GB/T 42218-2022Clinical laboratory medicine - In vitro diagnostic medical devices - Validation of user quality control procedures by the manufacturer (English PDF)
Also coversGBT42218-2022
Open the GB/T 42218-2022 preview as PDF
This is a limited preview
Buy now to download the full PDF
Issued by
State Administration for Market Regulation, China National Standardization Administration
Level / Type
National · Recommended
Issue date
December 30, 2022
Implementation date
July 1, 2023
Scope
GB/T 42218-2022 (Clinical laboratory medicine - In vitro diagnostic medical devices - Validation of user quality control procedures by the manufacturer) is available as an English-translated PDF.
GB/T 42218-2022 — This document describes the process by which manufacturers of in vitro diagnostic medical devices validate quality control procedures recommended to users. Quality Control Procedures The purpose of the test is to provide the user with assurance that the performance of the device is consistent with its intended use and the manufacturer's declaration. This document applies to all in vitro diagnostic medical devices.
Document preview — GB/T 42218-2022
National Standard of the People's Republic of China
- ICS
- 11.100.10
- Classification
- C 30
Issued by: State Administration for Market Regulation, China National Standardization Administration
Contents
- Preface
- Introduction
- 1 Scope1
- 2 Normative references1
- 3 Terms and Definitions1
- 4 Quality Control Procedures4
- 5 Validation of quality control procedures5
foreword
This document is in accordance with the provisions of GB/T 1.1-2020 "Guidelines for Standardization Work Part 1.Structure and Drafting Rules for Standardization Documents"
drafting.
This document is modified to adopt ISO 15198.2004 "Quality control provided to users by manufacturers of in vitro diagnostic medical devices for laboratory medicine".
Confirmation of Manufacturing Procedures.
Compared with ISO 15198.2004, this document has made the following structural adjustments.
--- Deleted 3.3, 3.7, 3.12, 3.16~3.20, 3.23 of ISO 15198.2004;
--- Merge 3.10 and 3.11 of ISO 15198.2004 into one term.
The technical differences between this document and ISO 15198.2004 and their reasons are as follows.
--- Deleted the normative references ISO 3534-1, ISO 5725-1, added GB/T 29791.1, JJF1001 (see Chapter 3).
---The basic vocabulary and common terms of international metrology (VIM), BIPM, IEC, IFCC, ISO, IUPAC, IUPAP,
OIML, the second edition is replaced by ISO /IEC Guide 99 to adapt to the national conditions of our country.
---Change the term names and contents originally quoted in 3.2, 3.4~3.6, 3.8~3.11, 3.22, 3.24~3.26 to GB/T 29791.1-
The corresponding term names and contents in.2013 are adapted to my country's national conditions.
--- 3.1, 3.14, adopt the definition of terms in JJF1001-2011, 3.13 adopt the definition of ISO 13485.2016, 3.21 adopt
The definition in ISO 14971.2019 to adapt to my country's national conditions.
Please note that some contents of this document may refer to patents. The issuing agency of this document assumes no responsibility for identifying patents.
Introduction
Manufacturers of in vitro diagnostic medical devices usually incorporate quality control procedures into their instructions for use. The quality control program serves to provide
Provide the user with assurance that the device will perform as specified so that the results are consistent with the intended diagnostic use. For some medical devices, quality control
Procedures may be an essential risk control measure. Through the design of the medical device quality control program, the quality control program can also be passed
The following aspects help users ensure the quality of results.
a) verify the suitability of the analytical system (sample, reagents, instrument and/or operator);
b) monitoring the precision and correctness of measurement results;
c) prevent false negative and false positive results;
d) identify fault conditions that may lead to inaccurate results; and/or
e) Eliminate faults requiring corrective action.
In addition, manufacturers usually design in vitro diagnostic medical devices so that they can detect potential failures and warn users to take corrective actions.
ability to measure. These internal control systems can potentially reduce or even eliminate the need for users to run quality control samples to monitor device performance.
need.
This document can be used as part of the design controls and risk management program for manufacturers of in vitro diagnostic medical devices. This document also enables manufacturers to
Provides validated quality control procedures for clinical diagnostic laboratory users.
This document describes how manufacturers can validate the quality control procedures for their devices. Validation procedures ensure that the manufacturer is running quality controls as intended.
control procedures, and to ensure that the manufacturer's recommendations meet certain special medical devices, such as separation systems, products with built-in electronic controllers, and devices
The need for products with chemical and/or biological controls. Information on validated quality control procedures enhances user confidence in the overall quality of the device
Understanding of quality assurance requirements and making informed choices about appropriate control procedures.
Although laboratory directors are ultimately responsible for determining appropriate quality control procedures for their laboratory, in vitro diagnostic medical device
Manufacturers are responsible for providing users with sufficient information on device performance, risk control methods, and performance verification methods in the instructions. because
Thus, in fact, quality control is a shared responsibility of both the manufacturer and user of in vitro diagnostic medical devices.
Due to the essential differences in design, technology, function and intended use of in vitro diagnostic medical devices, no matter now or in the future,
No single quality control program will cover all IVD medical devices. Established quality control practices over the years have provided the laboratory with
provides some level of assurance that the results are valid. Although these quality controls have been widely accepted by laboratories, regulatory agencies, and accreditation bodies
Practice, but these practices originated from the manual operation of laboratory analysis and the stage of laboratory self-preparation of reagents. Regarding current in vitro diagnostic medical devices
However, they may not always be optimal. Therefore, when a quality control program is required, it is the responsibility of the manufacturer to design and validate
Appropriate quality control procedures.
Quality system standards for medical device manufacturers are evolving with the times. For example, ISO 13485 and most regulations at the time included design
control and risk management requirements. Design control requires a risk analysis of the design, and requires that the design meet the user requirements before the product is released to the market.
identified in terms of user requirements and intended use. Quality control procedures in instructions for use are considered inseparable from the design of in vitro diagnostic medical devices
Therefore, it is necessary to design the confirmed requirements.
Laboratory Medicine In Vitro Diagnostic Medical Devices
The manufacturer's quality control for the user
program confirmation
1 Scope
This document describes the process by which manufacturers of in vitro diagnostic medical devices validate quality control procedures recommended to users. Quality Control Procedures
The purpose of the test is to provide the user with assurance that the performance of the device is consistent with its intended use and the manufacturer's declaration.
This document applies to all in vitro diagnostic medical devices.
2 Normative references
The contents of the following documents constitute the essential provisions of this document through normative references in the text. Among them, dated references
For documents, only the version corresponding to the date is applicable to this document; for undated reference documents, the latest version (including all amendments) is applicable to
this document.
GB/T 29791.1 Information provided by manufacturers of in vitro diagnostic medical devices (labelling) Part 1.Terms, definitions and general requirements
Seek (GB/T 29791.1-2013, ISO 18113-1.2009, IDT)
ISO 13485 medical device quality management system for regulatory requirements (Medicaldevices-Qualitymanagement
Note. YY/T 0287-2017 Medical Device Quality Management System is used for regulatory requirements (ISO 13485.2016, IDT)
Note. Application of YY/T 0316-2016 Medical Device Risk Management to Medical Devices (ISO 14971.2007 Corrected Edition, IDT)
JJF1001 General Metrology Terms and Definitions
3 Terms and Definitions
GB/T 29791.1, ISO 13485, ISO 14971, JJF1001, ISO /IEC Guide 99 and the following terms and definitions apply
used in this document.
3.1
Accuracyaccuracy
The degree of agreement between the measured value of a measurand and its true value.
[Source. JJF1001-2011,5.8]
3.2
bias
An estimate of the systematic measurement error.
[Source. GB/T 29791.1-2013, A.3.25]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — all pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 29791.1In vitro diagnostic medical devices -- Information supplied by the manufacturer (labelling) -- Part 1: Terms, definitions and general requirements
- GB/T 29791.1-2013In vitro diagnostic medical devices -- Information supplied by the manufacturer (labelling) -- Part 1: Terms, definitions and general requirements
ISO 18113 · ISO 13485 · YY/T 0287-2017 · ISO 13485.2016 · YY/T 0316-2016 · ISO 14971.2007
How to Buy GB/T 42218-2022
- 1Add to cart. Click the "Buy GB/T 42218-2022" button on this page. You can add more standards before checkout.
- 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
- 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 29791.1-2013 — In vitro diagnostic medical devices -- Information supplied by the manufacturer (labelling) -- Part 1: Terms, definitions and general requirements
GB/T 46830.2-2026 — Plastics - Liquid crystal polymer (LCP) moulding and extrusion materials - Part 2: Preparation of test specimens and determination of properties
GB/T 28593-2026 — Sand and dust weather warning levels
Secure payment via Stripe
Payments accepted
GB/T 42218-2022
$200.00