YY/T 0589-2016Electrolyte analyzer (English PDF)
电解质分析仪
This is a limited preview
Buy now to download the full PDF (23 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0589-2016 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
March 23, 2016
Implementation date
January 1, 2017
Scope
YY/T 0589-2016 is the English-translated version of 电解质分析仪.
YY/T 0589-2016 is the Chinese pharmaceutical industry standard for the electrolyte analyser, the instrument that measures sodium, potassium, chloride and related ions with ion-selective electrodes.
It gives the classification and the basic parameters, the requirements, the test methods, the marks, labelling and instruction manual and the rules for packaging, transport and storage, covering the measuring ranges, the accuracy and repeatability against control materials, the carry-over, the stability of the calibration, the electrodes and the alarms. Issued on 23 March 2016, it has applied since 1 January 2017.
Found what you need?
Document preview — YY/T 0589-2016
National Standard of the People's Republic of China
- Replacing
- YY/T 0589-2005
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope5
- 2 Normative references5
- 3 Classification and basic parameters6
- 4 Requirements7
- 5 Test method8
- 6 Signs, labeling, instruction manual12
- 7 Packaging, transportation and storage14
- Appendix A (Normative) Preparation method of fixed value mass control test solution15
1 Scope
This Standard specifies the classification and basic parameters, requirements, test methods, signs, labeling, instruction manual and packaging, transportation, and storage of electrolyte analyzers.
This Standard applies to electrolyte analyzers which use ion-selective electrodes as sensors (hereafter referred to as instruments); the instruments are suitable for human clinical electrolyte testing. The electrolyte module on the biochemical analyzer can refer to this Standard.
2 Normative references
The following documents are indispensable for the application of this document.
For dated references, only the dated version applies to this document. For undated references, the latest edition (including all amendments) applies to this document.
GB/T 191, Packaging - Pictorial marking for handling of goods
GB 4793.1, Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
GB/T 14710, Environmental requirement and test methods for medical electrical equipment
GB/T 18268.1, Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements
GB/T 18268.26, Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 26: Particular requirements - In vitro diagnostic (IVD) medical equipment
YY/T 0466.1, Medical devices - Symbols to be used with medical device labels, labelling and information to be supplied - Part 1: General requirements
YY 0648, Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment 4.2.4 Save and print the measurement data of the instrument.
4 Requirements
4.3 Appearance
The appearance shall meet the following requirements:
a) The whole machine shall be complete, without scratches and cracks;
b) The fastener connection shall be firm and reliable, without loosening;
c) There shall be no obvious defects on the surface of metal parts.
4.4 Safety
It shall meet the requirements of the applicable clauses in GB 4793.1 and YY 0648.
4.5 Environmental test
It shall meet the requirements of the applicable clauses in GB/T 14710.
4.6 Electromagnetic compatibility
It shall meet the requirements of the applicable clauses in GB/T 18268.1, GB/T 18268.26.
5 Test method
5.1 Test conditions
Normal working conditions of the instrument:
-- power supply: AC 220 V ± 22 V, 50 Hz ± 1 Hz; or other applicable power supply;
-- ambient temperature: 15 °C ~ 30 °C;
-- relative humidity: not more than 85%;
-- away from interference sources of strong electromagnetic fields.
Note: When the conditions in 5.1 are inconsistent with the manufacturer's product regulations, the product regulations shall prevail. The manufacturer shall state in the product standard.
5.2 Accuracy
For the sample, select a reference substance that has fixed value data based on serum. SD -- standard deviation;
-- the average value of the 10 continuously output measured values.
5.4 Linear
For the sample, select No. 1, No. 2, No. 3, No. 4 and No. 5 K+, Na+, Cl-fixed mass control test solution that is prepared in Table A.1 of Appendix A.
For the sample, select No. 1, No. 2, No. 3, No. 4 and No. 5 iCa2+, Li+ fixed mass control test solution that is prepared in Table A.2 of Appendix A.
Under normal working conditions of the instrument, first, perform the calibration according to the conventional test procedure; then, respectively measure each of the certain value of the mass control test solution continuously for 3 times in sequence; take the average value as the actual-measured value; for each analytical element, there is a group of measured value Y of different concentrations.
Between a group of actual-measured values of different concentrations of each analytical element, Yn, and a group of nominal values of different concentrations corresponding to it, X, firstly, use the linear regression analysis method to calculate the linear regression parameters, a (slope) and b (intercept), of each analytical element; derive the linear regression equation Y = a · X + b of each analytical element.
For each analytical element, substitute a known group of nominal values X of different concentrations into their corresponding linear regression equations to obtain a corresponding group of linear regression values Y.
Respectively calculate the absolute value of absolute deviation or relative deviation of the actual-measured value Yn relative to the linear regression value Y of different concentrations of each analytical element; calculate the correlation coefficient of linear regression |r|, which shall meet the requirements of 4.1.3.
5.5 Stability
For the sample, select No. 3 K+, Na+, Cl-fixed value mass control test solution that is prepared in Table A.1 of Appendix A.
For the sample, select No. 3 iCa2+, Li+ fixed value mass control test solution that is prepared in Table A.2 of Appendix A.
Under normal working conditions of the instrument, first, perform the calibration in accordance with the conventional test procedures; then, test the No. 3 fixed-value mass control test solution; perform the test, respectively, at 0 h, 4 h, and 8 h, and record the measured values of each analytical element; obtain three
Where: CHL -- Carrying pollution rate;
L1 ~ L4 -- the 1st ~ 4th measurement value of each group of low concentration solution;
H1 ~ H4 -- the 1st ~ 4th measurement value of each group of high concentration solution.
5.7 Function
Perform the function test according to the use method of the instruction manual and the operation prompt of the display, which shall comply with the provisions of 4.2.
5.8 Appearance
Visually inspect, which shall meet the requirements of 4.3.
5.9 Safety
Perform according to the test methods of the applicable clauses in GB 4793.1 and YY 0648.
5.10 Environmental test
Perform the test according to applicable test procedures and test methods of
GB/T 14710. During the high temperature storage, low temperature storage, damp heat storage and vibration collision tests, the electrode, pump tube and reagents can be withdrawn; the test results shall meet the requirements of 4.4.
5.11 Electromagnetic compatibility
Perform the test according to the applicable clauses in GB/T 18268.1 and GB/T 18268.26, which shall meet the requirements of 4.6.
6 Signs, labeling, instruction manual
6.1 Overview
The text of the signs, labeling and instruction manual must be in Chinese; other languages can be attached. The use of Chinese shall conform to the national language standards. The text, symbols, graphics, tables, numbers, photos, pictures, etc. of the instruction manual, labeling and packaging marks shall be accurate, clear and standardized. The symbols which are used in signs, labeling and instruction manual shall meet the requirements of YY/T 0466.1.
A.1 Equipment
A.1.1 Analytical balance (the maximum weighing is not more than 200 g; the accuracy is 0.1 mg) and weighing bottle with lid.
A.1.2 Volumetric flask (1 000 mL, grade A) and pipette (10 mL, grade A). A.1.3 Electric oven, beaker (100 mL) and dryer.
A.2 Reagent
A.2.1 Potassium chloride [KCl], guaranteed reagent (content >= 99.5%). A.2.2 Sodium chloride [NaCl], guaranteed reagent (content >= 99.5%). A.2.3 Sodium acetate [CH3COONa], guaranteed reagent (content >= 99.0%). A.2.4 Boric acid [H3BO3], guaranteed reagent (content >= 99.0%). A.2.5 Anhydrous lithium chloride [LiCl], guaranteed reagent (content >= 99.0%).
A.2.6 Anhydrous calcium chloride [CaCl2], guaranteed reagent (content >=
99.5%).
A.2.7 Sodium 3-(N-morpholino) propanesulfonate
[MOPSNa], guaranteed reagent (content >= 99.0%). A.2.8 3-(N-morpholino) propanesulfonic acid [MOPS], guaranteed reagent (content >= 99.0%).
A.3 Preparation requirements and methods
A.3.1 Preparation requirements
According to Table A.1, Table A.2, prepare 5-point fixed value mass control test solution.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 23 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 191Graphical symbols marking for handling and storage of packages
- GB 4793.1Safety requirements for electrical equipment for measurement,control,andlaboratory use-Part 1:General requirements
- GB/T 14710The environmental requirement and test methods for medical electrical equipment
- GB/T 18268.1Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements
- GB/T 18268.26Electrical equipment for measurement, control and laboratory use -- EMC requirements -- Part 26: Particular requirements -- In vitro diagnostic (IVD) medical equipment
- YY/T 0466.1Medical devices - Symbols to be used with information to be supplied by the manufacturer - Part 1: General requirements
Editions of YY/T 0589
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 0589-2016 | Electrolyte analyzer | current edition | Current |
| YY/T 0589-2005 | Electrolyte analyzer | previous edition | Superseded |
This page sells the current edition, YY/T 0589-2016. Earlier editions are listed for reference only.
How to Download YY/T 0589-2016
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB 4793.1-2007 — Safety requirements for electrical equipment for measurement,control,andlaboratory use-Part 1:General requirements
GB/T 14710-2009 — The environmental requirement and test methods for medical electrical equipment
GB/T 18268.1-2025 — Electrical equipment for measurement, control and laboratory use - EMC requirements - Part 1: General requirements