YY/T 1193-2011Follicle stimulating hormone (FSH) quantitative immunoassay kit (chemiluminescent immunoassay) (English PDF)
促卵泡生成激素(FSH)定量测定试剂盒(化学发光免疫分析法)
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 1193-2011 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
December 31, 2011
Implementation date
June 1, 2013
Scope
YY/T 1193-2011 is the English-translated version of 促卵泡生成激素(FSH)定量测定试剂盒(化学发光免疫分析法).
YY/T 1193-2011 is the Chinese pharmaceutical industry standard for the kit that measures follicle stimulating hormone quantitatively by chemiluminescent immunoassay.
It specifies the requirements, the test methods, the marking, the labelling, the instruction manual and the rules for packaging, transport and storage, covering the appearance, the traceability of the calibrator, the accuracy, the linear range, the detection limit, the precision, the hook effect and the stability of the reagent. Issued on 31 December 2011, it has applied since 1 June 2013.
Found what you need?
Document preview — YY/T 1193-2011
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Classification4
- 4 Requirements5
- 5 Test methods6
- 6 Marks, labels and instructions9
- 7 Packaging, transportation and storage11
- Bibliography12
Foreword
This standard was drafted in accordance with the rules specified in GB/T 1.1-2009.
1 Scope
This standard specifies the product classification, requirements, test method, marks, labels, instructions, packaging, transportation and storage etc. of the follicle stimulating hormone (FSH) quantitative immunoassay kit (chemiluminescent immunoassay).
This standard is applicable to follicle stimulating hormone (FSH) kit [hereinafter referred as "FSH kit"] of quantitative detection in human blood matrix or other body fluid components, based on the principle of chemiluminescent immunoassay. It includes the enzymatic and non-enzymatic chemiluminescent immunoassay detection kit in the carriers of micro-plates, pipes, magnetic particles, micro-beads and plastic beads. This standard does not apply to.
a) The calibrator and quality control product of follicle stimulating hormone intended for separate sale;
b) Biochip that uses the principle of chemiluminescent immunoassay.
2 Normative references
The articles contained in the following documents have become part of this document when they are quoted herein. For the dated documents so quoted, all the modifications (Including all corrections) or revisions made thereafter shall be applicable to this document.
GB/T 21415-2008 In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values assigned to calibrators and control materials
3 Classification
FSH kit can be divided into enzymatic and non-enzymatic chemiluminescent immunoassay kits according to different chemiluminescent principles. It can be divided into different kinds of chemiluminescent immunoassay kits taking micro-plates, tubes, magnetic particles, micro-beads and plastic beads as carrier according to the various solid phase carriers. According to different operation processes, it can be divided into manual operation method and instrument-automatic operation method. coefficient r of the kit, within the linear range specified by the manufacturing enterprise, shall not be less than 0.990.
4 Requirements
4.7 Repeatability
Use the samples with at least 2 concentration levels. Respectively repeat the detection for 10 times. The coefficient of variation (CV) shall not be more than 8.0% (instrument-automatic operation method) or not more than 12.0% (manual operation method).
4.8 Within-batch difference
Use 3 batches of kits to detect the same sample. The coefficient of variation (CV) among the 3 kits shall not be more than 15.0%.
4.9 Stability
The following methods can be used for verification.
a) Stability within the validity. The manufacturing enterprise shall specify the validity of the kit. Take the kit that has been expired to detect the accuracy, minimum detection limit, linearity and repeatability. It shall meet the requirements of 4.3, 4.4, 4.6 and 4.7;
b) Thermal stability test. The kit within the period of validity is placed at 37°C for 3d.
Test the accuracy, minimum detection limit, linearity and repeatability. It shall meet the requirements of 4.3, 4.4, 4.6 and 4.7.
Note 1. Thermal stability test cannot be used to derive the validity of products, unless it uses the derivation formula that is established and based on a large number of stability study data;
Note 2. Any arbitrary combination of a) and b) can be selected according to the product characteristics, but the selected method shall be capable of verifying the stability of the product, so as to ensure that the performance of the product can comply with the standard requirements within valid period.
5 Test methods
5.1 Appearance
Use the corrected-visual-acuity to visually inspect, under the natural light. It shall meet the requirements of 4.1.
5.2 Traceability
Traceability information provided by the manufacturing enterprise shall meet the requirements of 4.2.
e) Product batch number;
f) Expiration date; g) Storage conditions.
6 Marks, labels and instructions
6.2 Marks and labels of each component package of kit
It shall contain at least the following contents.
a) Product names and packaging specifications;
b) Name or trademarks of manufacturers;
c) Product batch number;
d) Expiration date; e) Storage conditions.
6.3 Instructions of kit
It shall include at least the following contents.
a) Product name;
b) Packaging specifications;
c) Intended use;
d) Detection principle;
e) Main compositions;
f) Storage conditions and expiration date;
g) Applicable instrument;
h) Sample requirement;
i) Detection method;
j) Reference value (Reference range);
k) Explanation of the detection result;
l) The limitations of detection methods;
m) Product performance index;
Bibliography 1. YY/T 0316-2008 Medical devices - Application of risk management to medical devices (ISO 14971. 2007, IDT)
2. YY/T 0466.1-2009 Devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements (ISO 15223.1-2009, IDT)
4. GB/T 9969-2008 General principles for preparation of instructions for use of industrial products […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Similar standards
How to Download YY/T 1193-2011
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 21415-2008 — In vitro diagnostic medical devices - Measurement of quantities in biological samples - Metrological traceability of values assigned to calibrators and control materials
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0479-2026 — Medical diagnostic X-ray tube assemblies and X-ray source assemblies - Determination of maximum symmetric radiation field
Secure payment via Stripe
Payments accepted
YY/T 1193-2011
$85.00