YY/T 1620-2018Cardiopulmonary bypass systems—Assessment method of red blood cell damage in continuous flow blood pumps (English PDF)

心肺转流系统 连续流血泵红细胞损伤评价方法

Preview PDF

This is a limited preview

Buy now to download the full PDF (23 pages)

Need quotation for your management?

Get a formal quotation for YY/T 1620-2018 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

September 28, 2018

Implementation date

October 1, 2019

Scope

YY/T 1620-2018 is the English-translated version of 心肺转流系统 连续流血泵红细胞损伤评价方法.

YY/T 1620-2018 is the Chinese pharmaceutical industry standard that gives a method for assessing the damage a continuous flow blood pump does to red cells.

It sets out the test circuit and the blood used, the flow and pressure conditions, the duration of circulation and the sampling, and the calculation of the index of haemolysis from the free haemoglobin measured, so that pumps can be compared.

It applies to continuous flow pumps of a cardiopulmonary bypass system rated at about 5 L/min or more. Issued on 28 September 2018, it has applied since 1 October 2019.

Found what you need?

Document preview — YY/T 1620-2018

National Standard of the People's Republic of China

ICS
11.040.40
Classification
C 45

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword
  • Introduction
  • 1 Scope
  • 2 Normative references
  • 3 Terms and Definitions
  • 4 General
  • 5 Test Overview
  • 6 Preparation before the test
  • 7 Test step
  • 8 Acceptance of criteria
  • Appendix A (Normative Appendix) Selection of blood for blood pump red blood cell injury test in vitro
  • Appendix B (informative) Determination of plasma free hemoglobin and total hemoglobin Reference

Foreword

This standard was drafted in accordance with the rules given in GB/T 1.1-2009.

Introduction

The purpose of using a blood pump is to replace or assist the function of the human heart to complete the extracorporeal circulation. Currently developed for roller pumps (roler Pump) and centrifugal pumps (centrifugalpumps) are continuous blood flow pumps commonly used in clinical extracorporeal circulation. Blood pump not only used in routine In cardiopulmonary bypass (extracorporeal circulation) for cardiac surgery, it can also be used for ventricular assist, extracorporeal membrane oxygenation, and the like.

Hemolysis is one of the important parameters for evaluating red blood cell damage caused by blood pumps. Hemolysis caused by blood pump depends mainly on the design and knot of the blood pump Structure and materials used. The International Organization for Standardization Surgical Implants Technical Committee Cardiovascular Subcommittee (ISO /TC150/SC2) has been issued Cloth ISO 18242:2016 Cardiovascularimplantsandextracorporealsystems-Centrifugalbloodpumps, In the terms of the requirements and methods, there is a special evaluation requirement for red blood cell damage, and the basic conditions of blood for testing (blood flow)

Rate, blood glucose, hemoglobin content); in the test procedure clause, the basic composition, test time, sampling point and test of the test tooling are given.

Project (including detection indicators for free hemoglobin). However, the hemolysis performance evaluation was not included in the relevant blood pump standards in China.

In the evaluation of the potential risk of blood pump, it undoubtedly left a safety hazard to the use. And in GB/T 16886.4 "Medical Device Biology Review Price Part 4. Selection of Tests for Interaction with Blood does not address fluid dynamics factors in continuous circulation of instruments such as blood pumps, Structure and other potential damage to red blood cells. Therefore, it is very necessary to establish a standardized in vitro hemolysis test for evaluating continuous blood flow pumps. law. Although the basic conditions for blood for testing are given in the ISO 18242 standard, detailed blood selection criteria for testing are not given. thing In fact, the different blood quality used for the test will seriously affect the results of the test; the detailed test tool design and continuous are not given in the standard.

The test method of the flow cycle 6h. ASTM F1830-1997 (R2013) Standard Practice for Assessment of HemolysisinContinuousFlowBloodPumps and ASTM F1841-1997(R2013)StandardPracticefor SelectionofBloodforinvitroEvaluationofBloodPumps gives detailed test blood selection conditions and specific Test tool design and continuous cycle 6h test method. This standard optimizes the design of the test tooling and supplements the free hemoglobin. The test method improves the test procedure.

1 Scope

This standard gives a specific method for evaluation of red blood cell damage in continuous blood pump.

This standard is applicable to the evaluation of red blood cell injury in continuous blood flow pump (nominally 5L/min or more) in cardiopulmonary bypass system.

2 Normative references

The following documents are indispensable for the application of this document. For dated references, only dated versions apply to this document. For undated references, the latest edition (including all amendments) applies to this document.

GB/T 16886.4 Biological evaluation of medical devices - Part 4. Selection of test for interaction with blood (GB/T 16886.4- 2003, ISO 10993-4:2002, IDT)

3 Terms and definitions

The following terms and definitions as defined in GB/T 16886.4 apply to this document.

3.1 Continuous blood flow pump

A device that produces a continuous flow of blood driven by a driving force.

3.2 Plasma free hemoglobin

Red blood cells destroy hemoglobin (globin or heme containing protein) released into the plasma.

3.3 Hemolysis

Destruction of red blood cell damage, leading to the phenomenon of hemoglobin entering the plasma.

3.4 Standard hemolysis index normalizedindexofhemolysis; NIH

The amount of plasma free hemoglobin (g) increased per pump of 100 L of blood, corrected for plasma volume by hematocrit, and Corrected with flow rate and cycle time.

3.5 Standard millimeter hemolysis index normalizedmiligramindexofhemolysis;mg.NIH

A standard milligram of hemolysis index expressed as milligrams of plasma free hemoglobin.

3.6 Modified hemolysis index modifiedindexofhemolysis; MIH

The ratio of the amount of hemoglobin released into the plasma to the normalized total amount of hemoglobin delivered by the blood pump.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 23 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download YY/T 1620-2018

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
23 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 1620-2018

$205.00