GB/T 14571.2-2018Test method of ethylene glycol for industrial use—Part 2: Determination of purity and impurities—Gas chromatography (English PDF)

工业用乙二醇试验方法 第2部分:纯度和杂质的测定 气相色谱法

Preview PDF

This is a limited preview

Buy now to download the full PDF (15 pages)

Need quotation for your management?

Get a formal quotation for GB/T 14571.2-2018 in 30 minutes.

Issued by

SAMR; SAC

Level / Type

National · Recommended

Issue date

March 15, 2018

Implementation date

October 1, 2018

Scope

GB/T 14571.2-2018 is the English-translated version of 工业用乙二醇试验方法 第2部分:纯度和杂质的测定 气相色谱法.

GB/T 14571.2-2018 is the Chinese national gas chromatography method for measuring the purity of ethylene glycol for industrial use and the impurities it contains.

It applies to samples with a purity of at least 98.0% by mass and covers 1,2-propanediol, diethylene glycol, triethylene glycol, 1,2-butanediol, 1,4-butanediol, 1,2-hexanediol, ethylene carbonate and 1,3-dioxane-2-methanol, with detection limits of 0.0010% or 0.0020% by mass.

The method uses a chromatograph with a flame ionization detector, calibration solutions prepared by weighing and correction factors applied through corrected area normalization.

It also sets out the reagents and gases, sampling, test steps, repeatability and reproducibility, quality control samples and the test report. It replaces GB/T 14571.2-1993, was issued on 15 March 2018 and has applied since 1 October 2018.

Found what you need?

Document preview — GB/T 14571.2-2018

National Standard of the People's Republic of China

ICS
71.080.60
Classification
G 16
Replacing
GB/T 14571.2-1993

Issued by: State Administration for Market Regulation; Standardization Administration of the PRC

Contents

  • Foreword3
  • 1 Scope5
  • 2 Normative references5
  • 3 Principle6
  • 4 Reagents or materials6
  • 5 Instruments and devices7
  • 6 Sample7
  • 7 Test steps8
  • 8 Processing of test data10
  • 9 Precision11
  • 10 Quality assurance and control12
  • 11 Test report12

Foreword

GB/T 14571 consists of the following parts.

GB/T 14571.1, Test method of ethylene glycol for industrial use - Part 1. Determination of acidity - Titration method

GB/T 14571.2, Test method of ethylene glycol for industrial use - Part 2. Determination of purity and impurities - Gas chromatography

GB/T 14571.3, Ethylene glycol for industrial use - Determination of content of total aldehydes present - Spectrophotometric method

GB/T 14571.4, Ethylene glycol for industrial use - Determination of ultraviolet transmittance - Ultraviolet spectrophotometric method

GB/T 14571.5, Test method of ethylene glycol for industrial use - Part 5. Determination of chloride ion - Ion chromatography

This Part is Part 2 of GB/T 14571.

This Part was drafted in accordance with the rules given in GB/T 1.1-2009. This Part replaces GB/T 14571.2-1993 "Ethylene glycol for industrial use. Determination of diethylene glycol and triethylene glycol. Gas chromatographic method".

Compared with GB/T 14571.2-1993, main changes in this Part are as follows.

- modified standard's name;

- modified tiles of related clauses (see Clause 3~Clause 11 of this Edition, Clause 3~Clause 9 of Edition 1993);

- modified scope (see Clause 1 of this Edition, Clause 1 of Edition 1993);

- added relevant standards to normative references (see Clause 2 of this Edition, Clause 2 of Edition 1993);

- modified principle (see Clause 3 of this Edition, Clause 3 of Edition 1993);

- deleted packing column; deleted capillary column type and chromatographic conditions (see Table 1 of this Edition, Table 1 of Edition 1993);

- added calculation of ethylene glycol purity (see Clause 8 of this Edition);

- added contents related to determinations of 1,2-propanediol, 1,2-butanediol, 1,4-butanediol, 1,2-hexanediol, ethylene carbonate, and 1,3-dioxane-2-methanol (see Clause 1, Clause 7 and 4.5.2 of this Edition);

- modified quantification method from external standard method and internal standard method to corrected area normalization method (see Clause 8 of this Edition, 7.3 and Annex A of Edition 1993);

- modified precision data of method (see Table 2 of this Edition, Table 2 of Edition 1993);

- added quality assurance and control (see Clause 10 of this Edition);

- deleted Annex A (See Annex A of Edition 1993).

1 Scope

This Part of GB/T 14571 specifies principle, reagents or materials, instruments and devices, samples, test procedures, test data processing, precision, quality assurance and control, test report of gas chromatography for determination of purity and impurities in ethylene glycol for industrial use.

This Part is applicable to determine the ethylene glycol sample for industrial use of which purity is not less than 98.0% (mass fraction). Where, the detection limit of 1,2-propanediol and triethylene glycol is 0.0020% (mass fraction). The detection limit of 1,3-dioxane-2-methanol, diethylene glycol, 1,2-butanediol, 1,4-butanediol, 1,2-hexanediol and ethylene carbonate is 0.0010% (mass fraction).

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

GB/T 3723, Sampling of chemical products for industrial use - Safety in sampling

GB/T 6678, General Principles for Sampling Chemical Products

GB/T 6680, General rules for sampling liquid chemical products

3 Principle

Under specified conditions, inject an appropriate amount of sample into the chromatograph that is equipped with hydrogen flame ionization detector (FID).

4 Reagents or materials

WARNING - Gases in 4.1~4.4 are high-pressure compressed gas or pressurized flammable gas. Most of standard reagents in 4.5 are flammable or toxic liquids. Pay attention to safety in use.

4.5 Reagents.

4.1 Carrier gas

Helium or nitrogen; purity is not less than 99.99% (volume fraction); dried and purified by silica gel and 5A molecular sieve.

4.4 Auxiliary gas

Nitrogen; purity is not less than 99.99% (volume fraction); dried and purified by silica gel and 5A molecular sieve.

5 Instruments and devices

5.1 Gas chromatograph. equipped with hydrogen flame ionization detector. The peak height of the lowest measured concentration of impurities specified in this Part shall be at least twice the noise. Dynamic linear range meets quantitative requirements.

5.3 Analytical balance. resolution is 0.1mg.

5.4 Sample introduction device. 10µL micro syringe or liquid autosampler.

5.5 Recording device. electronic integrator or chromatography workstation.

6 Sample

Perform sampling according to the provisions of GB/T 3723, GB/T 6678, GB/T 6680.

7 Test steps

7.1 Instrument preparation

According to the operation instructions of the instrument, install and age the chromatographic column in chromatograph.

7.2 Preparation of calibration solution

Use weighing method to prepare the calibration solution that contains high purity ethylene glycol (4.5.1) and impurities to be tested (4.5.2). Each component shall be accurately weighed to 0.0001g.

7.3 Determination of correction factor

Under the chromatographic conditions recommended in Table 1, take an appropriate amount of calibration solution (7.2) into chromatograph. Repeat the determination 3 times. See Figure 1 for typical chromatogram. Measure all chromatographic peak areas. If 1,2-propanediol and ethylene glycol are not separated from the baseline, the chromatographic peak of 1,2-propanediol shall be chamfered according to the tailing peak.

7.4 Determination of sample

Under the chromatographic conditions recommended in Table 1, take an appropriate amount of sample to be tested into chromatograph.

8 Processing of test data

8.1 Calculation of correction factor

Calculate the correction factor (fi) of each impurity relative to ethylene glycol according to formula (1).

8.3 Expression of analysis results

For any sample, the content of each component is expressed as the arithmetic mean of the results of two parallel determinations.

9 Precision

9.1 Repeatability

In the same laboratory, the same operator uses the same device. According to the same test method, test the same test object independently of each other in a short time. The absolute difference between the two independent test results obtained is not greater than the repeatability limit (r) listed in Table 2. The premise is that the case greater than the repeatability limit (r) does not exceed 5%.

9.2 Reproducibility

In any two laboratories, the different operator uses the different device. According to the same test method, test the same test object independently.

10 Quality assurance and control

10.1 The laboratory shall periodically analyze the quality control samples to ensure the accuracy of the test results.

10.2 The quality control sample shall be stable and representative of the sample being analyzed. The quality control sample can use the calibration solution that is self-prepared according to 7.2 or the certified standard solution sold at market.

11 Test report

The report shall contain the following information. […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Referenced standards

Similar standards

Editions of GB/T 14571.2

EditionTitleRevisionStatus
GB/T 14571.2-2018Test method of ethylene glycol for industrial use—Part 2: Determination of purity and impurities—Gas chromatographycurrent editionCurrent
GB/T 14571.2-1993Test method of ethylene glycol for industrial use—Part 2: Determination of purity and impurities—Gas chromatographyprevious editionSuperseded

This page sells the current edition, GB/T 14571.2-2018. Earlier editions are listed for reference only.

How to Download GB/T 14571.2-2018

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

GB/T 14571.2-2018

$125.00