YY/T 0894-2013Medical electrical equipment - Dosimetric instruments as used in brachytherapy - Instruments based on well-type ionization chambers (English PDF)
医用电气设备近距离放射治疗用剂量仪器基于井型电离室的仪器
This is a limited preview
Buy now to download the full PDF (16 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0894-2013 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
October 21, 2013
Implementation date
October 1, 2014
Scope
YY/T 0894-2013 is the English-translated version of 医用电气设备近距离放射治疗用剂量仪器基于井型电离室的仪器.
YY/T 0894-2013 is the Chinese standard "Medical electrical equipment - Dosimetric instruments as used in brachytherapy - Instruments based on well-type ionization chambers".
Its scope clause reads: This standard specifies the well type ionization chamber measurements and ancillary components (see Chapter 3 defined) performance and certain structural requirements for the determination of a The amount of, for example. after a given type of appropriate calibration source for brachytherapy photon radiation field intensity in the air kerma rate Degrees or reference air kerma rate, or photons and beta radiation field at a depth of water absorbed dose.
The criteria include quantitative methods adapted brachytherapy sources. This amount may be air kerma rate or intensity at 1m from the source Reference air kerma rate, or at a certain depth (for example.
2mm or 5mm) absorbed dose to water. The measurement of the amount by Various existing well type ionization chamber or system implementation. This standard applies to include and beta-ray photons, including the intended use of low-dose, high Dose rate brachytherapy intravascular measurement products.
This standard does not apply to medical applications of nuclear instrumentation. Limited to the application of this In the well type ionization chamber as a detector of the instrument.
The instrument uses a measuring chamber (inserted into a body cavity) or organizations radioactive output (insert human tissue) sealed sources.
The purpose of this standard are. a) the establishment of the well type chamber system satisfactory performance requirements, and b) providing a standardized method for determining compliance with the performance.
This standard does not involve security Well type ionization chamber system. Well type ionization chamber system are not covered by this standard environment for patients.
Well type Electrical ionization chamber system security, see IEC 61010-1. Electrometer measurement system, see IEC 60731.
Found what you need?
Document preview — YY/T 0894-2013
National Standard of the People's Republic of China
- ICS
- 11.040.50
- Classification
- C 43
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Introduction
- 1 Scope
- 2 Normative references
- 3 Terms and definitions
- 4 General requirements
- 4.1 Performance Requirements
- 4.2 measuring assembly
- 4.3 of 5 types of source
- 4.4 was measured
- 4.5 Reference conditions and standard test conditions
- 4.6 General test conditions
- 4.8 Test 7 member
- 5.1 Insert the source location and repeatability
- 5.2 effective length
- 5.4 Settling Time
- 5.6 Stability
- 6 affected by the amount of the impact of changes in the limits
- 6.1 Overview
- 6.2 ionization chamber - Composite loss
- 6.3 Operating voltage
- 6.4 air pressure
- 6.5 air pressure changes and the radiation detector equilibration time
- 6.6 Temperature and humidity
- 6.7 length of the response
- 6.8 Electromagnetic Immunity
- 7 logo
- 7.1 Well type ionization chamber assembly
- 7.2 measuring assembly
- 8 random file
- 8.1 Overview
- 8.2 Use of the instrument
- 8.3 Document
- References
1 Scope
This standard specifies the well type ionization chamber measurements and ancillary components (see Chapter 3 defined) performance and certain structural requirements for the determination of a
The amount of, for example. after a given type of appropriate calibration source for brachytherapy photon radiation field intensity in the air kerma rate
Degrees or reference air kerma rate, or photons and beta radiation field at a depth of water absorbed dose.
The criteria include quantitative methods adapted brachytherapy sources. This amount may be air kerma rate or intensity at 1m from the source
Reference air kerma rate, or at a certain depth (for example. 2mm or 5mm) absorbed dose to water. The measurement of the amount by
Various existing well type ionization chamber or system implementation. This standard applies to include and beta-ray photons, including the intended use of low-dose, high
Dose rate brachytherapy intravascular measurement products. This standard does not apply to medical applications of nuclear instrumentation. Limited to the application of this
In the well type ionization chamber as a detector of the instrument.
The instrument uses a measuring chamber (inserted into a body cavity) or organizations radioactive output (insert human tissue) sealed sources.
The purpose of this standard are.
a) the establishment of the well type chamber system satisfactory performance requirements, and b) providing a standardized method for determining compliance with the performance.
This standard does not involve security Well type ionization chamber system. Well type ionization chamber system are not covered by this standard environment for patients. Well type
Electrical ionization chamber system security, see IEC 61010-1. Electrometer measurement system, see IEC 60731.
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein
Member. For undated references, the latest edition (including any amendments) applies to this document.
GB/T 2900.81-2008 Electrotechnical terminology Nuclear Instruments Physical phenomena and basic concepts (IEC 60050-393.2003, IDT)
GB/T 19629-2005 Medical electrical equipment - X-ray diagnostic imaging used in the ionization chamber and (or) semiconductor detector dose
Meter (IEC 61674.1997, IDT)
GB/T 20012-2005 Medical electrical equipment dose area product meter (IEC 60580.2003, IDT)
IEC 60417 Graphical symbols for electrical equipment (Graphicalsymbolsforuseonequipment)
IEC 60601-1.2005 Medical electrical equipment. Part 1. Basic safety and essential performance General requirements (Medicalelectrical equipment-Part 1. Generalrequirementsforbasicsafetyandessentialperformance)
IEC 60731.1997 Medical electrical equipment used in radiotherapy ionization chamber dosimeter (Medicalelectrical equipment-Dosimeterswithionizationchambersasusedinradiotherapy)
IEC /T R60788.2004 Medical electrical equipment - defined indexes (Medicalelectricalequipment-Glossaryof definedterms)
IEC 61187 Electrical and electronic measuring equipment - Documentation (Electricalandelectronicmeasuringequipment-Doc- umentation)
ISO /IEC Guide 99 international basic and common terminology Metrology (VIM) […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 16 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 19629-2005Medical electrical equipment - Dosimeters with ionization chambers and/or semiconductor detectors as used in X-ray diagnostic imaging
- GB/T 20012-2005Medical electrical equipment - Dose area product meters
GB/T 2900.81-2008 · IEC 60417 · IEC 60601-1:2005 · IEC 60731:1997 · IEC/T R60788.2004 · IEC 61187 · ISO/IEC Guide 99
Similar standards
How to Download YY/T 0894-2013
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 19629-2005 — Medical electrical equipment - Dosimeters with ionization chambers and/or semiconductor detectors as used in X-ray diagnostic imaging
GB/T 20012-2005 — Medical electrical equipment - Dose area product meters
YY/T 0297-2026 — General requirements for clinical investigation of medical devices