YY/T 1746-2020Absorbable surgical suture—Test method for breaking strength in vitro hydrolytic degradation (English PDF)

可吸收性外科缝线 体外水解后断裂强力试验方法

Preview PDF

This is a limited preview

Buy now to download the full PDF (20 pages)

Need quotation for your management?

Get a formal quotation for YY/T 1746-2020 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

September 27, 2020

Implementation date

September 1, 2021

Scope

YY/T 1746-2020 is the English-translated version of 可吸收性外科缝线 体外水解后断裂强力试验方法.

YY/T 1746-2020 is the Chinese pharmaceutical industry standard that gives the method for determining how much of its breaking strength an absorbable surgical suture retains after a period in a simulated body environment.

It sets out the specimens and their knots, the medium and its temperature and pH, the intervals of immersion, the tensile test that follows and the calculation of the retained strength as a fraction of the original, so that absorption profiles of different sutures can be compared. Issued on 27 September 2020, it has applied since 1 September 2021.

Found what you need?

Document preview — YY/T 1746-2020

National Standard of the People's Republic of China

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope4
  • 2 Principle4
  • 3 Apparatus and reagents4
  • 4 Test procedure5
  • 5 Test report7
  • References8

1 Scope

This Standard specifies the test method for the retention value of breaking strength of absorbable surgical suture (hereinafter referred to as suture) in the in vitro simulated environment.

This Standard applies to the absorbable surgical suture which is processed from synthetic polymers that are hydrolyzed and degraded.

2 Principle

Test the retention value of suture breaking strength in different time periods in an in vitro simulated environment, which provides reference information for whether the suture can maintain sufficient tension during the critical period of wound healing.

3 Apparatus and reagents

3.1 Material testing machine

The tensile speed is 300 mm/min ± 10 mm/min; the test gauge length is 130 mm ± 5 mm.

Note: If the suture length cannot meet the requirements of the specified test gauge length, the test gauge length can be adjusted according to the actual length of the suture; the corresponding stretching speed is set to 2 times the adjusted test gauge length. If the adjusted test gauge length is set to 50 mm, the stretching speed is set to 2×50 = 100 mm/min.

3.2 Electric-heating constant-temperature oscillating water tank or constant-temperature oscillator

The temperature is 37 °C ± 1 °C; the oscillation frequency: 60 r/min (or other suitable oscillation methods).

3.3 pH meter

The accuracy of the pH meter is not less than 0.02.

3.4 Container

4 Test procedure

4.2.4 Calculate the average of the 10 recorded force values, which is the initial breaking strength.

4.3 Test of residual breaking strength

4.3.1 Take 10 sutures of the same batch as the initial test value (4.2); put them into the container (3.4), so that the sutures are completely immersed in the test solution (3.5) (the ratio of the volume of the test solution to the mass of the suture is at least 10 mL:1 g).

Note: If the suture length can meet the test conditions, take the same suture to test the initial breaking strength value and the residual breaking strength value, respectively.

4.3.3 At the end of the test period, take out the suture and remove the water on the surface of the suture until there is no dripping.

Note: The method of removing water shall not affect the breaking strength of the suture.

4.3.5 Calculate the average of the 10 recorded force values, which is the residual breaking strength.

4.3.4 Carry out the test according to steps 4.2.1 ~ 4.2.3

4.3.2 Incubation

Place the container in the electric-heating constant-temperature. oscillating water tank (3.2); select the corresponding test period according to the material of the suture and the intended use. During the test period, the pH value of the test solution in the container shall be monitored as needed. If the pH value of the buffer exceeds the specified limit value (7.4 ± 0.2), use the 0.1 mol/L sodium hydroxide solution or dilute phosphoric acid to adjust or replace the buffer, to make it meet the requirements (3.5).

4.4 Result calculation

Calculate the retention value of breaking strength according to Formula (1):

Where: BSR -- Retention value of breaking strength; F0 -- Initial breaking strength; F1 -- Residual breaking strength. […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 20 pages — is available in the English PDF.

Similar standards

How to Download YY/T 1746-2020

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
20 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 1746-2020

$105.00