Scope
GB/T 35952-2018 is the English-translated version of 化妆品中十一烯酸及其锌盐的测定 气相色谱法.
GB/T 35952-2018 sets out a gas chromatographic method for measuring undecylenic acid and zinc undecylenate in cosmetics - creams, emulsions and lotions - and in hand sanitizers.
Both substances are methylated with a sulfuric acid-methanol solution, the methyl undecylenate formed is extracted into n-hexane, filtered through a 0.45 µm filter and injected into the gas chromatograph, and the result is quantified by the external standard method.
The standard gives the reagents and materials, the instruments, the sample treatment, the chromatographic conditions on an HP-5 capillary column or a column of equivalent polarity, the standard working curve, the result calculation, recovery and precision, and the allowable difference.
The detection limit is 12 mg/kg and the quantitation limit 40 mg/kg, both expressed as acid. It was issued on 6 February 2018 and has applied since 1 September 2018.
Found what you need?
Document preview — GB/T 35952-2018
National Standard of the People's Republic of China
- ICS
- 71.100.70
- Classification
- Y 42
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Principle4
- 4 Reagents and materials4
- 5 Instruments5
- 6 Determination steps5
- 7 Result calculation7
- 8 Recovery and precision7
- 9 Allowable difference8
- Appendix A (Informative) Standard substance information of undecylenic acid and zinc undecylenate9
- Appendix B (Informative) Gas chromatogram of methyl esterification of undecylenic acid standard solution10
1 Scope
This Standard specifies the gas chromatographic method for the determination of undecylenic acid and zinc undecylenate in cosmetics (creams, emulsions, lotions) and hand sanitizers.
This Standard applies to the determination of undecylenic acid and zinc undecylenate in cosmetics (creams, emulsions, lotions) and hand sanitizers.
The detection-limit of this method for undecylenic acid and zinc undecylenate is 12 mg/kg (0.001 2%), and the quantitation-limit is 40 mg/kg (0.004%) (both in terms of acid).
2 Normative references
The following documents are indispensable for the application of this document.
For dated references, only the dated version applies to this document. For undated references, the latest edition (including all amendments) applies to this document.
GB/T 6682, Water for analytical laboratory use. Specification and test methods
3 Principle
After the undecylenic acid and zinc undecylenate are methylated by the sulfuric acid-methanol solution, use n-hexane to extract the produced methyl undecylenate; filter the n-hexane layer through a 0.45 µm filter; then, inject it into the gas chromatograph for detection; use the external standard method to quantify.
4 Reagents and materials
Unless otherwise specified, all the reagents are analytical reagents, and the water is grade-1 water which is specified by GB/T 6682.
4.1 Methanol: chromatographic pure. […]
6 Determination steps
6.2 Sample treatment
Accurately weigh 0.25 g of sample; add 5 mL of 5% sulfuric acid-methanol solution (4.7); perform ultrasonic extraction for 10 min. The methyl esterification steps are the same as 6.1.
6.3 Determination
6.3.1 Chromatographic conditions
The chromatographic conditions are shown as below:
a) chromatographic column: HP-5 capillary column (30.0 m × 0.25 mm × 0.25 µm) or a column of equivalent polarity;
b) heating procedure: keep the initial temperature of 100°C for 2 min; rise the temperature to 240°C at 15°C/min; then, keep it for 5 min;
c) temperature of the injector: 250°C; temperature of the detector: 260°C;
d) carrier gas flow velocity: 1 mL/min.
e) injection volume: 1 µL;
f) split ratio: 30:1.
6.3.2 Drawing of the standard working curve
Inject the undecylenic acid standard working solution (6.1) into the gas chromatograph; detect it according to the chromatographic conditions of 6.3.1; respectively take the chromatographic peak area of the undecylenic acid methyl ester as the ordinate, and the total content of undecylenic acid in the system before methyl esterification as the abscissa to draw the standard working curve of undecylenic acid. See Figure B.1 in Appendix B for the gas chromatogram of methyl esterification of undecylenic acid standard solution.
6.3.3 Sample determination
Inject the sample solution (6.2) into the gas chromatograph; measure according to the chromatographic conditions of 6.3.1. The peak area of the chromatographic peak can be used to determine the total undecylenic acid content from the standard curve; the response value of the measured substance in the sample solution shall be within the linear range of the standard curve. When the content of the measured substance is higher than the upper […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
Similar standards
How to Download GB/T 35952-2018
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 6682-1992 — Water for analytical laboratory use--Specification and test methods
GB/T 6682-2008 — Water for analytical laboratory use - Specification and test methods
GB/T 10003-2026 — Biaxially oriented polypropylene (BOPP) film
GB/T 35952-2018
$125.00