YY/T 1555.2-2018Particular requirements for the silicone gel filled breast implant—Performance requirement of silicone gel filler - Part 2: Limits requirement for extractable substances (English PDF)

硅凝胶填充乳房植入物专用要求 硅凝胶填充物性能要求 第2部分:可浸提物质限量要求

Preview PDF

This is a limited preview

Buy now to download the full PDF (17 pages)

Need quotation for your management?

Get a formal quotation for YY/T 1555.2-2018 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

December 20, 2018

Implementation date

January 1, 2020

Scope

YY/T 1555.2-2018 is the English-translated version of 硅凝胶填充乳房植入物专用要求 硅凝胶填充物性能要求 第2部分:可浸提物质限量要求.

YY/T 1555.2-2018 is the Chinese pharmaceutical industry standard for the extractable substances of the silicone gel filler of a breast implant.

It is the second part of the YY/T 1555 series and sets the extraction methods, the methods of determination and the limits: the volatile and low molecular weight siloxanes, the platinum residues and the other extractables, with the solvent, the conditions and the analytical technique fixed so that results are comparable. Issued on 20 December 2018, it has applied since 1 January 2020.

Found what you need?

Document preview — YY/T 1555.2-2018

National Standard of the People's Republic of China

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative references4
  • 3 Terms and definitions4
  • 4 Limit requirements5
  • 5 Detection methods6
  • Appendix A (Normative) Preparation method of extract7
  • Appendix B (Normative) Determination of residual amount of oligosiloxane substances: gas chromatography (GC method)8
  • Bibliography11

1 Scope

This Part of YY/T 1555 specifies the extraction methods, detection methods and limit requirements for extractable substances in the silicone gel filler used for artificial mammary implant.

This Part applies to artificial mammary implants with silicone gel as the inner filler.

2 Normative references

The following documents are indispensable for the application of this document.

For dated references, only the dated version applies to this document. For undated references, the latest edition (including all amendments) applies to this Standard.

YY 0334-2002, General specification for surgical implants made of silicone elastomer

YY/T 1457, Non-active surgical implants - Determination method for oligomeric siloxanes in mammary implants

3 Terms and definitions

The following terms and definitions apply to this document.

3.1 Mammary implant

An implant with a shell, which is used for increasing the volume of the breast or replacing the breast. The shell can be filled with objects by a manufacturer or a surgeon. [YY 0647-2008, definition 3.5]

3.2 Filler

The contents of the implant.

Appendix B (Normative)

Determination of residual amount of oligosiloxane substances: gas chromatography (GC method)

B.1 Principle

Use gas chromatography to analyze the small molecules in the extract. When the vaporized sample is carried by the mobile phase into the chromatographic column for operation, it is repeatedly distributed between the two phases. Due to the different distribution coefficients of each component between the mobile phase and the stationary phase, the running speed of each component in the chromatographic column is different. After a certain column length, they are separated from each other. They leave the chromatographic column and enters the detector in sequence. The detector can convert the sample component information into an electrical signal; the size of the electrical signal is proportional to the amount or concentration of the measured component, so as to realize the detection of the measured component.

B.2 Apparatus

B.2.1 Gas chromatograph, with hydrogen flame ionization detector. B.2.2 Electronic balance (precision of 0.1 mg).

B.3 Reagents

B.3.1 Acetone, required to be analytical grade and above. B.3.2 Octamethylcyclotetrasiloxane (D4), required to be analytical grade and above. B.3.3 Decamethylcyclopentasiloxane (D5), required to be analytical grade and above.

B.4 Test methods

B.4.1 Standard solution preparation

Take about 0.05 g of D4, D5 and other oligosiloxane substances respectively; accurately weigh them; use acetone to dilute the solution to 50 mL; then, use acetone to dilute the solutions successively, to prepare mixed standard solutions of D4 and D5 oligosiloxane substances of at least 5 concentration gradients from 1 µg/mL to 50 µg/mL.

B.4.2 Gas phase determination

Chromatographic conditions:

Chromatographic column: capillary column (30 m × 0.25 mm × 0.25 µm) with 5% phenyl-95% polydimethylsiloxane as the fixed liquid;

Column temperature: 60 °C (hold for 3 min), raise to 300 °C at 10 °C/min (hold for 10 min); Inlet temperature: 280 °C; Detector temperature: 300 °C;

Carrier gas: helium or nitrogen (purity above 99.99%), flow rate of 1.0 mL/min; Injection volume: 1 µL.

Take the blank solution, standard solution, and sample solution (see A.3 for the preparation method) for the gas chromatograph test. Take a standard solution of an intermediate concentration and inject consecutively 5 times, to calculate the relative standard deviation (RSD) of the peak area of oligosiloxane substances.

B.5 Result analysis

B.5.1 Use the external standard method to analyze the test results; establish a standard working curve and a regression equation; calculate the residual amount of oligosiloxane substances in the sample according to Formula (B.1).

The measurement results are expressed as the ratio of the mass (µg) of oligosiloxane substances to the mass (g) of the test mass; the unit is µg/g, namely, mg/kg.

Where:

X - the residual amount of oligosiloxane substances, in micrograms per gram (µg/g, namely, mg/kg); c - the concentration value of oligosiloxane substances in the sample test solution, which is read on the standard curve, in micrograms per milliliter (µg/mL);

V - the volume of the added extraction solvent acetone, in milliliters (mL); m - the mass of the sample, in grams (g).

B.5.2 If the concentration value read on the standard curve is less than the minimum concentration value of the standard solution, the result is expressed by Formula (B.2):

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 17 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download YY/T 1555.2-2018

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included