fraction of cobalt in %.
Clause 5 covers the flame atomic absorption route and requires an atomic absorption spectrometer fitted with a cobalt hollow cathode lamp whose characteristic concentration for cobalt is not greater than 0.035 micrograms per millilitre, whose standard deviation over ten absorbance measurements does not exceed 1.0 % of the mean absorbance at the highest standard concentration and 0.5 % of that mean at the lowest, and whose working curve, divided into 5 equal concentration segments, has a ratio between the highest and the lowest absorbance difference of not less than 0.7; the Mo matrix used must contain less than 0.0001 % cobalt.
Clause 2 cites GB/T 6682, GB/T 12806, GB/T 12808, GB/T 12809 and GB/T 12810, and the page breaks off inside clause 5, so clause 6 Test report is not shown.
The front matter states that it replaces GB/T 4325.7-1984, while the page's version table instead lists GB/T 4325.8-1984, on the dimethylglyoxime photometric determination of nickel content, as obsolete; it was issued on May 9, 2013 and has applied since February 1, 2014.
Found what you need?
Document preview — GB/T 4325.8-2013
National Standard of the People's Republic of China
- ICS
- 77.120.99
- Classification
- H 63
- Replacing
- GB/T 4325.7-1984
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword3
- 1 Scope6
- 2 Normative references6
- 3 General6
- 4 Method 1 - Cobalt reagent spectrophotometry7
- 5 Method 2 - Flame atomic absorption spectrometry10
- 6 Test report14
1 Scope
This part of GB/T 4325 specifies the determination method of cobalt in molybdenum.
This part is suitable for the determination of cobalt in molybdenum powder, molybdenum bar, molybdenum trioxide and ammonium molybdate. Method 1's measurement range: 0.0006% ~ 0.010%; method 2's measurement range: 0.0010% ~ 0.010%.
2 Normative references
The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) are applicable to this standard.
GB/T 6682 Water for analytical laboratory use - Specification and test methods (ISO 3696)
GB/T 12806 Laboratory glassware - One-mark volumetric flasks (ISO 1042)
GB/T 12808 Laboratory glassware - One-mark volumetric flasks (ISO 648)
GB/T 12809 Laboratory glassware - Principles of design and construction of volumetric glassware (ISO 384)
GB/T 12810 Laboratory glassware - Volumetric glassware - Methods for use and testing of capacity (ISO 4787)
3 General
3.1 Unless otherwise specified, only confirmed analytical reagents are used in the analysis; the water used is distilled or deionized water or water of equivalent purity, which shall meet the requirements of GB/T 6682.
3.2 The instruments used shall be within the verification period; their performance shall meet the technical parameter indicators required by the verification; the glass container shall use the Class A products as specified in GB/T 12808, GB/T 12809, GB/T 12806; the specific method of use shall refer to the requirements of GB/T 12810.
4 Method 1 - Cobalt reagent spectrophotometry
4.1 Method summary
The sample is decomposed with hydrogen peroxide or sodium hydroxide solution; the main molybdenum is complexed with potassium and sodium tartrate. The cobalt reagent forms an orange-red complex with cobalt in a phosphate buffer solution at a pH of 6 ~ 7; the absorbance is measured in a phosphoric acid medium.
4.2 Reagents
4.2.1 Hydrogen peroxide (rho = 1.10 g/mL), excellent grade pure.
4.2.2 Nitric acid (1 + 1).
4.2.3 Sulfuric acid (1 + 1).
4.2.4 Phosphoric acid (rho = 1.70 g/mL), excellent grade pure.
4.2.5 Sodium hydroxide solution (200 g/L), excellent grade pure.
4.2.6 Potassium sodium tartrate solution (400 g/L), excellent grade pure.
4.2.7 4-[(5-Chloro-2-pyridine)azo]-1,3-diaminobenzene (cobalt reagent) ethanol solution (0.025%), stored in a brown bottle.
4.2.8 Buffer solution (pH 6 ~ 7): Weigh 35.5 g of disodium hydrogen phosphate and dissolve it in 500 mL of water; weigh 34 g of potassium dihydrogen phosphate and dissolve it in 500 mL of water; mix the two; prepare for use.
4.2.9 Phenolphthalein ethanol solution (1 g/L).
4.2.10 pH test paper (pH 1 ~ 14).
4.2.11 Cobalt standard storage solution: Weigh 0.1000 g of metallic cobalt (mass fraction >= 99.99%) in a 250 mL beaker; add 10 mL of nitric acid (4.2.2) to dissolve it; add 10 mL of sulfuric acid (4.2.3); heat until sulfuric acid white smoke appears; remove and cool it; use water to rinse the watch glass and the beaker wall; add 30 mL of water; boil it; cool it; use water to transfer it into a
4.5.4.1
4.5.4.1.2 Add 4 mL of potassium sodium tartrate solution (4.2.6), 1.5 mL of sulfuric acid (4.2.3); add 1 ~ 2 drops of phenolphthalein ethanol solution (4.2.9); use sodium hydroxide solution (4.2.5) to neutralize it until the solution is light purple; add sulfuric acid (4.2.3) to fade the purple (pH 6 ~ 7); use pH test paper to check it. The solution is transferred to a 50 mL colorimetric tube by a small amount of water; control the volume within 30 mL.
4.5.4.1.3 Add 1.5 mL of cobalt reagent ethanol solution (4.2.7); mix well; add 2.5 mL of buffer solution (4.2.8); mix well; put the colorimetric tube in a boiling water bath and heat for 5 min; remove and cool it to room temperature; accurately add 8 mL of phosphoric acid (4.2.4); use water to dilute to the mark; mix well.
4.5.4.1.4 Move part of the solution (4.5.4.1.3) into a 2 cm cuvette; along with the blank solution as a reference, measure its absorbance at a wavelength of 550 nm by a spectrophotometer; find out the corresponding content of cobalt from the working curve.
4.5.4.2 Molybdenum trioxide, ammonium molybdate
Place the sample (4.5.2) in a 150 mL beaker; add 5 mL of sodium hydroxide solution (4.2.5); heat to dissolve until the solution is clear; cool it down. The following shall be carried out according to 4.5.4.1.2 ~ 4.5.4.1.4.
4.5.5 Drawing of working curve
4.5.5.1 Pipette 0 mL, 0.10 mL, 0.20 mL, 0.50 mL, 0.80 mL, 1.00 mL of cobalt standard solution (4.2.12), respectively, into a set of 50 mL colorimetric tubes.
Add 4 mL of potassium sodium tartrate solution (4.2.6); add 1 ~ 2 drops of phenolphthalein ethanol solution (4.2.9); add sodium hydroxide solution (4.2.5) dropwise to neutralize the solution until the solution is light purple; then add sulfuric acid (4.2.3) to make the purple fade (pH 6 ~ 7); use pH test paper to check it. Add water to control the volume within 30 mL. The following shall proceed according to 4.5.4.1.3.
4.5.5.2 Move part of the solution (4.5.5.1) into a 2 cm cuvette; take the reagent blank as a reference; measure its absorbance at a wavelength of 550 nm by a spectrophotometer. Draw a working curve with the content of cobalt as the abscissa and absorbance as the ordinate.
4.6 Calculation of analysis results
The cobalt content is calculated as the mass fraction of cobalt wCo; the value is expressed in % and calculated according to formula (1): […]
5 Method 2 - Flame atomic absorption spectrometry
5.2 Reagents
5.2.1 Hydrogen peroxide (rho = 1.10 g/mL), excellent grade pure.
5.2.2 Hydrochloric acid (rho = 1.19 g/mL), excellent grade pure.
5.2.3 Nitric acid (1 + 1).
5.2.4 Citric acid solution (500 g/L), excellent grade pure.
5.2.5 Cobalt standard storage solution: Weigh 0.1000 g of metallic cobalt (>= 99.99%) in a 250 mL beaker; use a small amount of water to m; add 20 mL of nitric acid (5.2.3); cover a watch glass; heat it at a low temperature to dissolve it completely; remove and cool to room temperature; transfer to a 1000 mL volumetric flask; use water to dilute to the mark; mix it uniformly. This solution 1 mL contains 100 µg of cobalt.
5.2.6 Cobalt standard solution: Pipette 10.00 mL of cobalt standard stock solution (5.2.5) into a 100 mL volumetric flask; add 10 mL of nitric acid (5.2.3); use water to dilute to the mark; mix well. 1 mL of this solution contains 10 µg of cobalt.
5.2.7 Mo matrix: The mass fraction of cobalt contained is less than 0.0001%.
5.3 Apparatus
Atomic absorption spectrometer, which is equipped with cobalt hollow cathode lamp.
Under the best working conditions of the instrument, those which can achieve the following indicators can be used:
- Characteristic concentration: In a solution consistent with the measurement solution matrix, the characteristic concentration of cobalt shall not be greater than 0.035 µg/mL;
- Precision: Use the highest concentration of standard solution to measure the absorbance 10 times; the standard deviation shall not exceed 1.0% of the average absorbance; use the lowest concentration of standard solution (not the "zero" concentration standard solution) to measure the absorbance 10 times; the standard deviation shall not exceed 0.5% of the average absorbance of the highest concentration;
- The working curve's linearity: Divide the working curve into 5 equal segments according to the concentration, and; the ratio of the absorbance difference between the highest segment and the lowest segment shall not be less than 0.7. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 6682Water for analytical laboratory use - Specification and test methods
- GB/T 12806Laboratory glassware -- One-mark volumetric flasks
- GB/T 12808Laboratory glassware - One mark pipettes
- GB/T 12809Laboratory glassware -- Principles of design and construction of volumetric glassware
- GB/T 12810Laboratory glassware - Capacity methods testing and use of volumetric instruments
Similar standards
Editions of GB/T 4325.8
| Edition | Title | Revision | Status |
|---|---|---|---|
| GB/T 4325.8-2013 | Methods for chemical analysis of molybdenum - Part 8: Determination of cobalt content - 5-Cl-PADAB spectrophotometry and flame atomic absorption spectrometry | current edition | Current |
| GB/T 4325.7-1984 | Methods for chemical analysis of molybdenum - Part 8: Determination of cobalt content - 5-Cl-PADAB spectrophotometry and flame atomic absorption spectrometry | previous edition | Superseded |
This page sells the current edition, GB/T 4325.8-2013. Earlier editions are listed for reference only.
How to Download GB/T 4325.8-2013
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 12806-2011 — Laboratory glassware -- One-mark volumetric flasks
GB/T 12808-2015 — Laboratory glassware - One mark pipettes
GB/T 12809-2015 — Laboratory glassware -- Principles of design and construction of volumetric glassware
Secure payment via Stripe
Payments accepted
GB/T 4325.8-2013
$165.00