YY/T 1282-2022Intravenous catheter for single use (English PDF)

一次性使用静脉留置针

Preview PDF

This is a limited preview

Buy now to download the full PDF (18 pages)

Need quotation for your management?

Get a formal quotation for YY/T 1282-2022 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

May 18, 2022

Implementation date

June 1, 2023

Scope

YY/T 1282-2022 is the English-translated version of 一次性使用静脉留置针.

YY/T 1282-2022 is the Chinese pharmaceutical industry standard for the single-use intravenous catheter inserted into a peripheral vein for infusion or transfusion.

It sets the requirements that make the catheter work safely with the infusion and transfusion equipment it is joined to: the materials and the biological evaluation, the dimensions and the flow rate, the strength of the catheter and of its bond to the hub, the conical fitting, freedom from leakage, the needle point and its protection, sterility, and the packaging and marking.

It replaces YY 1282-2016, was issued on 18 May 2022 and has applied since 1 June 2023.

Found what you need?

Document preview — YY/T 1282-2022

National Standard of the People's Republic of China

Replacing
YY 1282-2016

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword4
  • Introduction5
  • 1 Scope6
  • 2 Normative references6
  • 3 Terms and definitions7
  • 4 Structure8
  • 5 Physical requirements8
  • 5.1 Particulate contamination8
  • 5.2 Specifications8
  • 5.3 Catheter unit9
  • 5.3.1 Outer surface9
  • 5.3.2 Matching9
  • 5.3.3 Bending resistance10
  • 5.3.4 Peak tensile force10
  • 5.3.5 Connection strength11
  • 5.3.6 Infusion port11
  • 5.3.7 Multi-way stopcock11
  • 5.3.8 Extension tube11
  • 5.3.9 Blocking device11
  • 5.4 Needle unit12
  • 5.4.1 Needle tube12
  • 5.4.2 Needle tip12
  • 5.4.3 Connection strength12
  • 5.5 Lubricant12
  • 5.6 Flow rate12
  • 5.7 Venting12
  • 5.8 Flashback13
  • 5.9 No leakage13
  • 5.10 Adhesion13
  • 5.11 Corrosion resistance13
  • 5.12 Radiographic detectability13
  • 5.13 Protective piece14
  • 5.14 Anti-needlestick protection device14
  • 6 Chemical requirements14
  • 7 Biological requirements14
  • 7.1 Biocompatibility14
  • 7.2 Sterility14
  • 7.3 Bacterial endotoxins14
  • 8 Marks15

1 Scope

This document specifies the requirements for intravenous catheter for single use (hereinafter referred to as "intravenous catheter") for insertion into the human body's peripheral venous system for infusion or blood transfusion, to ensure compatibility with corresponding infusion and blood transfusion equipment.

This document applies to the intravenous catheter inserted into the human body's peripheral venous system for infusion or blood transfusion.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

GB/T 1962.2, Conical fittings with a 6% (Luer) taper for syringes, needles and certain other medical equipment -- Part 2: Lock fittings

GB/T 6682, Water for analytical laboratory use -- Specification and test methods

GB 8368, Infusion sets for single use -- Gravity feed

GB/T 14233.2-2005, Test methods for infusion, transfusion, injection equipment for medical use -- Part 2: Biological test methods

GB/T 16886.1, Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process

GB/T 18457, Stainless steel needle tubing for the manufacture of medical devices -- Requirements and test methods

GB/T 19633.1, Packaging for terminally sterilized medical devices -- Part 1. Requirements for materials, sterile barrier systems and packaging systems

YY 0285.1-2017, Intravascular catheters -- Sterile and single-use catheters -- Part

1 General requirements

YY 0285.5-2018, Intravascular catheters -- Sterile and single-use catheters -- Part

5 Over-needle peripheral catheters

YY 0581.1-2011, Infusion access adapters. Part 1: Needle access adapters (Heparin plugs)

YY 0581.2-2011, Infusion access adapter. Part 2: Needleless access adapters

YY 0585.2-2019, Fluid lines for use with pressure infusion equipment and accessories for single use -- Part 2: Accessories

YY/T 0615.1, Requirements for medical devices to be designated "STERILE". Part

1 Requirements for terminally sterilized medical devices

ISO 23908, Sharps injury protection -- Requirements and test methods -- Sharps protection features for single-use hypodermic needles, introducers for catheters and needles used for blood sampling

3 Terms and definitions

For the purposes of this document, the following terms and definitions apply.

3.1 Distal end

The last part that the catheter is inserted into the patient's vein as far as possible.

3.2 Outside diameter of the catheter

The maximum outer diameter of the catheter.

3.3 Effective length of the catheter

The maximum length the intravenous catheter can be inserted into a patient's vein.

NOTE. See l in Figure 1 of YY 0285.1-2017.

3.4 Infusion port

The interface intended to be connected to infusion, transfusion or injection equipment.

NOTE. Generally, there are 6% Luer conical connectors, puncture connectors (heparin caps), needle-free interface connectors (including positive pressure types), etc.

3.5 Multi-way stopcock

An assembly that can open a certain passage while closing all other passages, thereby achieving switching between multiple passages.

3.6 Vent fitting

A connector used to expel gas and limit or better prevent blood flow, divided into fixed and movable types.

3.7 Extension tube

A pipeline in which one end of the intravenous catheter is connected to the catheter seat, and the other end is connected to the infusion port.

3.8 Catheter unit

An assembly consisting of the catheter, catheter hub, extension tube (if any), and any integral connectors.

3.9 Needle unit

An assembly consisting of a needle barrel and needle hub and/or any integral accessories.

3.10 Flashback

Blood flows into the needle hub or catheter.

3.11 High-pressure angiographic catheter

An intravenous catheter for injecting contrast medium into a peripheral vein.

4 Structure

A typical intravenous catheter consists of two parts. a catheter unit and a needle unit.

5 Physical requirements

5.1 Particulate contamination

When tested according to Annex A, the contamination index of the intravenous catheter should not exceed 90.

5.2 Specifications

5.2.1 The outer diameter of the catheter of the intravenous catheter shall comply with the requirements of Table 1.The catheter unit shall have a color code in accordance with the requirements of Table 1.

When the blocking device is in the blocking state, a hydraulic pressure of 20 kPa is passed through it for 15 s and no liquid should pass through.

5.4 Needle unit

5.4.1 Needle tube

If steel pipes are used, they should comply with the requirements of GB/T 18457.

5.4.2 Needle tip

When examined with normal or corrected vision at 2.5 times magnification, the needle tip should be sharp, without curling, notches or hooks.

NOTE 1.Annex C of YY 0285.5-2018 gives examples of typical needle tip geometries.

NOTE 2.Annex B provides a method for evaluating the puncture performance of an intravenous catheter and catheter when they are in harmony.

5.4.3 Connection strength

When tested according to Annex A of YY 0285.5-2018, the needle tube should not be loose in the needle seat.

5.5 Lubricant

If catheters and needles are lubricated, there should be no visible accumulation of lubricant on the outer surfaces of the catheters and needles as observed with normal or corrected vision.

5.6 Flow rate

When tested according to Annex E of YY 0285.1-2017, the flow rate of a conduit with a nominal outer diameter less than 1.0 mm should be no less than 80% of the manufacturer's nominal value. The flow rate of a conduit with a nominal outer diameter greater than or equal to 1.0 mm should be no less than 90% of the manufacturer's nominal value.

5.7 Venting

If the intravenous catheter has a vent fitting, when tested according to Annex D of YY 0285.5-2018, there should be no liquid leakage from the vent fitting within 15 s.

5.13 Protective piece

The catheter end of the intravenous catheter and the Luer taper connector should have a protective piece. The protective piece should be firm but easy to remove.

5.14 Anti-needlestick protection device

The intravenous catheter should be equipped with anti-needlestick protection devices.

If there is an anti-needlestick protection device, it should comply with the requirements of ISO 23908.

6 Chemical requirements

When testing the test solution prepared in Annex F according to the method specified in GB 8368, it shall comply with the requirements of GB 8368.

7 Biological requirements

7.1 Biocompatibility

Intravenous catheters should be subjected to biological evaluation according to the requirements of GB/T 16886.1.The evaluation results should indicate that they are free of biological hazards.

7.2 Sterility

The sterility of intravenous catheters should comply with the requirements of YY/T 0615.1.

7.3 Bacterial endotoxins

When testing according to Chapter 4 of GB/T 14233.2-2005, use an endotoxin-free syringe to draw 5 mL of extraction medium and connect it to an indwelling catheter.

Inject the solution into the lumen of the intravenous catheter until full, then seal the distal end of the catheter. Place the syringes together in a 37°C incubator and allow them to incubate for 1 h. Push the remaining extraction medium in the syringe through the lumen of the intravenous catheter. Cut the catheter at the base. Place the cut catheter in the collection solution. Place the syringe in a 37°C incubator and incubate for 1 h.

Collect the entire extract for testing. The bacterial endotoxin limit should be less than 0.5 EU/mL.

NOTE. YY/T 0618 provides Test methods for bacterial endotoxins of medical devices - Routine monitoring and alternatives to batch testing.

8 Marks

8.1 Product marks

High-pressure angiographic catheters should be marked with a high-pressure resistance mark.

NOTE. Words or colors such as "High pressure resistant" or "P" can be used for identification.

8.2 Single packaging mark

At least the following clearly identifiable information should be provided on the single package.

a) Description of contents (name, catheter outer diameter and length, etc.);

b) For single use only, or equivalent;

c) Sterility;

d) Method of sterilization;

e) Radiopaque, if applicable;

f) Color code, unless the product's color code is visible through the individual packaging;

g) Batch number;

h) Date of manufacture, expiration date, or date of expiration;

i) Nominal flow rate;

j) Maximum nominal pressure for high-pressure angiographic catheter;

k) Manufacturer's name and address.

NOTE 1.The graphic symbols given in YY/T 0466.1 can be used to meet the above requirements.

NOTE 2.ISO 7000 symbol 2725 can be used to indicate the presence of a substance of concern by replacing "XXX" with the abbreviation of the substance. If the substance of concern is not present, a cross is placed on the symbol.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 18 pages — is available in the English PDF.

Referenced standards

Similar standards

Editions of YY/T 1282

EditionTitleRevisionStatus
YY/T 1282-2022Intravenous catheter for single usecurrent editionCurrent
YY 1282-2016Intravenous catheter for single useprevious editionSuperseded

This page sells the current edition, YY/T 1282-2022. Earlier editions are listed for reference only.

How to Download YY/T 1282-2022

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
18 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 1282-2022

$300.00