GB/T 37049-2018Test method for the content of metal impurity in electronic grade polysilicon—Inductively coupled-plasma mass spectrometry method (English PDF)

电子级多晶硅中基体金属杂质含量的测定 电感耦合等离子体质谱法

Preview PDF

This is a limited preview

Buy now to download the full PDF (15 pages)

Need quotation for your management?

Get a formal quotation for GB/T 37049-2018 in 30 minutes.

Issued by

SAMR; SAC

Level / Type

National · Recommended

Issue date

December 28, 2018

Implementation date

April 1, 2019

Scope

GB/T 37049-2018 is the English-translated version of 电子级多晶硅中基体金属杂质含量的测定 电感耦合等离子体质谱法.

GB/T 37049-2018 is a Chinese national test method for measuring trace metal impurities in electronic-grade polysilicon by inductively coupled plasma mass spectrometry (ICP-MS).

It covers iron, chromium, nickel, copper, zinc and sodium at contents below 5 ng/g, as defined in GB/T 12963. The sample is dissolved by heating in a mix of nitric and hydrofluoric acid, the silicon is driven off as silicon tetrafluoride, and the residue is taken up in nitric acid and measured by ICP-MS.

The standard lists the interfering factors, from operator cleanliness and instrument contamination to isobaric interferences and reagent purity, and specifies the reagents, instruments, test environment, sample preparation, test procedure, calculation, analysis of results and test report. It was issued on 28 December 2018 and has applied since 1 April 2019.

Found what you need?

Document preview — GB/T 37049-2018

National Standard of the People's Republic of China

ICS
77.040.30
Classification
H 17

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative references4
  • 3 Terms and definitions4
  • 4 Summary of method5
  • 5 Interfering factors6
  • 6 Reagents7
  • 7 Instruments and equipment7
  • 8 Test environment8
  • 9 Sample preparation8
  • 10 Test procedure8
  • 11 Calculation of analysis results10
  • 12 Analysis of test results11
  • 13 Test report11

1 Scope

This standard specifies a method for determining the content of trace metallic impurities in electronic-grade polysilicon, using inductively coupled plasma mass spectrometry (ICP-MS).

This standard applies to the determination of iron, chromium, nickel, copper, zinc, sodium contents in the range of less than 5 ng/g for metallic impurities, as defined in GB/T 12963.

2 Normative references

The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) is applicable to this standard.

GB/T 12963 Electronic-grade polycrystalline silicon

GB/T 14264 Terminology of semiconductor materials

GB/T 25915.1-2010 Cleanrooms and associated controlled environments - Part 1: Classification of air cleanliness

JJF 1159 Calibration specification for quadrupole inductively coupled plasma mass spectrometers

SEMI C1 Guide for the analysis of liquid chemicals

SEMI F63 Guide for ultrapure water used in semiconductor processing

3 Terms and definitions

The terms and definitions defined in GB/T 14264 and JJF 1159, as well as the following terms and definitions, apply to this document.

3.1 Resolution

Expressed as the peak width at 10% of the peak height of a specific element's mass peak.

3.2 Detection limit

The lowest limiting mass concentration of a specific element, that the mass spectrometer can measure within a certain confidence level.

3.3 Instrument detection limit

The concentration value corresponding to three times the standard deviation of the signal average for a reagent blank (pure water).

3.4 Method detection limit

The minimum concentration of an analyte, that can be detected by a given analytical method under specific conditions with a reasonable confidence level.

3.5 Background noise

The signal response for a specific element's ions generated by the mass spectrometry detection system, when ions of that element have not been introduced.

3.6 Mass stability

The degree of shift in the center of a specific element's mass peak over an extended period.

4 Summary of method

4.1 The polycrystalline silicon sample is dissolved by heating in a mixed acid of nitric acid and hydrofluoric acid; the solution is then evaporated to dryness or near-dryness, causing the silicon in the solution to volatilize as silicon tetrafluoride (SiF4). The residue is subsequently dissolved in nitric acid solution and diluted to volume with pure water; the content of the metal elements to be analyzed is then determined, using an inductively coupled plasma mass spectrometer (ICP-MS). The total detection limit for the six elements. iron, chromium, nickel, copper, zinc, sodium, using the test method specified in this standard shall be less than 0.4 ng/g.

4.2 The test sample is introduced into the high-temperature plasma source via the sample introduction system, where it undergoes evaporation, dissociation, atomization, ionization within the high-temperature torch tube. The vast majority of metal atoms are converted into singly charged ions. These ions travel at high speed through the dual-cone interface into the ion lens, where they are focused into an ion beam by an electric field before entering the quadrupole ion separation system. Upon entering the quadrupole mass analyzer, the ions are separated sequentially based on their mass-to-charge ratios. Finally, they are detected by an ion detector; the resulting signal is amplified and measured by the signal detection system.

5 Interfering factors

5.1 The cleanliness-related proficiency of the testing personnel affects the results. Personnel shall undergo rigorous training in trace analysis, possess relevant cleanroom experience, be capable of controlling potential contamination at every stage of the experiment.

5.2 The Inductively Coupled Plasma Mass Spectrometer (ICP-MS) shall be dedicated to specific applications and maintained regularly to ensure high sensitivity and stability at low concentrations. Severe contamination may render the instrument difficult to restore to normal operating conditions; therefore, testing samples with metal impurity concentrations exceeding 10 ng/g shall be avoided. Otherwise, the instrument's sensitivity at lower concentrations shall be evaluated beforehand. Prior to each test, quality control samples shall be verified, including checks before, during, after the analysis, to ensure the instrument is functioning correctly.

5.3 When selecting the mass number for the target element, an appropriate interference-reduction mode shall be chosen, based on potential isobaric interferences, to ensure accurate results.

5.4 Materials used for testing shall meet standard requirements and be protected from contamination. Before testing each sample, the cleanliness of pure water, nitric acid, hydrofluoric acid, laboratory ware shall be verified against requirements; any non-compliance invalidates the test results.

5.5 Samples shall be representative. Surface layers shall be excluded, as surface contamination can affect results; surface etching may be performed if necessary. Sample volume and mass shall be uniform, to avoid affecting the detection results. The sample mass to be weighed shall be determined based on the expected impurity content.

5.6 Sample preparation shall strictly adhere to standard testing procedures, as contamination during this stage can severely compromise the test results.

5.7 Intermediate standard solutions and test solutions shall be prepared on the day of use; they should be prepared in small quantities and not stored for long periods, as contaminated standard solutions can significantly affect test results.

5.8 The cleanliness of the testing environment shall be verified; the cleanliness of the environment, particularly during sample preparation and testing, can significantly impact the detection limit and test results.

5.9 The detection limits of the equipment and method significantly affect the reliability of test results. Good repeatability within a single laboratory does not guarantee the accuracy and reliability of the equipment's test data; therefore, the inductively coupled plasma mass spectrometer (ICP-MS) shall be calibrated periodically in accordance with JJF 1159; spike recovery tests shall be used to verify the reliability of the testing.

6 Reagents

6.1 Purified water. Compliant with SEMI F63 specifications.

6.2 Nitric acid. Mass fraction of 65.0% ~ 68.0%; content of each individual metal impurity < 10 ng/L.

6.3 Hydrofluoric acid. Mass fraction of 30.0% ~ 50.0%; content of each individual metal impurity < 10 ng/L.

6.4 Sulfuric acid. Mass fraction of 95.0% ~ 98.0%; content of each individual metal impurity < 10 ng/L.

6.5 Standard stock solutions. Concentrations of Fe, Cr, Ni, Cu, Zn, Na, Y, Co are all 1 g/L; certified reference materials with metrological traceability (domestic or international) shall be used.

6.6 Sulfuric acid solution. Dilute sulfuric acid (6.4) with purified water to a mass fraction of 16.7%.

6.7 Nitric acid. Nitric acid and purified water mixed at a volume ratio of 3:97 or another appropriate ratio.

6.8 Digestion solution. Nitric acid and hydrofluoric acid mixed at a volume ratio of 1:2 or another appropriate ratio.

7 Instruments and equipment

7.1 Inductively coupled plasma mass spectrometer (ICP-MS). Equipped with […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download GB/T 37049-2018

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 37049-2018

$150.00