Scope

YY/T 1566.1-2017 is the English-translated version of 一次性使用自体血处理器械第1部分:离心杯式血细胞回收器.

YY/T 1566.1-2017 is the Chinese pharmaceutical industry standard for the centrifuge bowl type blood cell recovery set used in surgery, the single-use disposable circuit of a cell salvage machine.

It is the first part of the YY/T 1566 series and sets the requirements that make the set safe with its machine, covering the bowl and the tubing, the reservoir and its filter, the anticoagulant and wash lines, the connectors, the strength and tightness under centrifugation and pressure, the particulate contamination, the materials and their biological evaluation and sterility. Issued on 17 July 2017, it has applied since 1 July 2018.

Found what you need?

Document preview — YY/T 1566.1-2017

National Standard of the People's Republic of China

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword4
  • Introduction5
  • 1 Scope6
  • 2 Normative references6
  • 3 Terms and definitions7
  • 4 Structure9
  • 5 Physical requirements9
  • 5.1 Appearance9
  • 5.2 Particle contamination10
  • 5.2.1 Separation bowl10
  • 5.2.2 Pipeline system10
  • 5.2.3 Red blood cell collection bag10
  • 5.3 Sealing10
  • 5.4 Connection strength11
  • 5.5 Frictional heat11
  • 5.6 Noise11
  • 5.7 Residual amount11
  • 5.8 Color scale12
  • 5.9 Piercer12
  • 5.10 Drip bucket12
  • 5.11 Flow regulator12
  • 5.12 Pump tube) elasticity12
  • 5.13 Position limit clip) (if any)12
  • 5.14 Fixture13
  • 5.15 Interface13
  • 5.16 Protective sleeve13
  • 5.17 Red blood cell collection bag13
  • 6 Chemical requirements13
  • 6.1 Separation bowl and pipeline system13
  • 6.1.1 Reducing substances13
  • 6.1.2 Metal ions14
  • 6.1.3 pH14
  • 6.1.4 Evaporation residue14
  • 6.1.5 UV absorbance14
  • 6.2 Red blood cell collection bag14
  • 6.3 Ethylene oxide residues14
  • 7 Biological requirements15
  • 7.1 Biological compatibility15
  • 7.2 Sterile15
  • 7.3 Bacterial endotoxin15
  • 8 Type test15
  • 9 Markings15
  • 9.1 General15
  • 9.2 Single package marking16
  • 9.3 Shelf packaging marking16
  • 10 Packaging17
  • Appendix A (Informative) Descriptions18
  • Appendix B (Normative) Physical test21
  • Appendix C (Normative) Chemical test References25

1 Scope

This part of YY/T 1566 specifies the requirements for the centrifuge bowl type blood cell recovery set used in surgical operations. It is a single-use product to ensure the safety of use with autologous blood recovery device.

This part applies to centrifuge bowl type blood cell recovery set that are intended to be used in conjunction with other autologous blood processing devices during surgery, and are expected to be used in conjunction with autologous blood recovery device. It consists of a tube system, a separation bowl, a red blood cell collection bag, and a waste solution bag. It does not include a blood collection filter (reservoir).

2 Normative references

The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) are applicable to this standard.

GB/T 6682 Water for analytical laboratory use - Specification and test methods

GB 8368 Infusion sets for single use, gravity feed

GB 14232.1-2004 Plastics collapsible containers for human blood and blood component - Part 1. Conventional containers

GB/T 14233.1-2008 Test methods for infusion reinfusion injection equipment for medical use - Part 1. Chemical analysis methods

GB/T 14233.2 Test methods for infusion, reinfusion, injection equipment for medical use - Part 2. Biological test methods

GB/T 16886.1 Biological evaluation of medical devices - Part 1. Evaluation and testing within a risk management process

5 Physical requirements

5.1 Appearance

5.1.5 The interface of the separation bowl, clamps, and other component

surfaces shall be smooth and free from burrs.

5.2 Particle contamination . 10

5.2.1 Separation bowl . 10

When measured in accordance with B.1, the number of particles 15 µm ~ 25 µm in 200 mL eluent shall not exceed 6.00 particles/mL, particles greater than 25 µm shall not exceed 3.00 particles/mL.

5.2.2 Pipeline system . 10

When measured in accordance with B.2, the number of particles 15 µm ~ 25 µm on the surface area per square centimeter shall not exceed 1.00, the number of particles larger than 25 µm shall not exceed 0.50.

5.2.3 Red blood cell collection bag . 10

The production of red blood cell collection bags shall avoid particle contamination.

When tested in accordance with the provisions of B.3, red blood cell collection bag shall have no visible particles.

Note. Work is under way to establish limits on the number and size of particles.

The limits and test methods given in pharmacopoeias (such as the limits and methods of preparations specified in the European Pharmacopoeia) are currently available.

5.3 Sealing . 10

5.3.1 When the separation bowl is subject to particle contamination test in accordance with B.1, the separation bowl shall be free from leakage.

Note. The sealability of the separation bowl and the particle contamination test are combined.

5.3.2 SEAL the other ends of the pipelines in the pipeline system, LEAVE one end and LEAD in 50 kPa air pressure for 15 s. There shall be no continuous air bubble leakage in the underwater test.

6 Chemical requirements

6.1 Separation bowl and pipeline system

6.1.2 Metal ions .. 14

When measured by atomic absorption spectrophotometry (AAS) or equivalent, the total content of antimony, chromium, copper, lead, and tin in the extract shall not exceed 1 µg/mL, the content of cadmium shall not exceed 0.1 µg/mL.

When tested in accordance with C.3, the color present in the extract shall not exceed the standard control solution with a mass concentration rho(Pb2+) = 1 µg/mL.

6.1.3 PH . 14

When tested in accordance with C.4, the difference between the pH of the test solution prepared in accordance with C.1.3 and the blank solution shall not exceed 1.5.

6.1.4 Evaporation residue .. 14

When tested in accordance with C.5, the total amount of evaporation residue in 50 mL of the test solution prepared in accordance with C.1.3 shall not exceed 2 mg.

6.1.5 UV absorbance . 14

When tested in accordance with C.6, the absorbance of the test solution prepared in accordance with C.1.3 in the range of 250 nm ~ 320 nm shall not exceed 0.1.

6.2 Red blood cell collection bag .. 14

The test solution prepared in accordance with C.1.4 shall meet the requirements of GB 14232.1-2004.

6.3 Ethylene oxide residues .. 14

If ethylene oxide is used for sterilization, when tested in accordance with Appendix C, the residual ethylene oxide residuals per set of centrifuge bowl type blood cell recovery set (excluding the waste solution bag) shall not exceed 4.0 mg.

Appendix A (Informative) Descriptions

A.1 Scope

Appendix A of ANSI/AAMI AT6.2013 "Autologous blood reinfusion devices" describes the clinical application of autologous blood reinfusion as follows.

Autologous blood reinfusion refers to the collection, storage, and reinfusion of blood products from the same patient. There are many cases where an autologous blood reinfusion is ideally useful or life-saving, especially when allogeneic blood is not immediately available. These conditions include the following types of clinical autologous blood reinfusion.

a) Collection, storage and reinfusion during optional surgery or emergency during surgery;

b) Collection, storage and reinfusion in the event of a life-threatening injury or major blood loss.

All autologous blood reinfusion devices will have blood contact with the surrounding environment, so it will use anticoagulants, purification (filtration), washing and reinfusion. Therefore, all kinds of instruments used in these links have dangerous sources. These dangerous sources should be recognized by a standard. Due to the group of patients who use these devices, most reinfusion devices do not have an advantage in avoiding dangerous sources. Because most reinfusions need to handle one or more circulating blood volumes, such a procedure will amplify the deleterious effects on blood (proteins, forming elements, etc.). Moreover, since autologous blood reinfusion devices more or less form the extracorporeal blood circuit, regardless of the intended use or application of these devices, the performance criteria of these devices should consider the performance of such aspects as the extracorporeal circuit of cardiac surgery and non-cardiac surgery. The blood collected from the open wound surface is in contact with external substances other than air, thromboplastin, chips, fat, and other circulatory systems, so it is contaminated blood. Air can be removed through the reservoir, which is usually not a problem in autologous blood reinfusion devices equipped with an external oxygenator.

However, when using the direct blood contact interface to quickly introduce blood into the patient with air pressure, air pollution becomes a problem.

Because there is no air capture device in the oxygenator, it is dangerous to use

Note. When the manufacturer gives the centrifuge speed, the test is carried out in accordance with the centrifuge speed given by the manufacturer. Otherwise, the test is carried out in accordance with the centrifuge speed of 5600 r/min.

B.5 Residual amount

At 23 °C ± 2 °C ambient conditions, WEIGH (accurate to 0.1 g) separation bowl, PLACE it in the autologous blood recovery set, in accordance with conventional use method or instruction manual, INSTALL and CONNECT the centrifugal bowl type blood cell recovery set, USE the distilled water at 23 °C ± 2 °C as a circulation solution. MAKE adjustment and control in accordance with autologous blood recovery set operating procedures, MAKE it circulate once at the centrifuge speed 5600 r/min, after the reinfusion is completed, WEIGH the separation bowl again, USE the difference between the two weighing results divided by the water density (taken as 1 g/mL) to obtain the separation bowl blood residual amount.

Note. When the manufacturer gives the centrifuge speed, the test is carried out in acc...

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 27 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download YY/T 1566.1-2017

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
27 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 1566.1-2017

$125.00