Scope

GB/T 27757-2011 is the English-translated version of pH值测定用参比电极.

GB/T 27757-2011 is the Chinese standard "Reference electrodes for the measurement of pH value". Its scope clause reads: This standard specifies the pH value determined by the reference electrode was classified, requirements, test methods, inspection rules, marking, packaging, transportation, storage and so on.

This standard applies to mercury/calomel, silver/silver chloride electrode material within the general type of saturated potassium chloride reference electrode, the other reference electrode Pole for implementation (hereinafter referred to as electrode). Its clauses include requirements.

Found what you need?

Document preview — GB/T 27757-2011

National Standard of the People's Republic of China

ICS
71.040.01
Classification
N 53

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • 1 Scope
  • 2 Normative references
  • 4 Requirements
  • 4.1 Electrode normal operating conditions
  • 4.2 electrode potential bias
  • 4.3 electrode stability

1 Scope

This standard specifies the pH value determined by the reference electrode was classified, requirements, test methods, inspection rules, marking, packaging, transportation, storage and so on.

This standard applies to mercury/calomel, silver/silver chloride electrode material within the general type of saturated potassium chloride reference electrode, the other reference electrode

Pole for implementation (hereinafter referred to as electrode).

2 Normative references

The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein

Member. For undated references, the latest edition (including any amendments) applies to this document.

GB/T 191-2008 Packaging - Pictorial signs (ISO 780.1997, MOD)

GB/T 2828.1-2003 Sampling procedures for inspection - Part 1. by acceptance quality limit (AQL) retrieval batch test sample

Programme (ISO 2859-1.1999, IDT)

GB/T 2829-2002 periodic inspection Sampling procedures and tables (suitable for process stability test)

GB/T 11606-2007 analytical instruments environmental test methods

Category 3

Divided by the use of premises.

a) Laboratory;

b) line type.

4 Requirements

4.1 Electrode normal operating conditions

Electrode should be able to work under the following conditions.

a) Ambient temperature. 5 °C ~ 40 °C;

b) Relative humidity. <=90%;

c) the measured solution temperature. 0 °C ~ 70 °C;

d) The junction is no serious corrosion and contamination of water.

4.2 electrode potential bias

Electrode potential deviation shall not exceed ± 3mV.

4.3 electrode stability

Stability of the electrode should meet the following requirements.

a) Laboratory. ± 2mV/7h;

b) On-Line. ± 3mV/24h.

GB/T 17360-2020 | GB/T 25929-2010 | GB/T 25930-2010 | GB/T 21187-2026 | GB/T 21186-2026 | GB/T 27756-2011 | […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Similar standards

How to Download GB/T 27757-2011

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Fast delivery

    After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Fast delivery
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 27757-2011

$135.00