Scope

GB/T 3286.11-2022 is the English-translated version of 石灰石及白云石化学分析方法 第11部分:氧化钙、氧化镁、二氧化硅、氧化铝及氧化铁含量的测定 波长色散X射线荧光光谱法(熔铸玻璃片法).

GB/T 3286.11-2022 is the part of the Chinese series on the chemical analysis of limestone and dolomite that determines calcium oxide, magnesium oxide, silicon dioxide, aluminium oxide and iron oxide by wavelength dispersive X-ray fluorescence spectrometry.

It gives the principle, the apparatus, the reagents, the preparation of the fused bead, the calibration with reference materials, the measurement, the calculation of each oxide, the precision and the test report, with the determination range of every element tabulated.

It applies to limestone and dolomite, the fluxes and raw materials of the steel, cement and glass industries. Quarries, steelworks and cement plants use it. Issued on 9 March 2022, it has applied since 1 October 2022.

Found what you need?

Document preview — GB/T 3286.11-2022

National Standard of the People's Republic of China

ICS
73.080
Classification
D 52

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • Foreword3
  • Introduction5
  • 1 Scope6
  • 2 Normative references6
  • 3 Terms and definitions7
  • 4 Principle7
  • 5 Reagents and materials7
  • 6 Instruments and equipment8
  • 7 Sampling and sample preparation8
  • 8 Preparation of fused cast bead9
  • 9 Preparation of instruments10
  • 10 Analysis steps10
  • 11 Calculation and expression of results13
  • 12 Precision13
  • 13 Test report14
  • Annex A (informative) Recommended X-ray spectrometer measurement conditions15
  • Annex B (normative) Procedure for acceptance of sample analysis results16
  • Annex C (informative) Raw data of precision tests17

1 Scope

This document specifies the determination of calcium oxide, magnesium oxide, silicon dioxide, aluminium oxide and iron oxide content in limestone and dolomite by wavelength dispersive X-ray fluorescence spectrometry.

This document applies to the determination of calcium oxide, magnesium oxide, silicon dioxide, aluminium oxide and iron oxide content in limestone and dolomite. The determination range is shown in Table 1.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

GB/T 2007.2 General rules for the sampling and sample preparation of minerals in bulk - Sample preparation by manual method

GB/T 6682 Water for analytical laboratory use - Specification and test methods

GB/T 8170 Rules of rounding off for numerical values & expression and judgement of limiting values

GB/T 15000.3 Directives for the work of reference materials - Part 3: Reference materials - Characterization and assessment of homogeneity and stability

GB/T 16597 Analytical methods of metallurgical products - General rule for X-ray fluorescence spectrometric methods

JJG 810 Wavelength dispersive X-ray fluorescence spectrometers

YB/T 082 Specification for certified reference materials for metallurgical product analysis

3 Terms and definitions

There are no terms or definitions to be defined in this document.

4 Principle

The sample is ignited at high temperature and melted with borate to prepare fused cast beads. The sample on the surface layer is irradiated by primary X-rays to produce characteristic X-ray fluorescence. After spectroscopy by the crystal, the detector measures the X-ray fluorescence intensity at the 2theta angle corresponding to the selected characteristic wavelength.

5 Reagents and materials

Unless otherwise stated, only approved analytical reagents are used in the analysis, and the water for test shall be distilled water of Grade 3 or above specified in GB/T 6682 or water with a purity equivalent to it.

5.1 Ammonium iodide. solid, analytical reagent.

5.2 Lithium borate mixed flux (lithium tetraborate. lithium metaborate = 67.33). solid; it shall be in an anhydrous dry state, otherwise it needs to be burned at 500 °C ± 10 °C for 4 hours, cooled to room temperature, and then placed in a desiccator for later use.

5.4 Argon methane gas (90 % Ar - 10 % CH4). selected according to the instrument.

6 Instruments and equipment

6.1 Melting furnace

The temperature field of the melting furnace shall be uniform. It shall be able to maintain a temperature of at least 1100 °C ± 10 °C. It can choose an electric melting furnace, a gas melting furnace or a high-frequency induction melting furnace.

6.3 Crucible and mold

The crucible and mold (or the crucible that is also used as a mold) are made of non- wetting platinum-gold alloy (95 % Pt - 5 % Au). The crucible shall have a certain thickness to prevent deformation after heating. The bottom of the mold shall be kept flat. For directly formed crucibles, there shall be a flat bottom.

6.4 Balance

The minimum division is 0.1 mg.

6.5 High-temperature furnace

The maximum working temperature for a long time shall reach a temperature of 1100 °C.

6.6 Oven

It is able to control the temperature at 105 °C ± 5 °C.

7 Sampling and sample preparation

7.1 Sampling and sample preparation shall be carried out according to the provisions of

GB/T 2007.2.All samples shall pass through the 0.125 mm sieve.

7.2 Limestone and dolomite samples shall be dried in an oven at 105 °C ± 5 °C for 2 hours before analysis, and then placed in a desiccator to cool to room temperature. The retention time of the dried sample shall not exceed 8 hours, otherwise it shall be dried again.

8 Preparation of fused cast bead

8.1 Ignition treatment of the sample

Weigh 1.50 g ± 0.01 g of the sample that has been dried and cooled to room temperature (see Clause 7) and place it in a square porcelain dish. Place the dish into a high- temperature furnace (see 6.5) and ignite at 1050 °C ± 50 °C until constant weight.

8.2 Calculation of loss on ignition

Take out the sample from the high-temperature furnace, place it in a desiccator, cool it to room temperature, weigh it, and calculate the loss on ignition of the sample according to formula (1).

8.3 Preparation of the fused bead of the sample

Accurately weigh 0.7000 g ± 0.0005 g of the sample that has been ignited and dried and cooled to room temperature (see 8.2), place it in a glass stirring dish, add 7.0000 g ± 0.0005 g of borate mixed flux (see 5.2) powder, and mix well. Completely transfer it to a platinum-gold crucible (see 6.3).

8.4 Fused cast bead

Transfer the crucible shown in 8.3 to the melting furnace (see 6.1), pre-melt at 1050 °C ± 20 °C for 2 minutes, melt for 13 minutes, take it out, and directly shape or mold it into a measurable glass piece.

9 Preparation of instruments

9.1 Working environment of instruments

The working environment of the instrument shall comply with the provisions of GB/T 16597.

10 Analysis steps

10.1 Measurement conditions Select appropriate

measurement conditions based on the type of instrument used, element to be analyzed, coexisting elements, and their content variation ranges.

10.2 Plotting and confirmation of calibration curve

10.2.1 Plotting of calibration curve

Under selected working conditions, use an X-ray fluorescence spectrometer to measure the fused cast beads of a series of reference materials/standard samples. The lowest number of points on the working curve shall be no less than 8 points, and each sample bead shall be measured at least twice.

10.3 Analysis of the unknown sample

10.3.1 Standardization of instruments

Regularly perform standardization confirmation on the instrument, usually using a fixed sample to check whether there is a significant change in the X-ray intensity of the element corresponding to the component to be measured (see Annex A). If there is a significant change, it indicates that the instrument has drifted.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 23 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download GB/T 3286.11-2022

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
23 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 3286.11-2022

$255.00