GB/T 27503-2011Test solutions of electrolytic conductivity analyzer - Preparation method of sodium chloride solutions (English PDF)
电导率仪的试验溶液 氯化钠溶液制备方法
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for GB/T 27503-2011 in 30 minutes.
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
October 31, 2011
Implementation date
January 1, 2012
Scope
GB/T 27503-2011 is the English-translated version of 电导率仪的试验溶液 氯化钠溶液制备方法.
GB/T 27503-2011 gives the method for preparing the sodium chloride test solutions of the electrolytic conductivity analyzer.
Clause 1 states that the standard specifies the reagents, the instruments or equipment and the preparation steps used to prepare the sodium chloride solution, together with the correspondence between the concentration of the solution and its resistivity, and that it applies to the preparation of the sodium chloride solution, called the solution below, required for testing a conductivity meter.
Clause 2, "Reagents", calls for analytical reagent sodium chloride and for distilled or deionized water of conductivity not greater than 0.2 x 10^-6 S/cm.
Clause 3, on the instruments or apparatus required, calls for a balance of maximum weighing not more than 200 g with a verification scale value of 0.1 mg and an electric oven covering 0 to 300 degrees Celsius.
Clause 4, on the preparation steps, sets the environmental conditions as an ambient temperature of (20 plus or minus 2) degrees Celsius and a relative humidity below 85%, and has the sodium chloride dried in the electric oven at a controlled temperature of 220 to 240 degrees Celsius for at least 2 h, cooled to room temperature, then weighed and added to a certain mass of water and mixed thoroughly to obtain the solution of the required mass fraction, with an uncertainty not exceeding 0.25%.
Clause 5, on the resistivity and conductivity of the solution, gives in Table 1 the resistivity of solutions of different concentrations at 18 degrees Celsius, in Table 2 the temperature coefficient of the solution over the temperature range of 0 to 140 degrees Celsius and in Table 3 correction values of that temperature coefficient in the low concentration range, and gives formula (1) for calculating the resistivity of the solution at other temperatures.
The draft on the page breaks off inside clause 5, and the page lists no normative references and no normative references clause for this standard.
The foreword states that the standard was drafted according to the rules given in GB/T 1.1-2009, was proposed by the China Machinery Industry Federation, is centrally managed by SAC/TC124/SC6 of the national industrial process measurement and control instruments standardization technical committee, and was drafted by Shanghai Precision Scientific Instrument Co., Ltd., the Shanghai Measurement and Testing Technology Research Institute and East China Normal University among others. It was issued on 2011-10-31 and has applied since 2012-01-01.
Found what you need?
Document preview — GB/T 27503-2011
National Standard of the People's Republic of China
- ICS
- 71.040.01
- Classification
- N 53
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword
- 1 Scope
- 2 Reagents
- 3 Instruments or apparatus Preparation of the solution required for
- 5 Solution resistivity and conductivity
Foreword
This standard was drafted in accordance with GB/T 1.1-2009 given rules. The standard proposed by China Machinery Industry Federation.
The standard measurement and control of the instruments Standardization Technical Committee Technical Committee (SAC/TC124/SC6) by the National Industrial Process Centralized.
1 Scope
This standard specifies the correspondence between the sodium chloride solution was prepared using reagents, instruments or equipment, preparation steps and solution concentration and resistivity.
Preparation of sodium chloride solution (hereinafter referred to as a solution) is required for this standard applies to the conductivity meter test.
2 Reagents
Preparation of the solution required reagents.
a) Sodium chloride. AR;
b) Water. water or a laboratory conductivity no greater than 0.2 x 10^-6S/cm of distilled or deionized water.
3 Instruments or apparatus
Preparation of the solution required for instruments or apparatus.
a) balance. the largest weighing not more than 200g, verification scale value of 0.1mg;
b) electric oven. (0 ~ 300) °C.
4 Preparation Step
4.1 Environmental conditions
Solution should be prepared in the following environmental conditions.
a) Ambient temperature. (20 ± 2) °C;
b) Relative humidity. less than 85%.
4.2 Solution Preparation
After the sodium chloride placed in an electric oven, the temperature control (220 ~ 240) °C range, dried for at least 2h, cooled to room temperature, depending on the count Count, weigh a certain amount of sodium chloride was added to a certain quality of water and thoroughly mixed to obtain the desired mass fraction of the solution, the uncertainty does not exceed More than 0.25%.
5 Solution resistivity and conductivity
5.1 Solution resistivity
Table 1 shows the different concentrations of the solution 18 °C when the resistivity;
Table 2 (0 to 140) °C temperature coefficient within the temperature range of the solution [alpha];
Table 3 is a correction value in the low concentration range of 5% solution of the temperature coefficient alpha;
The resistivity of the solution by the formula (1) calculation of the other temperatures. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Similar standards
How to Download GB/T 27503-2011
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.