Scope

YY/T 1074-2002 is the English-translated version of 外科植入物 不锈钢产品点蚀电位.

YY/T 1074-2002 is the Chinese pharmaceutical industry standard that gives the method for measuring the pitting corrosion potential of stainless steel surgical implants, the potential at which the passive film breaks down and pitting begins.

It defines the measurement, then sets the requirements for the test instruments and the measuring system, the preparation of the specimen, the test conditions and the steps to follow, and states how the result is expressed and what the test report must contain.

The measurement is made by the potentiodynamic method and is used to compare the corrosion resistance of implant steels. It replaces YY 91074-1999, was issued on 24 September 2002 and has applied since 1 April 2003.

Found what you need?

Document preview — YY/T 1074-2002

National Standard of the People's Republic of China

Replacing
YY 91074-1999

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative References4
  • 3 Terms4
  • 4 Test instruments and equipment5
  • 5 Preparation of sample5
  • 6 Test conditions and steps6
  • 7 Test results7
  • 8 Test report7
  • Appendix A8
  • Appendix B9

Foreword

This standard adopts the test solution and test temperature provisions of the simulated human physiological environment that are specified in No. 405, 406 "New Working Item" proposed by the International Organization for Standardization (ISO) Technical Committee 150 on October 24, 1997.For the test instruments and equipment, sample preparation, and test steps, reference is made to some provisions in GB/T 17899-1999 "Method of Pitting Potential Measurement for Stainless Steels". Appendix A and B of this standard are normative.

YY 91074-1999 (ZB C35 005-1985) "Corrosion Test Methods for Metallic Bone Plates and Bone Screws" shall be abolished, since the issuing date of this standard.

This standard was proposed and approved by China Food and Drug Administration.

1 Scope

This standard specifies the definitions of measuring method for pitting corrosion potential on stainless surgical products; and the requirements of the measuring system's test instruments and equipment, sample preparation, test conditions and steps, test result, and test report.

This standard is applicable to the measurement that adopts potentiodynamic method to measure the pitting corrosion potential of stainless surgical implant products in the simulated human physiological environment.

2 Normative References

The articles contained in the following documents have become part of this standard when they are quoted herein. For the dated documents so quoted, all the modifications (excluding corrections) or revisions made thereafter shall not be applicable to this standard. For the undated documents so quoted, the latest editions shall be applicable to this standard.

GB/T 1266-1986 Chemical reagents - Sodium chloride

3 Terms

For the purpose of this standard, the following terms and definitions shall apply.

3.1 Pitting corrosion

Pit-shaped corrosion that extends from the surface to the internal part of metal which may be developed into cavity.

3.2 Free corrosion potential (Ek)

It refers to the electrode potential of the designated electrolyte solution when no net current flows into or out of the research metal surface. cutting method to cut.

5 Preparation of sample

5.3 Selection of test surface. For plate products, select rolling surface. For rod or wire products, select drawing surface. The test surface shall be away from edge, sharp corner, hole, groove, thread, mark, and welding spot. If there is no flat surface, select smooth spherical surface or cylinder surface.

5.4 Coating of sample. The surface of sample shall be wiped and cleaned by alcohol and acetone. After drying process, use insulating sealant (704 silicone rubber, epoxy resin and vinyl resin) to coat or inset the sample, so as to make the final exposed test surface to be about 1 cm2.If the sample is too small, the test surface can be less than 1 cm2.No crevice or pit is allowed on the insulated coating surface.

6 Test conditions and steps

6.1 Test conditions

6.1.1 Test solution. 9 g/L sodium oxide.

Dissolve 9 g of analytically pure sodium chloride that conforms to provisions of GB/T 1266 into 1 000 ml of distilled water or deionized water.

6.2 Test steps

6.2.1 Inject appropriate test solution into a clean electrolytic tank; then place the electrolytic tank into thermostatic water bath (temperature controlled at 37°C±1°C).

6.2.2 Install the working electrode (sample), saturated calomel electrode, and platinum electrode in the test solution of electrolytic tank, according to the diagram the electrochemical measurement system diagram (Appendix B), and then fix them.

6.2.3 Fully submerge the working electrode (sample), reference electrode, and auxiliary electrode into test solution. The upper edge of insulating part shall be at least 1cm higher than the test solution. The top end of luggin capillary of reference electrode shall be 1 to 2 mm away from the test surface of working electrode.

6.2.4 Connect the wire according to operation instructions. Put the switch and button at the required positions. Input the relevant parameters (select constant potential control, and record E-logi anode polarization curve).

6.2.5 After 10min, record the free corrosion potential. Start from the free corrosion potential, at 20mV/min potential scanning speed, forward, continuously and automatically adjust the potential, so as to make sample to be anodic polarization. When the anodic current density reaches 500µA/cm2~1 000µA/cm2, end the test.

Appendix A (Normative)

Supplementary explanation A.1 This standard is also applicable to the measurement by adopting manual step-method.

The potential variation speed is to keep constant potential for 1 min per 20 mV, or for 2 min per 40mV. List out the potential and current values; draw the E-logi anode polarization curve.

A.2 When the pitting corrosion potential cannot be determined, check whether the sample has crevice corrosion or coating is properly sealed. If there is crevice corrosion, the corrosion potential value at this point shall be omitted and void; and test shall be re-conducted.

A.3 The pitting corrosion potential measured by this standard only indicates the anti-corrosion level of the surface of the product, and represent the quality of the product and the processing level of surface treatment; it does not replace the random defect test of the product surface.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.

Referenced standards

Similar standards

Editions of YY/T 1074

EditionTitleRevisionStatus
YY/T 1074-2002Implants for surgery--Measuring method for pitting corrosion potential on stainless productscurrent editionCurrent
YY 91074-1999Implants for surgery--Measuring method for pitting corrosion potential on stainless productsprevious editionSuperseded

This page sells the current edition, YY/T 1074-2002. Earlier editions are listed for reference only.

How to Download YY/T 1074-2002

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
15 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 1074-2002

$85.00