YY/T 0927-2014Guidance of determination for di(2-ethylhexyl) phthalate(DEHP) released from PVC medical devices (English PDF)
聚氯乙烯医疗器械中邻苯二甲酸二(2-乙基己基)酯(DEHP)溶出量测定指南
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0927-2014 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
June 17, 2014
Implementation date
July 1, 2015
Scope
YY/T 0927-2014 is the English-translated version of 聚氯乙烯医疗器械中邻苯二甲酸二(2-乙基己基)酯(DEHP)溶出量测定指南.
YY/T 0927-2014 is the Chinese pharmaceutical industry standard that gives the chemical method for determining how much of the plasticizer DEHP is released from a medical device made of PVC when the device is in contact with the liquid it is used with.
It sets out the simulating liquid and the conditions and duration of the contact, the extraction and the analysis of the liquid, and the calculation and reporting of the amount released. Issued on 17 June 2014, it has applied since 1 July 2015.
Found what you need?
Document preview — YY/T 0927-2014
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- Introduction4
- 1 Scope6
- 2 Normative references6
- 3 General principles6
- 4 Principles for extract solution preparation6
- 5 Determination of DEHP release by gas chromatography / mass spectrometer method (GC/MS)8
- 6 Determination of DEHP release by UV method12
Foreword
This standard was drafted in accordance with the rules given in GB/T 1.1-2009.
1 Scope
This standard provides the chemical analysis methods for release of plasticizer DEHP after medical device which takes PVC as raw material is contacting with liquid of clinical use, under simulation of actual conditions of use.
2 Normative references
The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.
GB/T 6682 Water for analytical laboratory use - Specification and test methods
3 General principles
3.1 Unless otherwise specified, room temperature in this standard shall be 10°C~30°C.
3.2 Test water in this standard shall meet requirements in GB/T 6682.
3.3 Unless otherwise specified, container used in this standard shall be borosilicate glass container.
4 Principles for extract solution preparation
4.2 Preparation of extract solution of device in contact with blood or blood components
4.1 General principle
The preparation of DEHP extract solution in PVC medical devices shall consider the clinical usage of those devices, then choose appropriate method for extract solution preparation (e.g., extract solvent, time, temperature and mode of action).
4.3 Preparation of extract solution which takes sodium chloride injection
and/or glucose injection diluted lipid drugs as extract solvent Clinically, sodium chloride injection and glucose injection are the most commonly used drug infusion carriers.
4.4 Preparation of extract solution which takes other extract solvents
When devices are expected to contact with other mediums in clinical use excepts the ones mentioned in 4.2 and 4.3, for example, nutrients or other body fluids, it shall use the demonstrated and worst practical infusion liquid (or other alternative liquids) which represents clinical use, as well as extraction manner and conditions to conduct the extraction. Then take some extract solution as test solution.
5 Determination
of DEHP release by gas chromatography / mass spectrometer method (GC/MS)
5.1 Principle
Use total ion chromatogram (TIC) scan mode of gas chromatography / mass spectrometer method (GC/MS) to compare DEHP in sample solution, so as to determine the nature, according to DEHP retention time, characteristic ion and abundance of characteristic ion in DEHP standard solution.
5.2 Reagents
5.2.1 Ethanol. analytically pure
5.3 Apparatus
5.3.1 Gas chromatograph-mass spectrometer (GC/MS)
5.3.2 Vacuum drying oven
5.4 Working standard solution
5.4.1 Take about 20 mg of standard DEHP. Accurately weigh (to 0.1 mg). Use hexane to dilute it to 10mL, and standard stock solution of which concentration is about 2000 µg/mL shall be obtained (store it in the fridge at 2°C~10°C; 1 month validity).
5.5 Pre-treatment of sample extract solution
5.5.1 Pre-treatment of blood-contacting-device extract solution
5.5.2 Extract solution which takes sodium chloride injection and/or
glucose injection diluted lipid drugs as extract solvent Take certain volume of extract solution prepared in 4.3 to dry in vacuum at 50°C.
After dying, cool it to room temperature. Then add hexane of equivalent volume.
Use vortex mixer to scroll and dissolve it for about 1 min. Take supernatant liquid as inspection solution.
5.6 GC / MS conditions (recommendatory)
Chromatographic column. HP-5MS silica capillary column 30m × 0.25mm × 0.25µm.
5.7 Determination and result analysis
5.7.1 Qualitative analysis
Determine DEHP standard solution, so as to confirm the retention time of DEHP chromatographic peak under current conditions of temperament (unit is minute; the example is shown in Figure 1), DEHP characteristic ions (including m/z=70, 83, 104, 112, 149, 167 and 279), and abundance of characteristic ion (the example is shown in Figure 2). The retention time of chromatographic peak detected in sample solution (the example is shown in Figure 3) is same with the retention time of chromatographic peak of standard solution.
5.7.2 Quantitative analysis
Use external standard method. Quantitative selective ion of selective DEHP is […]
6 Determination of DEHP release by UV method
6.2 Preparation of working standard solution
6.2.1 Use liquid specific gravity balance (or device of equivalent accuracy) to prepare ethanol solution of which density is 0.8050 g/mL~0.8123 g/mL (at 20°C).
6.2.2 Take about 1 g of standard DEHP (make it accurate to 0.1 mg). Use ethanol solution (density is 0.8050 g/mL~0.8123 g/mL) to dilute it to 100 mL.
And standard stock solution of which concentration is 10000 µg/mL shall be obtained.
6.3 Determination and result calculation
6.3.1 At 272 nm, use extract solvent (refer to 4.2.1) as reference solution to measure absorbance values of standard solution, and obtain working curve and regression equation. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
Similar standards
How to Download YY/T 0927-2014
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 6682-1992 — Water for analytical laboratory use--Specification and test methods
GB/T 6682-2008 — Water for analytical laboratory use - Specification and test methods
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0927-2014
$125.00