GB 16361-2012Specification for patient radiological protection and quality control in nuclear medicine (English PDF)
临床核医学患者防护与质量控制规范
Open the GB 16361-2012 preview as PDF
This is a limited preview
Buy now to download the full PDF (43 pages)
Issued by
Ministry of Health of the People's Republic of China; Standardization Administration of China
Level / Type
National · Mandatory
Issue date
June 29, 2012
Implementation date
October 1, 2012
Scope
GB 16361-2012 is the English-translated version of 临床核医学患者防护与质量控制规范.
GB 16361 covers the branch of medicine in which the radiation source is given to the patient rather than aimed at them: clinical nuclear medicine, diagnostic imaging and radionuclide therapy alike, excluding the implantation of sealed seed sources. Clause 4 sets what a nuclear medicine unit must have before it may practise - the rooms and their shielding under GBZ 120, verified activity meters and quality control instruments, a radiological protection and quality assurance organisation, contingency plans against a mistaken administration - and it reserves the prescribing of a therapeutic radiopharmaceutical to a qualified practitioner who has made the justification assessment case by case. Clause 5 is the longest: justification of each administration, the guidance levels and the smallest activity that will still answer the diagnostic question, and the specific rules for children, for women who are breastfeeding and for pregnant women, including the 100 mGy threshold below which a life-saving treatment need not end the pregnancy. Clause 6 adds the quality control of the equipment and of the procedures. Seven annexes carry the working numbers: the guidance levels for diagnosis (A), how long breastfeeding is to be interrupted (B), how long conception is to be deferred after treatment (C), the dose the patient receives (D, some twenty pages of it), the dose to the people around them (E, normative), and the performance and quality control testing of the activity meter (G). Issued on 29 June 2012, in force since 1 October 2012, replacing GB 16361-1996. This page is published from the official record of the standard and from its published table of contents. The full clause text, the tables and the numeric limits are in the document itself, which is delivered complete in English translation.
Document preview — GB 16361-2012
National Standard of the People's Republic of China
- ICS
- 13.280
- Classification
- C 57
- Replacing
- GB 16361-1996
Issued by: Ministry of Health of the People's Republic of China; Standardization Administration of China
Contents
- 1 Scope
- 2 Normative references
- 3 Terms and Definitions
- 4 Basic requirements
- 4.1 Clinical nuclear medicine diagnosis and treatment units
- 5.1 Justification judgment
- 5.2 Radiological Protection Optimization
- 5.2.1 General requirements
- 5.3 Guidance Levels for Medical Exposure 5.3.1
Foreword
The standard Chapter 4, Chapter 5 are mandatory, the rest are recommended. This standard was drafted in accordance with GB/T 1.1-2009 given rules. This standard replaces GB 16361-1996 "Clinical Nuclear Medicine Radiological protection standards for the patient." This standard compared with GB 16361-1996, the main technical changes are as follows:
--- Original standard is mainly based on national standards GB 4792-1984 "Radiation Health Protection basic standards"; this standard is based primarily on It has been replaced by a new version of the basic standards 4792-1984 GB of GB 18871-2002 "ionizing radiation protection and safety of radiation sources fundamental Standard "and GB Z179-2006" medical exposure Radiological Protection basic requirements. "
--- International standard original standard is mainly based on ICRP Publication 52 "ProtectionofthePatientinNuclear Medicine, Ann.ICRP17 (4), 1987 "; this standard is mainly based on the IAEA Safety Report No.40" ApplyingRa- diationSafetyStandardsinNuclearMedicine, 2005 ", as well as IAEA Safety Report No.63" Releaseof PatientsafterRadionuclideTherapy, 2009 "and the IAEA Technical Report No.454" QualityAssurancefor RadioactivityMeasurementinNuclearMedicine, 2006 ".
--- Standard name from "clinical nuclear medicine patient radiation health protection standards" by "protection of patients in clinical nuclear medicine and quality control System Specifications. "
--- Original standard Chapter 3 ~ Chapter 7 of the content integration, grouped into current standard Chapters 4 and 5, and there are more complementary and modify.
--- Increased in Chapter 6 concerning quality control requirements.
--- By the Appendix 2 to 8, and the corresponding references are given in the appendix. This standard is proposed and administered by the People's Republic of China Ministry of Health. This standard was drafted. Institute of Radiation Medicine, Chinese Academy of Medical Sciences, Sichuan Provincial Center for Disease Control and Prevention. The main drafters. Zhang Liangan, Zhang literary, Jiao Ling, Dingyan Qiu, He Ling, Yang Yi. This standard replaces the standards previously issued as follows:
--- GB 16361-1996. Protection and Quality Control of Nuclear Medicine clinical Specification
1 Scope
GB 16361 covers the branch of medicine in which the radiation source is given to the patient rather than aimed at them: clinical nuclear medicine, diagnostic imaging and radionuclide therapy alike, excluding the implantation of sealed seed sources. Clause 4 sets what a nuclear medicine unit must have before it may practise - the rooms and their shielding under GBZ 120, verified activity meters and quality control instruments, a radiological protection and quality assurance organisation, contingency plans against a mistaken administration - and it reserves the prescribing of a therapeutic radiopharmaceutical to a qualified practitioner who has made the justification assessment case by case. Clause 5 is the longest: justification of each administration, the guidance levels and the smallest activity that will still answer the diagnostic question, and the specific rules for children, for women who are breastfeeding and for pregnant women, including the 100 mGy threshold below which a life-saving treatment need not end the pregnancy. Clause 6 adds the quality control of the equipment and of the procedures. Seven annexes carry the working numbers: the guidance levels for diagnosis (A), how long breastfeeding is to be interrupted (B), how long conception is to be deferred after treatment (C), the dose the patient receives (D, some twenty pages of it), the dose to the people around them (E, normative), and the performance and quality control testing of the activity meter (G). Issued on 29 June 2012, in force since 1 October 2012, replacing GB 16361-1996. This page is published from the official record of the standard and from its published table of contents. The full clause text, the tables and the numeric limits are in the document itself, which is delivered complete in English translation.
This standard specifies the basic requirements for protection of clinical nuclear medicine patients, patient protection and safety, and quality control requirements. This standard applies to the radionuclides used for diagnosis and treatment of clinical nuclear medicine, but does not include the case of a particle source implantation.
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein Member. For undated references, the latest edition (including any amendments) applies to this document.
GB 18871-2002 ionizing radiation protection and safety of radiation sources basic standards GB Z120 Nuclear Medicine Radiological protection standards GB Z179 basic requirements for medical exposure Radiological Protection
3 Terms and Definitions
The following terms and definitions apply to this document.
3.1 Radiopharmaceuticals radiopharmaceutical For the purpose of diagnosis and treatment may be applied to the human body any compound containing radionuclides. In addition to the radioactive properties, chemical and biological On consideration, radioactive drugs and other drugs.
3.2 Radionuclide purity radionuclidicpurity Specific radioactivity short-lived radionuclides and daughters accounted for radioactivity total activity score.
3.3 Radiochemical purity radiochemicalpurity Radionuclides in the desired chemical form of the mass fraction.
Note. If you can not reach the radiochemical purity requirements may lead to target specific radiopharmaceutical tissue deterioration, resulting in radiation on non-target organs.
3.4 Chemical purity chemicalpurity Regardless of the presence of radioactive or not, according to the desired chemical form of the mass fraction. In general, radiopharmaceuticals, there are only of Learn toxicity or physiological process radiopharmaceutical impurities proposed chemical purity requirements.
3.5 Radioactivity specificradioactivity Radiopharmaceutical unit mass contained radionuclide activity and activity concentration of bound and free form.
3.6 Radionuclide generators radionuclidegenerator The device can be isolated from the decay of its long half-life radionuclide-containing preparations (known as the parent) of the short half-life nuclides (called daughters) generated by the vulgar Called "cow."
4.1 Clinical nuclear medicine diagnosis and treatment units
4.1.1 clinical nuclear medicine diagnosis and treatment unit (hereinafter referred to as nuclear medicine units) should follow the "radiation medical management regulations" for clinical Practicing nuclear medicine diagnosis and treatment conditions (hereinafter referred to as clinics) provided personnel and equipment, and other related requirements, and GB 18871- 2002 GB Z179 and other related technical standards.
4.1.2 should be equipped to adapt its services and the performance of qualified nuclear medicine diagnostic equipment (including related auxiliary equipment), and protection of patients Quality control testing equipment, personal protective equipment.
4.1.3 in accordance with relevant regulations, regularly test equipment for verification or calibration, to obtain competent and effective verification or calibration certificate. When it is detected Instrument after major repairs, you should re-test or calibration.
4.1.4 should be set to adapt its services and meet the protection requirements of GB Z120 laboratory examination room, injection rooms, treatment ward And waiting areas and other workplaces and their corresponding protection facilities.
4.1.5 It should be noted configured to adapt its services and reasonable structure of various professionals, including enforcement of the Nuclear Medicine have qualified Industry, physicians and clinical nuclear medicine technician, etc., and they are active in patient protection and safety and quality control knowledge, training and assessment.
4.1.6 There shall be established radiological protection and quality assurance management organizations, various medical technicians in their work should strictly follow the protection and safety to Demand and quality control requirements, and undertake corresponding responsibilities.
4.1.7 should develop the Nuclear Medicine Clinic quality assurance program, including strengthening the establishment of a sound management system, including the protection of patients, from the management system Ensure the correct implementation of the Nuclear Medicine Clinic of the degree and quality control procedures.
4.1.8 should be possible for the implementation of treatment failures or mistakes, development of contingency plans, and training and emergency drills will likely The consequences of failure or error is minimized.
4.2 Requirements for clinical nuclear medicine practitioners and associated personnel
4.2.1 only with a qualified medical practitioner to the patient prescribing radiation therapy and should be administered to the patient by virtue of prescription treatment Therapeutic radiopharmaceuticals.
4.2.2 The practitioner should be the legitimacy to judge case by case, only to meet the requirements of the legitimacy of radiopharmaceuticals to issue medical prescriptions.
4.2.3 practitioners and related personnel should be strictly controlled indications, should clinical nuclear medicine diagnosis and treatment when administered to a patient sensitive radiopharmaceuticals strict control.
4.2.4 Before the implementation of the Nuclear Medicine Clinic, practitioners and stakeholders have a responsibility to possible risks to patients orally or in writing inform Or their families.
5 patient protection and safety requirements
5.1 Justification judgment
5.1.1 In determining the nuclear medicine clinic should first make a legitimacy judgment, to ensure that the expected benefits of treatment according to clinical need will exceed the Treatment may bring potential danger.
5.1.2 All new clinical nuclear medicine techniques and methods of treatment, should be made through the use of pre-judging the legitimacy; it has been judged proper technical and square Law, when made new or important evidence and the need to re-judgment, they should re legitimacy judgment.
5.1.3 clinical nuclear medicine physicians should have the corresponding characteristics of medical diagnostic imaging technology and its indications, its use should be strictly controlled indications range. Even if new clinical nuclear medicine techniques and methods has been done to judge the legitimacy, in for new indications shall also be carried out prior to the legitimacy judgment.
5.1.4 diagnostic radiopharmaceuticals, reference should guide the level of medical exposure, using expected to reach the minimum discharge required diagnostic purposes Radionuclide administered amount. And note view past patient examination data, in order to avoid unnecessary duplication of tests.
5.1.5 lactating and pregnant women administered diagnostic radiopharmaceuticals, should pay special attention to the legitimacy of judgment. Because the special needs of pregnant women pregnant When women perform imaging, we should assess the absorbed dose to the fetus suffered.
5.1.6 unless it is life-saving situations, pregnant women should not receive treatment radiopharmaceuticals, particularly those containing 131I and 32P of radiopharmaceuticals. Therapeutic radiopharmaceuticals, should normally be at the end of pregnancy and lactation were. When life-saving drugs and radioactive therapy, if the fetus Accept the dose does not exceed 100mGy, may not terminate the pregnancy.
5.1.7 In order to avoid fetal and embryonic accidental radiation exposure, whether pregnant patients respond to an inquiry, inspection and evaluation. And will have Off advisory note posted in clinical nuclear medicine departments of the place, especially the entrance and waiting area.
5.1.8 only if there is significant clinical indications only can children in radionuclide imaging examination, and should be based on the child's weight, body The surface area or other applicable guidelines to minimize the amount of radioactive drug administration should be as short as possible to select the half-life of the radionuclide.
5.1.9 In addition to the above provisions, it should also meet other requirements in GB Z179 legitimacy judgment.
5.2.1 General requirements
5.2.1.1 practitioner when issued radiopharmaceutical prescription should be.
a) In the diagnosis can achieve the desired objective case, the patients received a dose as low as possible;
b) To make full use of existing information, to avoid all unnecessary duplication of irradiation;
c) there should be an actual medical exposure of nuclear medicine and radiopharmaceutical treatment prescription consistent verification procedures;
d) in the implementation of nuclear medicine diagnostic tests, should refer to the appropriate medical exposure guidance levels (see Appendix A).
5.2.1.2 practitioners, technicians and other imaging staff can be reached in an acceptable image quality, the patient should take By dose as low as possible. Therefore, we should adopt the following measures.
a) Depending on the choice of appropriate characteristics of patients administered radiopharmaceutical and its activity are available, particularly for children and organ Patients with functional impairment;
b) non-blocking inspection organs using a radioactive drug absorption, and speed up the process;
c) Note that the use of appropriate image acquisition and processing technology;
d) practicing physicians, technicians and other imaging staff, in addition should have the appropriate professional skills, but also accept nuclear medicine diagnosis and treatment Quality control and technical training requirements and protection of knowledge.
5.2.1.3 unless there was a clear clinical indications for pregnant women receive treatment when administered radiopharmaceuticals should control. Women Administration of radiopharmaceuticals should be avoided during pregnancy.
5.2.2 diagnostic nuclear medicine in clinical optimization requirements
5.2.2.1 should technical specifications and related standards establish operating conditions, nuclear medicine equipment, in order to obtain the best image when, Patients receiving the smallest dose.
5.2.2.2 Nuclear Medicine staff to deal with the patient has been administered radiopharmaceuticals provide written and oral guidance, so that they are also after discharge Can effectively limit their exposure suffered by nurses and the public to reduce their family members, if not adults and pregnant women, in particular, his or her spouse Close.
5.3 Guidance Levels for Medical Exposure 5.3.1
5.3.2 practitioners should refer to the principles guiding the use of medical exposure level (see Appendix A) to ensure that the activity of the co-administered reason.
5.3.2 Guidance Levels for the following principles.
a) When the dose or activity significantly lower than the corresponding level of guidance, can not provide useful diagnostic information to patients or doctors expected When treatment benefits should be required to take corrective action;
b) When the dose or activity significantly exceeds the appropriate level of guidance, you should consider whether the guidance has reached the level of radiation protection optimization, or Medicine Practice activities are held in a good level;
c) in Appendix A of the guidance applies only to the le...
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 43 pages — is available in the English PDF.
Referenced standards
Similar standards
GB 18871-2002|GBZ 120|GBZ 179|GB 16348-2010|GB 16362-2010
Editions of GB 16361
| Edition | Title | Revision | Status |
|---|---|---|---|
| GB 16361-2012 | Specification for patient radiological protection and quality control in nuclear medicine | current edition | Current |
| GB 16361-1996 | Radiological protection standards for the patient in clinical nuclear medicine | previous edition | Superseded |
This page sells the current edition, GB 16361-2012. Earlier editions are listed for reference only.
How to Buy GB 16361-2012
- 1Add to cart. Click the "Buy GB 16361-2012" button on this page. You can add more standards before checkout.
- 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
- 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards