Scope
GB/T 41063-2021 is the English-translated version of 玻璃纤维 密度的测定.
GB/T 41063-2021 sets out how to determine the density of glass fibre. It gives three test methods: method A with a gas pycnometer, which is the arbitration method, method B with a liquid pycnometer, and method C by buoyancy, weighing a glass rod in air and in distilled water.
For each method it lists the apparatus, such as analytical balances, constant temperature water baths holding 23 °C ± 0.1 °C and helium or nitrogen as the measuring gas, explains how specimens are prepared and freed of wetting agent by burning, and gives the operating steps and the density calculation.
It also covers humidity conditioning and the test environment, the test report and, in Annex A, the calibration of the pycnometer sample chamber and additional chamber volumes. It was issued on 31 December 2021 and has applied since 1 July 2022.
Found what you need?
Document preview — GB/T 41063-2021
National Standard of the People's Republic of China
- ICS
- 59.100.10
- Classification
- Q 36
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Terms and definitions4
- 4 Humidity conditioning and test environment5
- 5 Test methods5
- 6 Test report11
- Annex A (normative) Calibration of pycnometer sample chamber and additional chamber volumes12
1 Scope
This document specifies the humidity conditioning and test environment, test method and test report for determination of glass fiber density.
This document describes three test methods for determining the density of glass fibers, namely:
- Method A: Gas pycnometer method;
- Method B: Liquid pycnometer method;
- Method C: Buoyancy method.
Method A is the arbitration test method.
This document applies to the determination of glass fiber density.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
GB/T 6682, Water for analytical laboratory use - Specification and test methods
GB/T 9914.2, Test method for reinforcement products - Part 2: Determination of combustible-matter content for glass fibre
3 Terms and definitions
For the purposes of this document, the following terms and definitions apply.
3.1 Density; rho
The mass of a substance per unit volume at a specified temperature.
NOTE: Density is measured in kilograms per cubic meter (kg/m3), kilograms per liter (kg/L) or grams per cubic centimeter (g/cm3).
5 Test methods
5.1.2.2 Constant temperature water bath
The temperature of the sample chamber can be maintained at 23°C ± 0.1°C.
5.1.2.3 Analytical balance
It is readable to 0.1mg. The maximum allowable error is 0.5mg. The measuring range is 200g.
5.1.2.4 Measuring gas
It is recommended to use helium with a purity of not less than 99.99%. Other non-corrosive and non-adsorbed gases can also be used, such as nitrogen with a purity of not less than 99.99%.
5.1.2.5
Suitable tools, such as scissors or knife.
5.1.2.6
Vacuum pump (optional).
5.1.3 Specimen
Unless otherwise agreed, the specimen shall be free of wetting agent. To remove the wetting agent, the burning method specified in GB/T 9914.2 can be used. If the interested parties agree, the test specimen containing wetting agent can also be used.
Take a specimen of appropriate mass according to the volume of the sample chamber.
Cut the fibers with a suitable tool (5.1.2.5) to no higher than the height of the sample chamber.
5.1.4 Operation
5.1.4.1 Check the gas specific gravity meter (5.1.2.1) to ensure that there is no gas leakage in the instrument.
5.1.4.2 Weigh the mass (m1) of the sample chamber with the analytical balance (5.1.2.3), accurate to 0.1mg. Put the fiber in the sample chamber. Fill to more than 60% of the sample chamber capacity. Weigh the mass (m2) of the sample chamber and the specimen, accurate to 0.1mg.
5.1.4.3 Put the weighed sample chamber and specimen into the gas specific gravity meter. Seal the sample chamber. Adjust the constant temperature water bath (5.1.2.2) to keep the temperature of the sample chamber at 23°C ± 0.1°C. Access with the gas (5.1.2.4). Multiple pulse purges to remove air from instrument sample chamber and additional chamber. The vacuum pump (5.1.2.6) can also be used to remove air from the instrument sample chamber and additional chamber.
5.1.4.4 Open all valves. Fill the sample chamber and additional chamber with gas at standard atmospheric pressure. The pressure sensor should now read zero.
5.1.4.5 Close the valve of the additional chamber. Fill the sample chamber with gas until the desired pressure is reached. 1 standard atmosphere is recommended. Close the intake valve. The pressure sensor reading is recorded as p1. […]
5.2.2.4
Distilled water, which shall comply with the provisions of GB/T 6682.
5.2.2.5 Pycnometer
The capacity is 50mL.
5.2.2.6 Analytical balance
It is readable to 0.1mg. The maximum allowable error is 0.5mg. The measuring range is 200g.
5.2.2.7 Ultrasonic cleaner with flat bath
The power is not less than 400W.
5.2.2.8
Electric furnace, laboratory use, commercially available.
5.2.2.9 Constant temperature water bath
The temperature control accuracy is ±0.1°C.
5.2.3 Specimens
Specimens should be free of wetting agent. To remove the wetting agent, the burning method specified in GB/T 9914.2 can be used. Weigh 35g±2g of specimen. After removing the wetting agent, use an agate mortar to grind it to a length of 1mm~3mm.
Transfer it to a beaker. Use 3 times the specimen volume of absolute ethanol (5.2.2.1) to wash. Place still. Pour out the solvent. Repeat washing 2~3 times. Dry in the blast drying oven (5.2.2.2) at a temperature of 105°C±2°C for not less than 1h. Cool to room temperature in the desiccator (5.2.2.3).
5.2.4 Operation
5.2.4.1 Use distilled water (5.2.2.4) to rinse the pycnometer (5.2.2.5). Dry to constant weight together with the stopper at 105°C±2°C. Cool to room temperature in the desiccator. Use the analytical balance (5.2.2.6) to weigh its mass (m), accurate to 0.1mg.
5.2.4.2 Put the 8g~10g of specimen in the pycnometer through the funnel. Wipe off the powder stuck to the outside and mouth of the pycnometer. Weigh the mass (m1) of the pycnometer and the specimen, accurate to 0.1mg.
5.2.4.3 Fill 3/4 pycnometer with distilled water. Eliminate air bubbles by cleaning with the ultrasonic cleaner (5.2.2.7) or heating with the electric furnace (5.2.2.8). After removing the bubbles, place the pycnometer in the constant temperature water bath (5.2.2.9) at 23°C ± 0.1°C for at least 1h. Fill with 23°C±0.1°C distilled water. Use a stopper to plug. Set the volume to constant. Dry and weigh its mass (m2), accurate to 0.1mg.
5.2.4.4 Remove the specimen from the pycnometer. Filled with distilled water after de-bubbling. Place in a constant temperature water bath at 23°C±0.1°C for at least 1h. Take out. Use 23°C±0.1°C distilled water to set the volume tot constant. Dry and weigh its mass (m3), accurate to 0.1mg.
5.2.4.5 The density of the specimen is calculated according to formula (2): […]
5.3.2.8
Thermometer, with an accuracy of 0.1°C, capable of measuring the temperature of water.
5.3.2.9 Beaker
Appropriate volume (250mL~750mL). It can be placed in the balance. The gondola or wire loop specimen holder can be immersed in distilled water.
5.3.2.10 Support frame
Place the beaker across the balance tray.
5.3.2.11 Hanging basket, connected with the balance tray
The gondola is made of degreased wire with a diameter of less than 0.2mm.
5.3.3 Specimens
Perform sampling during glass fiber production drawing. Use the nickel-plated crucible tongs (5.3.2.1) to cut a section of glass rod directly from under the bushing. The mass is 2g±1g. Place in the specimen dish (5.3.2.2) to cool naturally. Repeat the interception to get multiple backups. Pick at least 5 glass rods that are free of visible bubbles, stones, streaks and so on. Use the file (5.3.2.3) to trim both ends into a smooth and flat specimen with slightly rounded edges. After repeated washing with distilled water (5.3.2.4) 2~3 times, put it in the blast drying oven (5.3.2.5) with a temperature of 105°C ± 2°C and dry it for not less than 1h. Cool to room temperature in the desiccator (5.3.2.6).
5.3.4 Operation
5.3.4.1 Use the analytical balance (5.3.2.7) to weigh the mass (m1) of the glass specimen, accurate to 0.1mg.
5.3.4.2 Place the beaker (5.3.2.9) with distilled water and the thermometer (5.3.2.8) inserted on the balance support (5.3.2.10). Measure and record the temperature of the water in the beaker. The temperature of the water should be 23°C ± 0.1°C.
5.3.4.3 Place the specimen in the hanging basket (5.3.2.11). Immerse the basket and the specimen into the water in the beaker. Weigh the mass of the specimen in water (m2), accurate to 0.1mg. During the test, the hanging basket and the specimen should not touch the beaker.
5.3.4.4 The density of the specimen is calculated according to formula (3):
Where, rho - The glass fiber density, in grams per cubic centimeter (g/cm3); […]
Annex A (normative) Calibration of pycnometer sample chamber and additional chamber volumes
A.1 Before starting the calibration procedure, open all valves to purge the instrument with gas. Purge the sample chamber and additional chamber. Fill both chambers with gas at standard atmospheric pressure. Make the pressure sensor read zero.
A.2 Close the valve of the additional chamber. Fill the sample chamber with gas until the desired pressure is reached. 0.1MPa is recommended. Close the intake valve. The pressure sensor reading is recorded as.
A.3 Open the valve of the additional chamber. Let the gas in the sample chamber enter the additional chamber. The pressure sensor reading is recorded as.
A.4 Put a standard sample of known volume (Vc) in the sample chamber. Repeat A.2 and A.3. Fill the sample chamber with gas again until the desired pressure is reached.
The pressure sensor reading is recorded as. After the gas enters the additional chamber, the pressure sensor reading is recorded as.
A.5 The volume of the sample chamber and the additional chamber can be calculated by formula (A.1) and formula (A.2), respectively:
Where, - The pressure of the empty sample chamber before and after the gas diffuses into the additional chamber, in Pascals (Pa);
- The pressure of the sample chamber where the standard sample is placed before and after the gas diffuses into the additional chamber, in Pascals (Pa); […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
Similar standards
How to Download GB/T 41063-2021
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 6682-2008 — Water for analytical laboratory use - Specification and test methods
GB/T 9914.2-2013 — Test method for reinforcement products - Part 2: Determination of combustible-matter content for glass fibre
GB/T 7251.4-2026 — Low-voltage switchgear and controlgear assemblies - Part 4: Particular requirements for assemblies for construction sites (ACS)
Secure payment via Stripe
Payments accepted
GB/T 41063-2021
$150.00