GB/T 30903-2026Inorganic chemicals for industrial use - Determination of impurity elements - Inductively coupled plasma mass spectrometry (English PDF)
无机化工产品 杂质元素的测定 电感耦合等离子体质谱法(ICP-MS)
This is a limited preview
Buy now to download the full PDF (18 pages)
Need quotation for your management?
Get a formal quotation for GB/T 30903-2026 in 30 minutes.
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
March 31, 2026
Implementation date
October 1, 2026
Scope
GB/T 30903-2026 is the English-translated version of 无机化工产品 杂质元素的测定 电感耦合等离子体质谱法(ICP-MS).
GB/T 30903-2026 is the Chinese national method for measuring metallic and non-metallic impurity elements in inorganic chemical products by inductively coupled plasma mass spectrometry.
It sets out the principle, the general provisions on reagents, water and impurity standard solutions, the apparatus, the test procedure, the precision and the recovery rate.
Sample preparation covers dissolution, fusion with alkali metal fluxes at 500 to 1 200 degrees Celsius, wet digestion in open vessels, sealed high-pressure vessels or a microwave, and separation and preconcentration; the test solution must carry not more than 0.2% total dissolved solids.
Informative annexes give a sub-boiling acid purification device, the way to determine the detection limit and the recommended analyte mass numbers and internal standards.
It replaces GB/T 30903-2014, was issued on 31 March 2026 and applies from 1 October 2026.
Found what you need?
Document preview — GB/T 30903-2026
National Standard of the People's Republic of China
- ICS
- 71.060.01
- Classification
- G 10
- Replacing
- GB/T 30903-2014
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword3
- 1 Scope5
- 2 Normative references5
- 3 Terms and definitions6
- 4 Principle7
- 5 General provisions7
- 6 Apparatus8
- 7 Test procedure8
- 8 Precision12
- 9 Recovery rate13
- Appendix A (Informative) Sub-boiling distilled acid purification device14
- Appendix B (Informative) Detection limit determination method16
- Appendix C (Informative) Recommended analyte mass numbers and internal standards17
- References18
1 Scope
This document provides reagents or materials, apparatus for the determination of metallic and non-metallic impurity elements in inorganic chemical products by inductively coupled plasma mass spectrometry (ICP-MS). It describes the principle, general provisions, apparatus, test procedures, precision, and recovery rate of the determination of metallic and non-metallic impurity elements in inorganic chemical products by inductively coupled plasma mass spectrometry (ICP-MS).
This document applies to the determination of the content of metallic and non-metallic impurity elements in inorganic chemical products by inductively coupled plasma mass spectrometry (ICP-MS).
2 Normative references
The provisions of the following documents constitute the essential clauses of this document through normative references in this text. Among them, for any dated reference, only the version corresponding to that date applies to this document; for any undated reference, the latest version (including all amendments) applies to this document.
GB/T 6379.2 Accuracy (trueness and precision) of measurement methods and results - Part 2: Basic method for the determination of repeatability and reproducibility of a standard measurement method
GB/T 14666-2025 Terms for analytical chemistry
3 Terms and definitions
The terms and definitions defined in GB/T 14666-2025, as well as the following terms and definitions, apply to this document.
3.1 Plasma
Ionized gas with a degree of ionization greater than 0.1%.
NOTE: It is composed of electrons, ions, atoms and molecules, with the numbers of electrons and ions being roughly equal, and the whole is neutral. [Source: GB/T 14666-2025, 7.1.22]
3.2 Inductively coupled plasma; ICP
The torch flame that is formed by applying high-frequency power to a coil coupled to a plasma torch.
3.3 Sampling cone
A crucial cone-shaped component for the interface between atmospheric pressure plasma and high vacuum mass analyzers.
NOTE: There is a cone-shaped hole of about 1 mm in the center of the sampling cone. Through this cone-shaped hole, ions in the plasma torch enter the first-stage vacuum chamber supported by a mechanical pump.
3.4 Skimmer cone
Another important cone-shaped component for the interface between atmospheric pressure plasma and a high vacuum mass analyzer.
NOTE: The center of the skimmer cone has a cone hole less than 1 mm in diameter, which is concentric with the axis of the sampling cone hole. Ions enter the mass analyzer through this cone hole.
3.5 Auxiliary gas
Argon gas flow that is introduced between the middle tube and the central tube of the torch.
NOTE: The auxiliary gas serves to support the plasma formed at the torch opening, control the position of the plasma flame, and protect the central tube.
3.6 Carry gas
The airflow introduced through the central tube.
NOTE: The role of the carry gas is to transport aerosols into the plasma.
3.7 Cooling gas
During the operation of the inductively coupled plasma mass spectrometer, the gas that provides a central channel for the sample (nebulized gas), participates in the plasma ignition process, forms a stable plasma torch flame, and isolates and protects the torch.
NOTE: The cooling gas is usually argon.
3.8 Interface
A device consisting of a sampling cone, a skimmer cone, etc., allows ions formed by plasma to enter the channel of a mass separator.
3.9 Mass spectral interference
Interference caused by one or more atomic ions having a similar mass-to-charge ratio to the analyte ions.
3.10 Polyatomic ion interference
Interference caused by complex ions composed of two or more atoms having a similar mass-to-charge ratio to the analyte ions.
4 Principle
The test solution is introduced by a carry gas (argon) into the nebulization system and nebulized, and then enters the plasma central region as an aerosol. In the high temperature and inert atmosphere, it is desolvated, vaporized, dissociated, and ionized, transforming into positively charged ions. These ions are then collected by an ion acquisition system and sent to a mass spectrometer. The mass spectrometer separates the ions based on their mass-to-charge ratio, and the ion signals are received by an electron multiplier, amplified, and detected. Qualitative or quantitative analysis of each element is performed based on the intensity of its mass spectrometric peaks.
5 General provisions
5.1 Reagents. Unless otherwise specified, all reagents refer to guaranteed reagents.
Hydrochloric acid and nitric acid shall be purified by sub-boiling distillation if necessary. The sub-boiling distillation acid purification apparatus is shown in Appendix A. Commonly used inorganic acids include nitric acid (HNO3), hydrochloric acid (HCl), hydrofluoric acid (HF), perchloric acid (HClO4), and sulfuric acid (H2SO4).
5.2 Water. The conductivity (25 °C) shall not exceed 0.0055 mS/m.
5.3 Impurity standard solutions. It shall be prepared using certified reference materials.
6 Apparatus
6.1 Inductively coupled plasma mass spectrometer. It is composed of a sample injection system, cooling system, vacuum system, ion source, interface, ion lens system, mass separator, detector, control and data processing system, etc., and can be equipped with collision/reaction cells, coupling equipment, etc.
6.2 Containers. The containers used for digestion should be selected based on whether the background value of the target element to be tested is within the allowable error range of the analysis. Generally, polytetrafluoroethylene (PTFE), fluoroplastic (PFA), or quartz vessels are used. Containers for holding the sample solution are generally made of polytetrafluoroethylene (PTFE) or fluoroplastic (PFA), or sample vials or quartz sample vials made of tetrafluoroethylene with perfluorinated alkyl side chains, low-density polyethylene (LDPE), high-density polyethylene (HDPE), or polypropylene (PP). Before use, the containers shall be soaked in a 20%~30% nitric acid solution for several hours, washed thoroughly with water, and then dried in an electric thermostatic drying oven.
7 Test procedure
7.1 Sample preparation
7.1.1 Dissolution method
For samples of easily soluble inorganic chemical products, weigh an appropriate amount of the specimen, add a certain amount of solvent (usually water or dilute acid) to dissolve it, so that the component to be tested is converted into a measurable form, and then dilute it to a certain volume of test solution for the determination of the content of the component to be tested.
7.1.2 Fusion method
For samples of insoluble inorganic chemical products, weigh an appropriate amount of the specimen, place it in a crucible, add 5~20 times the amount of alkali metal flux (such as lithium metaborate, lithium tetraborate, sodium carbonate, sodium hydroxide, sodium peroxide, corresponding potassium salts and alkali metal fluorides), and mix.
Then, heat and melt the mixture in a high-temperature furnace, generally at a temperature of 500 °C~1200 °C. Then, leach the mixture with water or acid to convert the component to be tested into a measurable form and dilute it to a certain volume of test solution for determining the content of the component to be tested.
7.1.3 Wet digestion method
Weigh an appropriate amount of inorganic chemical product specimen, add a certain amount of a mixture of different acids or mixed acids with hydrogen peroxide or other oxidants, heat and digest to convert the component to be tested into a measurable form, and dilute it to a certain volume of test solution for the determination of the content of the component to be tested.
Commonly used digestion reagents for wet digestion include HNO3, HNO3-HClO4, HNO3-H2SO4, H2SO4-KMnO4, H2SO4-H2O2, HNO3-H2SO4-HClO4, HNO3-H2SO4- V2O5, etc.
Wet digestion methods include open container digestion, sealed high-pressure digestion, and microwave digestion, as detailed below.
a) Open container digestion method. The specimen and digestion reagent are added to an open container and heated for digestion.
b) Sealed high-pressure digestion method. The specimen and digestion reagent are placed in a suitable container and digested under high temperature and high pressure.
c) Microwave digestion method. The specimen and digestion reagent are placed in a sealed container and digested using microwaves.
7.1.4 Separation and preconcentration methods
Inorganic chemical product samples are separated using methods such as solvent extraction, ion exchange, and coprecipitation/adsorption to remove potential matrix effects and interferences, achieve preconcentration, and lower the limit of quantitation.
7.2 Test solution requirements
After processing, samples of inorganic chemical products are diluted to a certain volume based on the content of the analyte and made into test solutions.
The sample weight is determined based on the mass concentration of the analyte in the specimen and the detection limit of the method; the mass concentration of the analyte in the test solution shall be at least three times the detection limit of that element. The method for determining the detection limit is given in Appendix B.
The test solutions used for measurements on the inductively coupled plasma mass spectrometer (ICP-MS) shall meet the following basic requirements.
a) The total dissolved solids (TDS) in the test solution are not greater than 0.2%.
b) The organic content in the test solution shall not be too high. Organic matter can cause severe matrix effects, and carbon particle deposition after organic […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 18 pages — is available in the English PDF.
Referenced standards
Normative references
Editions of GB/T 30903
| Edition | Title | Revision | Status |
|---|---|---|---|
| GB/T 30903-2026 | Inorganic chemicals for industrial use - Determination of impurity elements - Inductively coupled plasma mass spectrometry | current edition | Current |
| GB/T 30903-2014 | Inorganic chemicals for industrial use - Determination of impurity elements - Inductively coupled plasma mass spectrometry | previous edition | In force until 1 October 2026 |
This page sells the current edition, GB/T 30903-2026. Earlier editions are listed for reference only.
How to Download GB/T 30903-2026
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 14666-2025 — Terms for analytical chemistry
GB/T 6379.2-2004 — Measurement methods and results -- Accuracy (trueness and precision) -- Part 2: Determine the standard methods of measurement repeatability and reproducibility of the basic method
GB/T 6379.2-2026 — Accuracy (trueness and precision) of measurement methods and results - Part 2: Basic method for the determination of repeatability and reproducibility of a standard measurement method
Secure payment via Stripe
Payments accepted
GB/T 30903-2026
$255.00