YY/T 1706.1-2020Implants for surgery—Plasma-sprayed unalloyed titanium coatings on metallic surgical implants - Part 1: General requirements (English PDF)
外科植入物 金属外科植入物等离子喷涂纯钛涂层 第1部分:通用要求
This is a limited preview
Buy now to download the full PDF (22 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1706.1-2020 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
February 21, 2020
Implementation date
January 1, 2021
Scope
YY/T 1706.1-2020 is the English-translated version of 外科植入物 金属外科植入物等离子喷涂纯钛涂层 第1部分:通用要求.
YY/T 1706.1-2020 is the Chinese pharmaceutical industry standard that sets the general requirements for plasma-sprayed unalloyed titanium coatings on metal surgical implants, the rough titanium layer that encourages bone to grow onto a cementless implant.
It is the first part of the YY/T 1706 series and covers the powder and the substrate, the coating thickness, its porosity and roughness, the bond strength to the substrate and the abrasion resistance, together with the tests for each.
It applies to atmospheric and vacuum plasma spraying, and not to coatings of other materials or made by other processes. It was issued on 21 February 2020 and has applied since 1 January 2021.
Found what you need?
Document preview — YY/T 1706.1-2020
National Standard of the People's Republic of China
- ICS
- 11.040.40
- Classification
- C 35
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 Requirements
- 5 Test report
Foreword
YY/T 1706 "Surgical implants metal surgical implants plasma sprayed pure titanium coating" is divided into the following parts.
---Part 1: General requirements; This part is Part 1 of YY/T 1706.
This part uses the redrafting method to modify and adopt ISO 13179-1:2014 "Plasma sprayed pure titanium for surgical plant metal surgical implants Coating Part 1: General Requirements.
Compared with ISO 13179-1:2014, except for editorial changes, the main technical changes in this part are as follows.
---Regarding normative reference documents, this standard has made adjustments with technical differences to adapt to my country's technical conditions and adjustments.
The situation is collectively reflected in Chapter 2 "Normative Reference Documents", and the specific adjustments are as follows.
- Replace ISO 4287 with GB/T 3505 which is equivalent to adopting international standards;
- Replace ISO 10993-1 with GB/T 16886.1 which is equivalent to the international standard;
- Replace ISO 14971 with YY/T 0316 which is equivalent to the international standard;
- Use YY/T 0988.2 instead of ASTM F1580;
- Replace ASTM F1147 with YY/T 0988.11;
- Use YY/T 0988.12 instead of ASTM F1044;
- Use YY/T 0988.13 instead of ASTM F1160;
- Replace ASTM F1854 with YY/T 0988.14;
- Replace ASTM F1978 with YY/T 0988.15;
--- Chemical analysis step (4.2.3) adds "or recognized analysis method (existing ISO method or method recommended by national standards)" to Adapt to my country's national conditions and facilitate the implementation of standards.
Please note that certain contents of this document may involve patents. The issuing agency of this document is not responsible for identifying these patents.
Introduction
So far, there is no surgical implant material that has no rejection in the human body. However, materials prepared with reference to this standard should be used in long-term clinical applications.
In use, it has been shown that when the application is appropriate, an acceptable level of biological response can be expected.
1 Scope
This part of YY/T 1706 specifies the general requirements for plasma sprayed pure titanium coatings on metal surgical implants. This section applies to atmospheric or vacuum plasma spraying.
This section does not apply to coatings processed by materials other than pure titanium materials, or coatings processed by other technologies than plasma spraying technology. Industrial coating.
Note. The quality management system specified in YY/T 0287 can be used. Refer to ISO /IEC 17025 for the requirements of testing laboratory capabilities.
2 Normative references
The following documents are indispensable for the application of this document. For dated reference documents, only the dated version applies to this document. For undated references, the latest version (including all amendments) applies to this document.
GB/T 3505 Product Geometric Technical Specification (GPS) Surface Structure Profile Method Terms, Definitions and Surface Structure Parameters (GB/T 3505-2009, ISO 4287:1997, IDT)
GB/T 16886.1 Biological Evaluation of Medical Devices Part 1: Evaluation and Testing in the Process of Risk Management (GB/T 16886.1- 2011, ISO 10993-1:2009)
YY/T 0316 Application of medical device risk management to medical devices (YY/T 0316-2016, ISO 14971:2007)
YY/T 0988.2 Surgical implant coating Part 2: Titanium and titanium-6 aluminum-4 vanadium alloy powder
YY/T 0988.11 Surgical implant coating Part 11: Calcium phosphate coating and metal coating tensile test method
YY/T 0988.12 Surgical implant coatings Part 12: Calcium phosphate coating and metal coating shear test method
YY/T 0988.13 Surgical implant coating Part 13: Shearing and bending of calcium phosphate, metal and calcium phosphate/metal composite coating Fatigue test method
YY/T 0988.14 Surgical implant coatings. Part 14: Stereological evaluation methods for porous coatings
YY/T 0988.15 Surgical implant coatings Part 15: Test method for wear resistance of metal thermal spray coatings
ISO 5832-2 Metallic materials for surgical implants Part 2: Pure titanium
3 Terms and definitions
The following terms and definitions apply to this document.
3.1 Plasma spraying
Thermal spraying method using plasma jet to process coating.
3.2 Plasma sprayed pure titanium coating
A coating formed by spraying pure titanium on the surface of the substrate through a plasma spraying process.
4 Requirements
4.1 Plasma spray powder
The powder used in plasma spraying should meet the requirements of YY/T 0988.2.
4.2 Chemical analysis
4.2.1 Chemical composition
The chemical composition of the pure titanium plasma spray coating should meet the requirements of Table 1.
Note. Although this standard is applicable to both atmospheric and vacuum plasma sprayed coatings, vacuum plasma sprayed coatings may be more chemical than atmospheric plasma sprayed coatings. The component value is low.
4.2.2 Sample preparation
For chemical analysis of the coating, at least 5g samples should be taken from the coating deposited on the pure titanium substrate sample, and the pure titanium substrate sample should meet The requirements of ISO 5832-2.If it is impossible to prepare a 5g sample from the coating, the sample weight can be reduced without affecting the chemical analysis.
The technique of preparing coating samples should not cause contamination of the coating. If the implant has undergone a cleaning step after spraying the coating, before chemical analysis The same cleaning procedure should be used for the sample.
4.2.3 Chemical analysis steps
The content of iron (Fe) element can be inductively coupled plasma atomic emission spectrometry (ICP-AES) in accordance with ASTM E2371 Or recognized analytical methods (existing ISO methods or methods recommended by national standards) for determination.
Nitrogen (N), oxygen (O), carbon (C) and hydrogen (H) element content should be determined by the combustion method, which should be recognized as effective.
When declaring that the chemical analysis results meet the limit content specified in Table 1, the analysis accuracy of the 95% confidence level should be considered.
Example. If the measurement uncertainty of the oxygen (O) element content at the 95% confidence level is 1%, it should be pointed out that the coincidence of the measured value is <=9%.
4.3 Morphology
Taking into account the effectiveness of the morphological test, if possible the test should be carried out on the final product. If the geometry of the final product is not When the morphological test cannot be performed on the final product because the standard requirements can be met, the sample can be used and it should be proved that the sample can represent the final product.
The average coating thickness and tolerance should be determined in accordance with YY/T 0988.14, in micrometers (µm).
The roughness (Ra or Rt) should be measured in accordance with GB/T 3505, and the unit is micron (µm). The assessment length should be at least 8mm.
For coatings with an average thickness >=300µm, the average pore volume percentage and average pore intercept shall be calculated in accordance with YY/T 0988.14 Measure the average pore volume in different layers of coating thickness according to YY/T 0988.14 ("Tissue Interface Gradient Method") Ratio and average pore intercept.
4.4 Coating mechanical properties
4.4.1 General
Each test specified in 4.4.2~4.4.4 should use at least 5 test samples, and the matrix of the test samples should be the same as the matrix of the final product. with. Both the test sample and the final product should be subjected to the same pretreatment (for example, cleaning, sandblasting) and post-treatment (for example, cleaning, sterilization).
4.4.2 Static shear strength
When tested in accordance with YY/T 0988.12, the average static shear strength of the coating should be greater than 20MPa.
4.4.3 Shear fatigue strength
When tested in accordance with YY/T 0988.13, the maximum shear fatigue stress is at least 10MPa, and the coating should be able to withstand 10 million cycles No failure occurs.
4.4.4 Static tensile strength
When tested in accordance with YY/T 0988.11, the average static tensile strength of the surface coating should be greater than 22MPa.
4.4.5 Wear resistance
When tested in accordance with YY/T 0988.15, the wear amount of the surface coating after 100 cycles should be less than 65mg.
4.5 Verification control
According to YY/T 0316, the chemical composition, morphology and tensile strength of the coating should be regularly controlled within the period determined by the risk analysis.
The routine control procedures for coating should specify the type and time interval of each control test.
4.6 Validity of numerical rounding
In order to ensure the consistency of all specified numerical ranges under various circumstances, the calculated and observed values should be rounded to two decimal places to be valid digital.
5 Test report
The test report should include at least the following information.
a) The identification of the test sample includes the source, date of receipt, and form;
b) The number of test samples used in each test;
c) References of the test methods used;
d) Deviation from the test method used, if applicable;
e) Special conditions observed in the test;
f) The reason for excluding data from all data;
g) Chemical analysis results (see 4.2);
h) Morphological analysis results (see 4.3);
i) Average static shear strength (see 4.4.2);
j) Shear fatigue strength (see 4.4.3);
k) Average static tensile strength (see 4.4.4);
l) The amount of wear (see 4.4.5);
m) Test date;
n) The identification of the laboratory where the test is conducted;
o) Signatures of laboratory managers and operators. references [1] ISO 5961 Determination of Cadmium by Water Quality Atomic Absorption Spectrometry.
[2] YY/T 0287 Medical device quality management system is used for regulatory requirements.
[3] ISO 16429 Surgical Implant Evaluation of Metal Implant Materials and Medical Devices Long-term Corrosion Behavior Measurement of Open Circuit Potential method.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 22 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 3505Geometrical product specifications (GPS) -- Surface texture: Profile method -- Terms, definitions and surface texture parameters
- GB/T 16886.1Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
- YY/T 0316Medical devices-Application of risk management to medical devices
- YY/T 0988.2Coatings of surgical implants - Part 2: Titanium and titanium-6 aluminum-4 vanadium alloy powders
- YY/T 0988.11Coatings of surgical implants - Part 11: Tension testing method of calcium phosphate coating and metallic coatings
- YY/T 0988.12Coatings of surgical implants - Part 12: Shear testing method of calcium phosphate coating and metallic coatings
ISO 5832-2
Similar standards
How to Download YY/T 1706.1-2020
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 16886.1-2022 — Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
GB/T 3505-2009 — Geometrical product specifications (GPS) -- Surface texture: Profile method -- Terms, definitions and surface texture parameters
YY/T 0316-2003 — Medical devices - Application of risk management to medical devices