GB/T 23362.3-2009Methods for chemical analysis of high purity indium hydroxide - Part 3: Determination of antimony content - Atomic fluorescence spectrometry (English PDF)
高纯氢氧化铟化学分析方法 第3部分:锑量的测定 原子荧光光谱法
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for GB/T 23362.3-2009 in 30 minutes.
Issued by
SAMR; SAC
Level / Type
National · Recommended
Issue date
March 19, 2009
Implementation date
January 1, 2010
Scope
GB/T 23362.3-2009 is the English-translated version of 高纯氢氧化铟化学分析方法 第3部分:锑量的测定 原子荧光光谱法.
GB/T 23362.3-2009 gives the atomic fluorescence spectrometry method for the antimony content of high purity indium hydroxide. Clause 1 states that this part of GB/T 23362 specifies that determination and that the part applies to the antimony content of high purity indium hydroxide over a measuring range, as a mass fraction, of 0.00001 % to 0.0050 %.
Clause 2, method summary, has the sample dissolved with hydrochloric acid, the antimony pre-reduced to its trivalent state with a thiourea and ascorbic acid solution, then reduced to its hydride by potassium borohydride in a hydride generator and carried by argon into the quartz furnace atomizer, where the fluorescence intensity is measured on the atomic fluorescence spectrometer and the amount of antimony calculated from it.
Clause 3 lists the reagents, which are used with secondary distilled water or water of equivalent purity unless otherwise indicated, and gives the preparations for the thiourea and ascorbic acid solution, for the potassium borohydride solution at 20 g/L, for the antimony standard stock solution at 100 micrograms per millilitre and for the two standard solutions drawn from it, at 5 micrograms and at 500 nanograms per millilitre, as well as argon of a volume fraction of not less than 99.99 %.
Clause 4 requires an atomic fluorescence spectrometer fitted with a special antimony hollow cathode lamp and meeting stated specifications: a zero drift of not more than 5 %, a short-term stability of not more than 3 % RSD within 30 minutes, a detection limit of not more than 0.5 ng/mL and, for a curve over the range 0 to 60 ng/mL, a correlation coefficient of not less than 0.995.
Clause 5 has the specimen dried at 105 to 110 degrees Celsius for two hours and cooled to room temperature in a desiccator; the draft breaks off there, so the analysis steps and the calculation cannot be reported.
The foreword divides GB/T 23362 into six parts and names this one as Part 3. The page records no earlier edition. It was issued on 2009-03-19 and has applied since 2010-01-01.
Found what you need?
Document preview — GB/T 23362.3-2009
National Standard of the People's Republic of China
- ICS
- 77.120
- Classification
- H 66
Issued by: State Administration for Market Regulation; Standardization Administration of the PRC
Contents
- Foreword
- 1 Scope
- 2 Method summary
- 3 Reagents
- 4 Instrument
- 5 Sample
Foreword
GB/T 23362 "high-purity indium hydroxide chemical analysis method" is divided into six sections.
--- Part 1. Determination of arsenic content by atomic fluorescence spectrometry;
--- Part 2. Determination of tin content of phenyl spectrophotometric;
--- Part 3. Determination of antimony content - Atomic Fluorescence Spectrometry;
--- Part 4. aluminum, iron, copper, zinc, cadmium, lead and thallium content - Inductively Coupled Plasma Mass Spectrometry;
--- Part 5. Determination of chlorine content of thiocyanate mercury spectrophotometry;
--- Part 6. Determination of loss on ignition weighing method. This is Part 3.
1 Scope
This section GB/T 23362 specifies the determination method of high-purity indium hydroxide antimony content.
This section applies to the determination of high-purity indium hydroxide antimony content measurement range (mass fraction) of 0.00001% to 0.0050%.
2 Method summary
The sample is dissolved with hydrochloric acid, thiourea - ascorbic acid will be reduced to trivalent antimony advance in hydride generator, is reduced antimony potassium borohydride as hydrogen Thereof, with argon gas was introduced into the quartz furnace atomizer, the fluorescence intensity measured on the atomic fluorescence spectrometer, calculate the amount of antimony.
3 Reagents
Use only identified as gifted class pure reagents and secondary distilled water or its equivalent purity, unless otherwise indicated.
3.1 Hydrochloride (rho1.19g/mL).
3.2 Sulfate (rho1.84g/mL).
3.3 HCl (1 + 1).
3.4 thiourea - ascorbic acid solution. Weigh 25g and 25g of thiourea AR AR ascorbic acid dissolved in 500mL of water, mixed Uniform, the time is now equipped with.
3.5 solution of potassium borohydride (20g/L). Weigh 10g potassium borohydride, dissolved in 500mL of potassium hydroxide solution (5g/L), and mix well. If The precipitate is filtered and used. The time is now equipped with.
3.6 hydrochloric acid (1 + 19).
3.7 Antimony standard stock solution. Weigh 0.1000g antimony (mass fraction >=99.99%), placed in 100mL beaker. Sulfur was added 5mL Acid, dissolved by heating, cooled and transferred to 1000mL volumetric flask, diluted with hydrochloric acid (1 + 4) to the mark and mix. This solution containing 1mL 100µg antimony.
Antimony 3.8 Standard Solution A. Pipette 10.00mL antimony standard stock solution (3.7), placed in 200mL volumetric flask, diluted with hydrochloric acid (1 + 4) to Volume, and mix. This solution 1mL containing 5µg antimony.
3.9 Antimony standard solution B. Pipette 10.00mL antimony standard solution A (3.8), placed in 100mL flask, diluted with hydrochloric acid (1 + 4) to Volume, and mix. This solution 1mL containing 500ng antimony.
3.10 Argon (volume fraction >=99.99%).
4 Instrument
Atomic fluorescence spectrometer, equipped with antimony special hollow cathode lamp. The atomic fluorescence spectrometer should meet the following specifications.
--- Stability. Zero drift <=5%, short-term stability RSD<=3% 30min inside;
--- The detection limit <=0.5ng/mL;
--- Linear curve. curve in 0 ~ 60ng/mL range, the correlation coefficient should >=0.995.
5 Sample
The specimen should be 105 °C ~ 110 °C drying 2h, placed in a desiccator to cool to room temperature.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Similar standards
How to Download GB/T 23362.3-2009
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.