YY/T 0794-2010Particular specifications for computed radiography device (English PDF)
X射线摄影用影像板成像装置专用技术要求
This is a limited preview
Buy now to download the full PDF (18 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0794-2010 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
December 27, 2010
Implementation date
June 1, 2012
Scope
YY/T 0794-2010 is the English-translated version of X射线摄影用影像板成像装置专用技术要求.
YY/T 0794-2010 is the Chinese standard "Particular specifications for computed radiography device". Its scope clause reads: This standard specifies the terms and definitions, composition, requirements, test methods of computed radiography device (hereinafter referred to as CR).
This standard applies to CR. The device uses an image plate that forms a "latent image" after X-ray irradiation; forms a digital X-ray image by scanning the "latent image" on the image plate.
CR is used to obtain digital diagnostic images after X-ray device photography, excluding X-ray photography device. Its clauses include terms and definitions; composition; requirements.
Found what you need?
Document preview — YY/T 0794-2010
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Terms and definitions4
- 4 Composition6
- 5 Requirements6
- 6 Test method9
- Appendix A (Informative) Comprehensive test scale14
- Appendix B (Informative) Threshold contrast test phantom16
- Appendix C (Informative) Justification28
1 Scope
This standard specifies the terms and definitions, composition, requirements, test methods of computed radiography device (hereinafter referred to as CR).
This standard applies to CR. The device uses an image plate that forms a
"latent image" after X-ray irradiation; forms a digital X-ray image by scanning the "latent image" on the image plate. CR is used to obtain digital diagnostic images after X-ray device photography, excluding X-ray photography device.
2 Normative references
The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) are applicable to this standard.
GB 4793.1-2007 Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
GB 7247.1-2001 Safety of laser products - Part 1: Equipment classification, requirements and users guide
YY/T 0291-2007 Environmental requirements and test methods for medical x-ray equipment
YY/T 0796-2010 Medical electrical device - Exposure index of digital X-ray imaging system - Part 1: Definition and requirements of general X-ray photography
DICOM 3.0 Medical digital imaging and communication standard (third edition)
IEC 60788 Medical electrical device - Glossary of terms and definitions
3 Terms and definitions
The terms and definitions as defined in IEC 60788 as well as the following terms and definitions apply to this document.
4 Composition
CR is mainly composed of imaging plate, scanning device, control processing device.
5 Requirements
5.1 Working conditions
5.1.1 Environmental conditions
Unless otherwise specified in the product standard, the working environment conditions of CR shall meet:
a) Ambient temperature: 10 °C ~ 40 °C;
b) Relative humidity: 30% ~ 75%;
c) Atmospheric pressure: 700 hPa ~ 1060 hPa.
5.1.2 Power supply conditions
The working power supply conditions of CR shall meet:
a) For the power supply voltage and the number of phases specified in the product standard, the fluctuation of the grid voltage shall not exceed ±10% of the nominal value;
b) Power frequency: 50 Hz ± 1 Hz (products for export are specified by the manufacturer);
c) The power capacity as specified by the product standard.
5.2 Effective imaging area
The product standard shall specify the maximum size and deviation of the effective imaging area of the IP plate in the X and Y directions; the deviation shall not exceed ±5% of the nominal value.
5.3 Bit-depth
The product standard shall specify the bit-depth when the image is acquired; it shall not be less than 12 bits/pixel.
5.12 Security of management information
There shall be software measures taken to ensure the privacy and reliability of patient data.
5.13 Information management
There shall be software measures to manage patient information, at least with the functions of registration, modification, deletion, query, statistics, backup management.
5.14 Film printing
It has the function of transferring images to a film printer.
5.15 Medical digital imaging and communication standards
If it conforms to the DICOM 3.0 standard or other network communication standards, it shall provide a compliance statement in the accompanying documents.
5.16 Noise
The noise generated during operation under working conditions shall not exceed 70 dB (A weighting network) (not including non-continuous and non- periodic noise within 3 s).
5.17 Appearance
The appearance shall be neat and beautiful; the surface shall be smooth and clean; color shall be uniform. There shall be no defects such as scars, cracks, etc.
5.18 Environmental test
It shall meet the requirements of YY/T 0291-2007 (except IP plate). In the initial, intermediate or final testing, the device shall be able to be used normally without abnormal phenomena.
5.19 Safety
It shall meet the requirements of GB 4793.1-2007 and GB 7247.1-2001.
6.5 Number of IP plates processed per hour
The test steps are as follows:
a) Select five IP plates of the same specification;
b) Start timekeeping when the first IP plate is inserted into the card slot of the CR scanning device, until the last IP plate is withdrawn from the CR.
The operation time shall be shortened as much as possible for the part of the CR that requires manual operation;
c) Measure the total time required to complete the entire process of the IP plate’s processing cycle for 5 consecutive times; calculate the number of
IP plates processed in 1 h; the result takes an integer.
6.6 Threshold contrast
Use the test method specified in the product standard for testing, or test according to the following methods as recommended by this standard, the steps are as follows:
a) Use the threshold contrast phantom in Appendix B and select appropriate loading conditions, to obtain the clearest phantom image;
b) Refer to Appendix B, to obtain the reciprocal of the image quality value through calculation;
c) Repeat steps a) ~ b) to measure 3 times; take the average value;
d) The measurement conditions, effective imaging area of the IP plate, pixel size, bit-depth shall be recorded together with the results.
6.7 Spatial resolution
Load according to the loading conditions as specified in the product standard or automatic exposure control; meanwhile adjust the brightness and contrast to […]
6.12 Security of management information
Verify by checking the device.
6.13 Information management
Verify by checking the device.
6.14 Film printing
Verify by inspection.
6.15 Medical digital imaging and communication standards
Verify by checking its declaration of conformity.
6.16 Noise
The sound level meter’s probe is 1 m away from the surface of the CR system and 1.5 m from the floor. The sound level meter "A" class weighting network is used for measurement; use the maximum noise value for calculation.
6.17 Appearance
Observe visually.
6.18 Environmental test
It is performed according to the requirements of YY/T 0291-2007.
6.19 Security
The inspection shall be carried out according to the requirements of GB 4793.1-
2007 and GB 7247.1-2001.
C.1 Description about 3.1, 3.2, 3.3
Photostimulable Luminescence (PSL) is the basis of CR imaging. This feature makes CR different from other medical imaging device.
The imaging layer in the IP plate is a layer of barium halide fluoride crystals containing divalent europium ions. The compound in the crystal layer is irradiated by X-rays; the fluorescent substance absorbs the incident X-rays and stores its energy, forming a latent image and storing it in the crystal, which is a simulated image (this is the first excitation); then the laser beam scans the fluorescent screen with latent image (second excitation); the information stored on the screen is radiated from the converted light signal. The wavelength of the helium-neon laser beam used is 670 nm ~ 690 nm. The surface of the IP plate is scanned sequentially in the direction perpendicular to the IP plate. The phosphor refracted from the IP plate is excited by the secondary excitation (called Photostimulable Luminescence, PSL), producing fluorescence. The intensity of the fluorescence is precisely proportional to the latent image energy at the first excitation. The fluorescence is collected and guided along the high- efficiency light guide filmed by laser scanning; enters the photomultiplier tube;
is amplified by the A/D converter into a digital signal and stored on the computer.
The computer can further process the stored images and display them. The signal on the storage screen can be erased by strong light for the next use.
CR does not include X-ray machines. Its main purpose is to replace the traditional X-ray screen-film system, to realize the digital processing and storage of diagnostic images.
C.2 Description of 3.4 and 3.5
For the specifications of the IP plate, its effective imaging area, bit-depth, pixel size determines the image quality of X-ray digital images from the hardware.
C.3 Descriptions of 5.6
Image quality figure (IQF) is a method of quantifying images for threshold contrast evaluation. The following methods can be used for threshold contrast evaluation:
a) Draw a curve that passes through the threshold area, which is called the contrast-detail curve.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 18 pages — is available in the English PDF.
Similar standards
How to Download YY/T 0794-2010
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
YY/T 0463-2026 — Dentistry - Refractory investment and die materials
YY/T 0479-2026 — Medical diagnostic X-ray tube assemblies and X-ray source assemblies - Determination of maximum symmetric radiation field
YY/T 0794-2010
$295.00