YY/T 0741-2018Particular specifications for digital X-ray radiography system (English PDF)
数字化摄影X射线机专用技术条件
This is a limited preview
Buy now to download the full PDF (27 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0741-2018 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
September 28, 2018
Implementation date
October 1, 2019
Scope
YY/T 0741-2018 is the English-translated version of 数字化摄影X射线机专用技术条件.
YY/T 0741-2018 is the Chinese pharmaceutical industry standard for the digital radiography system.
It gives the terms and definitions, the composition of the system, the requirements and the test methods, covering the imaging performance of the detector and of the whole chain - resolution, contrast, noise and dynamic range - the accuracy of the loading factors and of the automatic exposure control, the dose indication, the image processing and the display and storage of the images, together with the accompanying documents. Issued on 28 September 2018, it has applied since 1 October 2019.
Found what you need?
Document preview — YY/T 0741-2018
National Standard of the People's Republic of China
- Replacing
- YY/T 0741-2009
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope5
- 2 Normative References5
- 3 Terms and Definitions6
- 4 System Composition8
- 5 Requirements8
- 6 Test Methods15
- Appendix A (Normative) Test Arrangement23
- Appendix B (Informative) Test Phantom24
- Appendix C (Informative) Drafting Explanation of Some Clauses27
Foreword
This Standard was drafted as per the rules specified in GB/T 1.1-2009.
This Standard replaced YY/T 0741-2009 General Specifications for Digital Medical X- Ray Radiography System. Compared with YY/T 0741-2009, this Standard has the major technical parameter changes as follows.
--- Modify the standard name from General Specifications for Digital Medical X-Ray Radiography System to Particular Specifications for Digital X-Ray Radiography System;
--- Modify the description of the Clause 4 System Composition;
--- Modify 5.3.6 requirements and test methods for automatic exposure control;
--- Modify 5.4 limits and test methods of some subclauses of imaging performance;
--- Modify 5.5 description of some subclause requirements of mechanical device performance;
--- Modify 5.6.2 information management, 5.6.3 requirements and test methods of imaging time;
--- Add 5.3.8 dose area product indication, 5.3.9 requirements and test methods of pediatric photography;
--- Add 5.4.3 dynamic range, 5.4.8 (long bone) image stitching performance, 5.4.9 detector calibration and stability test, 5.4.10 requirements and test methods of quantum detection efficiency;
--- Delete 5.6.4 requirements and test methods of preview time;
--- Add description of the integrated test card (see Appendix B).
Please note some contents of this document may involve patents. The issuing organization of this document shall not assume the responsibility for identifying these patents.
1 Scope
This Standard specifies the terms and definitions, system composition, requirements and test methods for digital X-ray radiography system (hereinafter referred to as DR system).
This Standard is applicable to the DR system for general X-ray photography. It includes but not limited to DR systems that use line scan or area scan detectors, such as.
--- DR system using flat panel detector (FPD);
--- DR system using area array CCD detector;
--- DR system using line array scanning CCD detector;
--- DR system using CMOS detector, etc.
Corresponding to the DR system using more than one digital X-ray image detector, this Standard is applicable to each digital X-ray image detector and the X-ray generating device used in its imaging.
This Standard is not applicable to system using X-ray image intensifier, system using imaging device of image plate for X-ray photography, mammary X-ray equipment, dental X-ray equipment, computed tomography equipment, mobile DR system.
2 Normative References
The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) are applicable to this document.
GB 9706.1 Medical Electrical Equipment - Part 1: General Requirements for Safety
GB 9706.3-2000 Medical Electrical Equipment - Part 2: Particular Requirements for the Safety of High-Voltage Generators of Diagnostic X-Ray Generators
GB 9706.11 Medical Electrical Equipment - Part 2: Particular Requirements for the Safety of X-Ray Source Assemblies and X-Ray Tube Assemblies for Medical Diagnosis
GB 9706.12-1997 Medical Electrical Equipment - Part 1: General Requirements for Safety - 3 Collateral Standard - General Requirements for Radiation Protection in Diagnostic X-Ray Equipment
3 Terms and Definitions
For the purposes of this document, the terms and definitions given in Normative References and the following apply.
3.1 Digital radiography system
It uses digital X-ray image detector technology to realize the X-ray photography; generally, it consists of X-ray generating device, digital X-ray imaging device, and ancillary equipment.
3.4 Imaging time
The time, from the exposure to the screen displaying, required for the normal image satisfying the diagnostic needs.
3.5 Information management
Software measures for managing the patient information, such as appointments, registrations, modifications, deletions, queries, statistics, backup management, etc.
3.6 Low contrast resolution
Under specified measurement conditions, the lowest contrast details between specified shape and area that can be distinguished from a uniform background condition, expressed in %.
3.10 Effect image area
Maximum effective area for detector imaging.
3.11 Artifact
A structure that is clearly visible on an image. It neither reflect the internal structure of the object, nor can it be explained by noise or system modulation transfer functions.
3.12 Erasure thoroughness
After the previous image signal is read, part of all of the previous image generated in the subsequent image due to incomplete signal removal.
4 System Composition
See Figure 1 for DR system composition.
5 Requirements
5.1 Working conditions
5.1.1 Environmental conditions
Unless otherwise specified, the working and environmental conditions of the DR system shall satisfy as follows.
5.1.2 Power supply conditions
The operating power supply conditions of the DR system shall satisfy as follows.
5.2 Electric power
5.2.1 Maximum output electric power
It shall be specified in the product technical requirements for the corresponding combination between voltage and current of X-ray tube with maximum output electric power.
5.3 Loading factors and control
5.3.1 X-ray tube voltage
The X-ray tube voltage shall meet the following requirements.
5.3.3 Loading time
If there is an indication for loading time, it shall meet the following requirements.
a) It shall be specified in product technical requirements for adjustment range and
5.3.5 Anti-overload
The DR system shall take the anti-overload measures, ensuring that the selection of loading factor shall not exceed the rated capacity of the X-ray tube. It shall satisfy the maximum loading factor combination given in the instruction manual.
5.3.6 Automatic exposure control
The DR system shall be equipped with Automatic Exposure Control (AEC) function.
The repeatability of the radiation output in the AEC mode shall satisfy the coefficient of variation for the air kerma measurement value no greater than 0.05.
5.4.4 Flat uniform
The manufacturer shall specify the maximum flat uniformity of DR, the used SID, and loading factor. Unless otherwise is specified by the manufacturer, it is specified by the image that the ratio between the standard deviation of the gray value on sampling point R and the mean value of the gray value on the sampling point Vm shall be no greater than 2.5%. Namely, formula (2).
5.6 Network and software
5.6.1 Network communication
It shall conform to the DICOM3.0 standard; the manufacturer shall provide a declaration of conformity for DICOM3.0 standard in accompanying file.
5.6.3 Imaging time
The imaging time shall be no more than 12s.
5.7 Plug and socket of high voltage cable
Unless otherwise specified, it shall meet the requirements for relevant type, basic dimension, marking, connection specified in GB/T 10151.
5.8 Appearance
The appearance of the DR system shall be neat and beautiful; the surface shall be smooth with uniform luster; free of scars, cracks, and the like defects.
5.9 Environmental test
It shall meet the requirement of YY/T 0291. The initial, intermediate or final test items shall include at least the contents of 5.4.1, 5.4.2, 5.4.3, 5.5.4.
6 Test Methods
6.1 Test conditions
6.1.1 Environmental conditions
It shall conform to the provisions of 5.1.1.
6.1.2 Power supply conditions
The test power supply conditions are as follows.
6.2 Electric power
6.2.1 Maximum output electric power
The DR system operating in an intermittent mode is loaded in the combination with the loading factors that cause the maximum output electric power; then observe whether there is an abnormality.
6.3 Loading factor and control
6.3.1 X-ray tube voltage
It shall be performed as per the following methods.
6.3.3 Loading time
It shall be performed as per the following methods.
6.3.4 Current time product
It shall be performed as per the following methods.
6.3.5 Anti-overload
It shall be performed as per the following methods.
Adjust the parameters of the DR system to the maximum load factor combination value specified in the instruction manual. When readjust any loading factor to the adjacent increase shift, the loading factor condition shall not be increased any more or the DR system is in the overload state. For the DR system with continuous adjustment of the X-ray tube voltage, when the operating voltage value of the anti-overload device is higher than the maximum X-ray tube voltage value (but not exceeding 5kV) specified in the instruction manual, it shall not be exposed and/or alarmed.
6.3.7 Correspondence between X-ray field and image receiving surface
It shall be performed as per the following methods.
6.3.8 Dose area product indication
It shall be performed as per the following methods.
Place the X-ray dosimeter detector in an appropriate position in the X-ray beam, such as on the image receiver surface or bed surface. Adjust the X-ray filed dimension to 15cm×15cm or other proper dimensions. Take photograph to dose area product that is displayed by the equipment is greater than 5µGy·m2; multiply the measured dose value by the ray filed area at the air kerma dosimeter detector. It can also be measured directly by the dose area product meter. Calculate the error between displayed value and the measured value.
6.3.9 Pediatric photography requirements
Actual operation and inspection.
6.4 Imaging performance
6.4.1 Spatial resolution
Test and arrange the multi-function test card shown in Figure B.1 according to Figure A.1.Adjust SID to be the normal clinical use distance; set the visual field of the image to be the maximum effect image area of the plate; select the typically clinical protocol specified by the manufacturer; manually set the X-ray tube voltage to be 75kV±7kV;
AEC automatic exposure. If required, adjust the window width and position; record the spatial resolution value and corresponding loading factor combination.
6.4.10 Quantum detection efficiency
Check the manufacturer's accompanying file.
6.5 Performance of mechanical device
6.5.1 Scope of mechanical motion
In practice, measure the range of rotation angle by the angle gauge; the minimum division scale of the angle gauge is no greater than 0.5°. The range of motion the longitudinal, lateral, and vertical directions is measured by a length gauge.
6.5.2 Length indication value
In practice, measure by the length gauge.
6.5.3 Angle indication value
In practice, measure by the angle gauge.
6.5.4 Braking
In practice, measure by a dynamometer.
6.5.5 Load bearing
The patient support device is in a horizontal state and in most unfavorable position at work; uniformly distribute 135kg load on the 168cm×37.5cm support surface for 1min; observe whether it can work normally.
For pedals and chairs, uniformly distribute 135kg load on the 0.1m² support surface for 1min; observe whether it can work normally.
6.7 Plug and socket of high voltage cable
In case when the manufacturer specified otherwise, inspection according to the method specified by the manufacturer; otherwise, inspect according to the relevant type, basic dimension, marking, connection specified in GB/T 10151.
6.8 Appearance
Visual examination.
6.9 Environmental test
The environmental test shall be performed as per the provisions of YY/T 0291.
6.10 Safety requirements
It shall be performed as per the provisions of GB 9706.1, GB 9706.3, GB 9706.11, GB 9706.12, GB 9706.14, GB 9706.15 and YY 0505.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 27 pages — is available in the English PDF.
Referenced standards
Normative references
- GB 9706.1Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- GB 9706.12-1997Medical electrical equipment Part 1: General requirements for safety 3. Collateral standard: General requirements for radiation protection in diagnostic X-ray equipment
GB 9706.3-2000 · GB 9706.11
Similar standards
Editions of YY/T 0741
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 0741-2018 | Particular specifications for digital X-ray radiography system | current edition | Current |
| YY/T 0741-2009 | Particular specifications for digital X-ray radiography system | previous edition | Superseded |
This page sells the current edition, YY/T 0741-2018. Earlier editions are listed for reference only.
How to Download YY/T 0741-2018
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 9706.1-2020 — Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
GB 9706.12-1997 — Medical electrical equipment Part 1: General requirements for safety 3. Collateral standard: General requirements for radiation protection in diagnostic X-ray equipment
YY/T 0297-2026 — General requirements for clinical investigation of medical devices
Secure payment via Stripe
Payments accepted
YY/T 0741-2018
$125.00