GB/T 4079-2025Test procedures for amplifiers and charge-sensitive preamplifiers used with detectors of ionizing radiation (English PDF)

用于电离辐射探测器的放大器和电荷灵敏前置放大器的测试方法

Preview PDF

This is a limited preview

Buy now to download the full PDF (31 pages)

Need quotation for your management?

Get a formal quotation for GB/T 4079-2025 in 30 minutes.

Issued by

SAMR; SAC

Level / Type

National · Recommended

Issue date

October 5, 2025

Implementation date

October 5, 2025

Scope

GB/T 4079-2025 is the English-translated version of 用于电离辐射探测器的放大器和电荷灵敏前置放大器的测试方法.

GB/T 4079-2025 is the Chinese national standard for testing main amplifiers and charge-sensitive preamplifiers used with ionizing radiation detectors. It defines the vocabulary of the field, from shaping amplifier and baseline restorer to walk, energy resolution and full width at half maximum, and lists the symbols used for pulse timing.

It fixes the test conditions: the reference environment of GB/T 8993, the test schematic with its 50 ohm terminating resistors, and the pulse generator, attenuator and capacitor box.

The remaining clauses give the test methods for the main parameters of the main amplifier and the key parameters of the preamplifier. It applies to amplifiers for semiconductor detectors, gas pulse ionization chambers and proportional counters, and to analogue main amplifiers for scintillation detectors.

It replaces GB/T 4079-1994, was issued on 5 October 2025 and has applied since the same day.

Found what you need?

Document preview — GB/T 4079-2025

National Standard of the People's Republic of China

ICS
27.120.20
Classification
F 80
Replacing
GB/T 4079-1994

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • Foreword3
  • 1 Scope5
  • 2 Normative references5
  • 3 Terms and definitions5
  • 4 Symbols9
  • 5 Test conditions10
  • 6 Test methods for the main parameters of the main amplifier13
  • 7 Test methods for key parameters of the preamplifier23
  • Bibliography31

1 Scope

This document describes the test methods for the main amplifier (shaping amplifier) and charge-sensitive preamplifier used in ionizing radiation detectors.

This document applies to main amplifiers and charge-sensitive preamplifiers for semiconductor detectors, gas pulse ionization chambers, proportional counters, and analog main amplifiers for scintillation detectors.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

GB/T 8993, Environmental conditions and testing procedures for nuclear instrumentation

3 Terms and definitions

For the purposes of this document, the following terms and definitions apply.

3.1 Charge-sensitive preamplifier

A preamplifier of which its output signal is proportional to the input charge and is essentially independent of the input capacitance. [Source: GB/T 4960.6-2008, 3.1.25]

3.2 Main amplifier; shaping amplifier; spectrum amplifier

A line pulse amplifier used for amplitude spectrum measurement.

NOTE: The main amplifier, also known as a shaping amplifier, typically features pulse shaping, anti-stacking, and baseline recovery functions, along with sufficiently good stability and a suitable spectral response width.

3.19 Differential nonlinearity (of measuring assembly)

(1) The maximum deviation of the slope of the output and input relationship curve from the slope of the reference line.

(2) Fluctuations in the uniformity of individual channel width in a multichannel analyzer.

NOTE: It is expressed as a percentage. [Source: GB/T 4960.6-2008, 3.2.41, modified]

3.20 Bipolar pulse

A raised pulse on each side of the baseline.

3.21 Baseline

The average level.

NOTE: In the case of non-overlapping pulses, the pulse deviates from this level and then returns to this level.

3.22 Baseline restorer; BLR

The line that quickly restores the baseline to its previous level after the amplifier outputs a pulse (or a series of pulses).

3.23 Offset

The DC voltage deviates from the specified voltage (or current) level.

NOTE: Unless otherwise stated, this level is the baseline.

3.24 Energy resolution (of a radiation spectrometer)

The smallest difference in energy between two particles that a radiation spectrometer can distinguish.

NOTE: Energy resolution is typically represented by a factor. This factor is the full width at half maximum (energy) of the peak in the monoenergetic particle distribution curve divided by the energy at the peak position. [Source: GB/T 4960.6-2008, 3.2.26]

3.25 Resolving time

The minimum time interval between two consecutive pulses that can still be distinguished. [Source: GB/T 4960.6-2008, 3.2.20]

3.26 Terminating resistor

A resistor connected across the output of an amplifier or signal generator.

NOTE: The purpose of terminating resistors is to eliminate signal reflections at these terminals.

3.27 Characteristic impedance

The internal resistance (impedance) of a network (such as a coaxial cable or an attenuator).

NOTE: When a resistor with the same impedance is used for termination, signal reflection can be avoided and the output signal is halved.

3.28 Walk

Changes in pulse height in the amplifier cause changes in the zero-crossing time.

3.29 Full width at half maximum; FWHM

On a distribution curve formed by a single peak, the distance between the x-coordinates of two points on the curve at half the peak value. [Source: GB/T 4960.6-2008, 3.2.27]

4 Symbols

The following symbols apply to this document.

t0.01: the minimum time interval between two consecutive, distinguishable pulses.

t1/2: the time interval between two points in a unipolar pulse waveform where the amplitude is half the peak amplitude, in microseconds (µs).

tp1: the time to peak of a bipolar pulse measured from 1% of the peak height of the first half-cycle to its peak center, in microseconds (µs).

tp2: the time to peak of a bipolar pulse measured from 1% of the peak height of the first half-cycle to the peak center of the second peak, in microseconds (µs).

tx0: the transit time of a bipolar pulse, in microseconds (µs).

5 Test conditions

5.1 Test environment conditions

The reference and standard test conditions for calibration and testing of the equipment under varying influence conditions covered in this document shall comply with the requirements of GB/T 8993.

5.2 Test schematic

The test schematic for general amplifier parameters is shown in Figure 1. The terminating resistor R0 for both the pulse generator and the amplifier is typically 50 ohm. The cable selection shall match the impedance of R0.

5.3 Test equipment requirements

5.3.1 Pulse generator
5.3.1.1 Basic requirements for pulse generators

The pulse generator has a maximum output amplitude Up greater than or equal to 10 V.

The amplitude of the output pulse can be adjusted in minimum intervals of 10 mV.

Unless otherwise specified, Up equals 10 V. Different Up values or adjustment methods shall only affect certain details of the measurement and shall not affect the overall measurement method.

The pulse generator shall have at least two outputs. one direct output and one attenuated output. Each output shall have its own termination resistor and the same internal resistance r0 (r0 is 0.5 ohm or less). There shall also be an output for triggering oscilloscope scanning.

The pulse generator can operate at the power supply frequency or other frequencies.

Operating at other frequencies can be used to detect AC hum interference in the system.

5.3.1.2 Verify the pulse generator of the preamplifier

A pulse generator with a flat-top pulse rise time (tr) less than or equal to 1 ns is generally used. Alternatively, a mercury relay tail pulse generator with a tr less than or equal to 5 ns can be used for all tests except for verifying the exponential decay performance of a charge-sensitive preamplifier.

Generally, the pulse generator's tr (trajectory value) shall match the detector's tr. When verifying preamplifiers connected to semiconductor and gas detectors, because the detector's collection time varies, the pulse generator's tr shall not exceed 1/3 of the shortest tr of the detector or preamplifier system. When verifying preamplifiers connected to scintillation detectors, the pulse generator's tr shall be less than the preamplifier's tr. If the scintillation detector's tr is greater than the preamplifier's tr, the pulse generator shall be readjusted to match the scintillation detector's tr.

The fall time of the pulse generator is faster than that of the preamplifier.

5.3.1.3 Verify the pulse generator of the main amplifier

When verifying overload or pole-zero cancellation, a tail pulse shall be used. For other tests, a rectangular pulse generator can be used, or a tail pulse generator can also be used. The tail pulse can be obtained from a flat-topped pulse by simply adding a capacitor between the pulse generator and the attenuator.

5.3.2 Attenuator

The attenuator can be part of the pulse generator or externally connected. Connected to the output of the pulse generator, each attenuator setting shall be matched to the coarse gain adjustment of the main amplifier. The accuracy and stability of the attenuator depend on the stability and accuracy of the connected resistors. The error caused by switching each setting on shall not exceed 1% of the output signal for that setting.

5.3.3 Capacitor box

Figure 1 shows the circuit diagram of the capacitor box. The capacitor box contains a test capacitor Cc and a set of parallel capacitors Ct switched by a switch. The test […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 31 pages — is available in the English PDF.

Referenced standards

Editions of GB/T 4079

EditionTitleRevisionStatus
GB/T 4079-2025Test procedures for amplifiers and charge-sensitive preamplifiers used with detectors of ionizing radiationcurrent editionCurrent
GB/T 4079-1994Test procedures for amplifiers and charge-sensitive preamplifiers used with detectors of ionizing radiationprevious editionSuperseded

This page sells the current edition, GB/T 4079-2025. Earlier editions are listed for reference only.

How to Download GB/T 4079-2025

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
31 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 4079-2025

$315.00