GB/T 26124-2011In vitro diagnostic reagent(kit) for clinical chemistry (English PDF)
临床化学体外诊断试剂(盒)
This is a limited preview
Buy now to download the full PDF (23 pages)
Need quotation for your management?
Get a formal quotation for GB/T 26124-2011 in 30 minutes.
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
May 12, 2011
Implementation date
November 1, 2011
Scope
GB/T 26124-2011 is the English-translated version of 临床化学体外诊断试剂(盒).
GB/T 26124-2011 is the Chinese standard "In Vitro Diagnostic Reagent (kit) for Clinical Chemistry".
Its scope clause reads: This Standard specifies the general technical requirements for the quality inspection of in vitro diagnostic reagent (kit) for clinical chemistry [hereinafter referred to as "reagent (kit)"], including terms and definitions, classification and naming, requirements, test methods, labels and instructions for use, packaging, transportation and storage.
This Standard is applicable to in vitro diagnostic reagents (kits) based on the principle of spectrophotometry used in the quantitative inspection of clinical chemistry items in medical laboratories.
This Standard does not apply to. a) Performance evaluation reagents (for example, reagents for research purposes only); b) In vitro diagnostic reagents for clinical chemistry in POCT (Point of Care Testing).
Its clauses include terms and definitions; naming and classification; requirements; test methods; markings, labels and instructions for use; packaging, transportation and storage.
It was issued by the General Administration of Quality Supervision, Inspection and Quarantine on MAY 12, 2011 and took effect on NOVEMBER 1, 2011.
Found what you need?
Document preview — GB/T 26124-2011
National Standard of the People's Republic of China
- ICS
- 11.100
- Classification
- C 44
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword3
- 1 Scope4
- 2 Normative References4
- 3 Terms and Definitions4
- 4 Naming and Classification9
- 5 Requirements9
- 6 Test Methods11
- 7 Markings, Labels and Instructions for Use15
- 8 Packaging, Transportation and Storage22
- Bibliography23
Foreword
This Standard was drafted in accordance with the rules in GB/T 1.1-2009 Directives for Standardization - Part 1: Structure and Drafting of Standards.
Please be noted that certain content of this document might involve patents. The institution issuing this document does not undertake the responsibility of identifying these patents.
1 Scope
This Standard specifies the general technical requirements for the quality inspection of in vitro diagnostic reagent (kit) for clinical chemistry [hereinafter referred to as "reagent (kit)"], including terms and definitions, classification and naming, requirements, test methods, labels and instructions for use, packaging, transportation and storage.
This Standard is applicable to in vitro diagnostic reagents (kits) based on the principle of spectrophotometry used in the quantitative inspection of clinical chemistry items in medical laboratories. This Standard does not apply to.
a) Performance evaluation reagents (for example, reagents for research purposes only);
b) In vitro diagnostic reagents for clinical chemistry in POCT (Point of Care Testing).
2 Normative References
The following documents are indispensable to the application of this document. In terms of references with a specified date, only versions with a specified date are applicable to this document. In terms of references without a specified date, the latest version (including all the modifications) is applicable to this document.
GB 3100 The International System of Units and Its Application
YY 0466-2003 Medical Devices - Symbols to be Used with Medical Device Labels, Labelling and Information to be Supplied
YY/T 0638 In Vitro Diagnostic Medical Devices - Measurement of Quantities in Biological Samples - Metrological Traceability of Assigned Values for Catalytic Concentration of Enzymes in Calibrators and Control Materials
3 Terms and Definitions
The following terms and definitions are applicable to this Standard.
3.1 In Vitro Diagnostic Reagent
In vitro diagnostic reagent refers to the chemical, biological or immunological components, solutions or preparations that are expected to be used by the manufacturer as an in vitro diagnostic medical device.
3.2 Clinical Chemistry Reagent
Clinical chemistry reagent refers to an in vitro diagnostic reagent based on the principle of spectrophotometry used in the quantitative inspection of clinical chemistry items in medical laboratories.
3.3 Kit
Kit refers to a set of components packaged together to complete a specified in vitro diagnostic test.
3.4 Accuracy
Accuracy refers to the degree of consistency between a measured value and an acceptable reference value.
3.5 Reference Material
Reference material refers to a material or substance that has one or multiple sufficiently uniform and well-determined characteristics to calibrate measurement devices, evaluate measurement methods or assign values to materials.
3.6 Certified Reference Material; CRM
Certified reference material refers to a reference material with an attached certificate; one or multiple characteristic values are determined by an established traceability procedure, so that it can be traced to an accurate reproducible measurement unit representing the characteristic value. Each certified characteristic value is accompanied by the uncertainty of a given confidence level.
4 Naming and Classification
4.1 Naming
The product name of in vitro diagnostic reagents (kits) for clinical chemistry shall include the Chinese name of the corresponding test item, and if necessary, the English (abbreviated) name may be added.
4.2 Classification
The composition, specifications and detection principle of the reagents (kits) shall be described.
5 Requirements
5.1 Appearance
The appearance shall comply with the normal appearance requirements specified by the manufacturer.
5.2 Net Weight
The net weight of liquid reagents shall be not less than the indicated value.
5.3 Reagent Blank
5.4 Analytical Sensitivity
When the reagents (kits) are used to test n units of tested object, the difference in absorbance ( A) or the change rate of absorbance ( A/min) shall comply with the range provided by the manufacturer.
5.5 Linearity Range
The analytical performance within the linearity range of the reagents (kits) shall comply with the following requirements. […]
5.6 Precision of Measurement
5.7 Accuracy
The accuracy of the reagents (kits) shall comply with the requirements specified by the manufacturer.
5.8 Stability
The following methods may be selected for the verification of stability.
6 Test Methods
6.1 Basic Requirements for Instruments
Spectrophotometer or biochemical analyzer. the wavelength range covers 340 nm to 700 nm. The biochemical analyzer shall be equipped with a constant-temperature device (precision ± 0.1 °C). The precision of absorbance measurement is above 0.001.
6.4 Reagent Blank
6.4.1 Use the designated blank samples to test the reagents (kits). At the main wavelength of the test, record the absorbance (A1) at the start of the test and the absorbance (A2) after about 5 min (t). The test result of A2 is the measured value of the […]
6.5 Analytical Sensitivity
Use samples with already-known concentration or activity to test the reagents (kits); record the absorbance change generated under the specified parameters of the reagents (kits). Convert it to the difference in absorbance ( A) of n units, or, the change rate of absorbance ( A/min) of n units, which shall comply with the requirements of 5.4.
6.6 Linearity Range
6.7 Precision of Measurement
6.7.1 Repeatability
Under repeatability conditions, use the control materials to test the reagents (kits); repeat the test for at least 10 times (n >= 10). Respectively calculate the average value ( x) and standard deviation (s) of the measured values. In accordance with Formula (1), calculate the coefficient of variation (CV), which shall comply with the requirements of 5.6.1.
6.7.2 Within-run difference between bottles
Use the control materials to respectively test 10 bottles with the same batch number.
In addition, calculate the average value ( x1) and standard deviation (s1) of the 10 measured values.
6.7.3 Between-run difference
Use the control materials to respectively test reagents (kits) with 3 different batch numbers. For each batch number, test 3 times. Respectively calculate the average value xi (i = 1, 2, 3) of the 3 tests of each batch. In accordance with Formula (4) and Formula (5), calculate the relative deviation (R), which shall comply with the requirements of 5.6.3.
6.9 Stability
The following methods may be selected for the verification of stability.
7 Markings, Labels and Instructions for Use
7.1 General Requirements
The markings, labels and instructions for use shall comply with the following requirements. […]
7.2 Markings and Labels of Outer Container
The markings and labels of the outer container shall include the following information.
7.3 Markings and Labels of Immediate Container
If the immediate container is also the outer container, then, the requirements of 7.2 also apply.
7.4 Instructions for Use
The instructions for use shall provide the following information.
8 Packaging, Transportation and Storage
8.1 Packaging
The packaging shall comply with the following requirements. […]
8.2 Transportation
Transport in accordance with the conditions specified in the contract.
8.3 Storage
Store in accordance with the specified conditions. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 23 pages — is available in the English PDF.
Referenced standards
Normative references
- GB 3100SI units and recommendations for the use of their multiples and of certain other units
- YY 0466-2003Medical devices -- Symbols to be used with medical device labels labelling and information to be supplied
- YY/T 0638In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values for catalytic concentration of enzymes assigned to calibrators and control materials
Similar standards
How to Download GB/T 26124-2011
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 3100-1993 — SI units and recommendations for the use of their multiples and of certain other units
YY 0466-2003 — Medical devices -- Symbols to be used with medical device labels labelling and information to be supplied
YY/T 0638-2008 — In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values for catalytic concentration of enzymes assigned to calibrators and control materials