YY/T 1529-2017ELISA analytical instruments (English PDF)
酶联免疫分析仪
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1529-2017 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
May 2, 2017
Implementation date
April 1, 2018
Scope
YY/T 1529-2017 is the English-translated version of 酶联免疫分析仪.
YY/T 1529-2017 is the Chinese standard "ELISA Analytical Instruments".
Its scope clause reads: This Standard defines the terms and definitions of ELISA analytical instruments (hereinafter referred to as the analytical instruments), and specifies the classification, requirements, test method, marking, labelling, operating manual, packaging, storage and transportation.
This Standard applies to the reading modules of ELISA analytical instruments and full-automatic ELISA analytical instruments. Its clauses include application scope; terms and definitions; classification; requirements; experimental method; marking, labelling and operating manual; packaging, storage and transportation.
It was issued by the China Food and Drug Administration on MAY 2, 2017 and took effect on APRIL 1, 2018.
Found what you need?
Document preview — YY/T 1529-2017
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Application Scope4
- 2 Normative References4
- 3 Terms and Definitions5
- 4 Classification5
- 5 Requirements6
- 6 Experimental Method7
- 7 Marking, Labelling and Operating Manual12
- 8 Packaging, Storage and Transportation12
- Substance for the ELISA Analytical Instruments13
- Bibliography14
1 Application Scope
This Standard defines the terms and definitions of ELISA analytical instruments (hereinafter referred to as the analytical instruments), and specifies the classification, requirements, test method, marking, labelling, operating manual, packaging, storage and transportation.
This Standard applies to the reading modules of ELISA analytical instruments and full- automatic ELISA analytical instruments.
2 Normative References
The following referenced documents are indispensable for the application of this document. For dated references, only the edition dated applies to this document. For undated references, the latest edition of the referenced documents (including all amendments) applies to This Standard.
GB/T 191, Packaging - Pictorial Marking for Handling of Goods
GB 4793.1, Safety Requirements for Electrical Equipment for Measurement,
Control and Laboratory Use - Part 1: General Requirements
GB/T 14710, Environmental Requirement and Test Methods for Medical Electrical
Equipment
GB/T 18268.1, Electrical Equipment for Measurement, Control and Laboratory Use
- EMC Requirements - Part 1: General Requirements
GB/T 18268.26, Electrical Equipment for Measurement, Control and Laboratory
Use - EMC Requirements - Part 26: Particular Requirements - In Vitro Diagnostic (IVD) Medical Equipment
GB/T 29791.3, In Vitro Diagnostic Medical Devices - Information Supplied by the
Manufacturer (Labelling) - Part 3: In Vitro Diagnostic Instruments for Professional
Use
JJG 861-2007, Verification Regulation of ELISA Analytical Instruments
YY 0648, Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use
The stability of absorbance of the instruments shall not exceed ± 0.005.
5.2.6 Sensitivity
Use the sensitivity concentration standard substance of concentration 5 mg/l for the
ELISA analytical instruments; and the measured absorbance value of the instruments shall be not less than 0.01.
5.2.7 Difference of channels
Using the air as the reference, measure the absorbance difference between 8 channels;
and the required result shall not be greater than 0.02.
5.3 Electrical safety requirements
As specified in the applicable clauses of GB 4793.1 and YY 0648.
5.4 Environmental test requirements
As specified in the applicable clauses of GB/T 14710.
5.5 EMC requirements
As specified in the applicable clauses of GB/T 18268.1 and GB/T 18268.26.
5.6 Software functions
The ELISA analytical instruments shall, at least, have the following functions:
a) having the data display function;
b) being capable of setting up test items, test methods and test dates flexibly;
c) being capable of continuously storing the test items and results in succession;
d) being capable of displaying and printing the test items, test data and test time.
6 Experimental Method
6.1 Operating conditions
6.1.1 Power supply requirements: voltage 220 V ± 22 V (a.c.); frequency 50 Hz ± 1
Hz.
6.1.2 Environmental temperature: + 10°C ~ +40°C.
6.1.3 Relative humidity: 30% ~ 80%.
6.1.4 The following items shall be carried out after the instrument is started and stabilized for 30 min. When the following experiments are carried out on the instrument,
Annex A (Informative)
Preparation Method of Sensitivity Solution Standard Substance for the
ELISA Analytical Instruments
For the sensitivity solution standard substance for the ELISA analytical instruments, use pure substance of potassium dichromate and sulfuric acid 0.05 mol/l as the matrix to prepare the solution standard substance 5 mg/l.
A.1 Preparation of sulfuric acid solution 0.05 mol/l
Add 400 ml of distilled water into a 500 ml beaker; use a pipette to absorb 2.7 ml of concentrated sulfuric acid into the beaker; stir up before pouring into a 1 000 ml volumetric flask; add distilled water to dilute to the scale line; and shake up as standby.
A.2 Preparation of potassium dichromate solution standard substance of
chromium content 5 mg/l
A.2.1 Step 1
Weigh accurately 0.282 9 g of potassium dichromate on the analytical balance to place into a 100 ml beaker; use sulfuric acid solution 0.05 mol/l to dissolve before transferring to a 500 ml volumetric flask; use a small amount of sulfuric acid solution 0.05 mol/l to wash the beaker for several times; pour the washings into the volumetric flask; use sulfuric acid solution 0.05 mol/l to dilute to the scale line; and shake up as standby.
The concentration of the solution is 200 mg/l.
A.2.2 Step 2
Use a pipette to transfer 2.5 ml of potassium dichromate solution 200 mg/l into a 100 ml volumetric flask; use sulfuric acid solution 0.05 mol/l to dilute to the scale line; and shake up. Then, the concentration of the solution is 5 mg/l.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 191Graphical symbols marking for handling and storage of packages
- GB 4793.1Safety requirements for electrical equipment for measurement,control,andlaboratory use-Part 1:General requirements
- GB/T 14710The environmental requirement and test methods for medical electrical equipment
- GB/T 18268.1Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements
- GB/T 18268.26Electrical equipment for measurement, control and laboratory use -- EMC requirements -- Part 26: Particular requirements -- In vitro diagnostic (IVD) medical equipment
- GB/T 29791.3In vitro diagnostic medical devices -- Information supplied by the manufacturer (labelling) -- Part 3: In vitro diagnostic instruments for professional use
Similar standards
How to Download YY/T 1529-2017
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 4793.1-2007 — Safety requirements for electrical equipment for measurement,control,andlaboratory use-Part 1:General requirements
GB/T 14710-2009 — The environmental requirement and test methods for medical electrical equipment
GB/T 18268.1-2025 — Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements