YY/T 0588-2017Flow cytometer (English PDF)

流式细胞仪

Preview PDF

This is a limited preview

Buy now to download the full PDF (19 pages)

Need quotation for your management?

Get a formal quotation for YY/T 0588-2017 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Recommended

Issue date

December 5, 2017

Implementation date

December 1, 2018

Scope

YY/T 0588-2017 is the English-translated version of 流式细胞仪.

YY/T 0588-2017 is the Chinese pharmaceutical industry standard for the flow cytometer, the instrument that counts and characterises cells one by one in a fluid stream by their scattered light and fluorescence.

It gives the terms and definitions, the technical requirements, the test methods and the rules for labels, marks, instructions for use, packaging, transport and storage.

Compared with the 2005 edition it renames fluorescence sensitivity as the detection limit of fluorescence and adds the limit for other lasers on the corresponding channel, renames forward scatter sensitivity as the forward scatter detection limit, reworks the instrument resolution requirement by dropping the peak width at half height and fixing requirements for FSC, FITC and PE, and extends the surface marker detection accuracy to CD16/CD56 and CD19.

It replaces YY/T 0588-2005, was issued on 5 December 2017 and has applied since 1 December 2018.

Found what you need?

Document preview — YY/T 0588-2017

National Standard of the People's Republic of China

Replacing
YY/T 0588-2005

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope5
  • 2 Normative References5
  • 3 Terms and Definitions6
  • 4 Technical Requirements7
  • 5 Test Methods9
  • 6 Labels, Marks and Instructions for Use12
  • 7 Package, Transportation and Storage12
  • Bibliography14

Foreword

This Standard was drafted as per the rules specified in GB/T 1.1-2009.

This Standard was finished through modifying the YY/T 0588-2005 Flow Cytometer; compared with YY/T 0588-2005, the major technical changes are as follows besides the editorial modifications.

--- Modify the normative reference documents;

--- Modify the definition of "polity";

--- Modify the name and requirement of "sensitivity of fluorescence" into "detection limit of fluorescence"; add the fluorescence detection limit requirement of other lasers against the corresponding channel fluorescence;

--- Modify the name of "forward scatter sensitivity" into "forward scatter detection limit";

--- Modify the requirement of "instrument resolution"; remove the peak width at half height requirement; specify the requirements of FSC, FITC, PE; other fluorescein meet the requirements of the manufacturer (see 4.5 instrument resolution requirements);

--- Modify "surface marker detection accuracy"; add the requirements of CD16/CD56, and CD19;

--- Modify "surface marker detection repeatability"; perform subsection requirements against the variation coefficient of CD3, CD4, CD8 positive percentage results; add the requirements for measuring CD16/CD56 and CD19 (see 4.9 surface marker detection repeatability requirements);

--- Modify the requirements of "carry-over", which shall be no greater than 0.5%;

--- Delete the "instrument function";

--- Add security requirement contents of GB 4793.9, YY 0648 (see 4.14 security);

--- Add electromagnetic compatibility requirement contents of GB/T 18268.1, GB/T 18268.26 (see 4.15 electromagnetic compatibility);

--- Modify the test method of "detection limit of fluorescence" (see 5.2 detection limit of fluorescence);

--- Modify the test method of "fluorescence linearity" (see 5.3 fluorescence linearity);

1 Scope

This Standard specifies the terms and definitions, product classification, technical requirements, test methods, mark, label and use instructions, package, transportation and storage of flow cytometer (FCM).

This Standard is applicable to the clinical use of flow cytometer for quantitative analysis and sorting of single-cell or other non-biological particle membrane surface and the internal biochemical and biophysical properties (only limited to the flow cytometer with sorting function).

2 Normative References

The following documents are essential to the application of this document. For the dated documents, only the versions with the dates indicated are applicable to this document; for the undated documents, only the latest version (including all the amendments) are applicable to this document.

GB/T 191 Packaging - Pictorial Marking for Handling of Goods

GB/T 4793.1 Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use - Part 1. General Requirements

GB/T 4793.9 Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use - Part 9. Particular Requirements for Automatic and Semi-Automatic Laboratory Equipment for Analysis and Other Purposes

GB/T 14710 Environmental Requirement and Test Methods for Medical Electrical Equipment

GB/T 18268.1 Electrical Equipment for Measurement, Control and Laboratory Use - EMC Requirements - Part 1. General Requirements

GB/T 18268.26 Electrical Equipment for Measurement, Control and Laboratory Use -EMC Requirements - Part 26. Particular Requirements - In Vitro Diagnostic (IVD) Medical Equipment

GB/T 29791.3 In Vitro Diagnostic Medical Devices - Information Supplied by the Manufacturer (Labelling) - Part 3. In Vitro Diagnostic Instruments for Professional

3.8 Carry-over

The carrying amount of the analyte transferred by the instrument from one test sample to the next, thereby erroneously cause an increase of the analyte concentration of the section test sample.

3.9 Standard particle

Microspheres of uniform size and/r labeled with consistent strength, constant fluorescein for calibration of flow cytometer.

4 Technical Requirements

4.1 Normal working conditions

The normal working conditions of the flow cytometer shall meet the following requirements.

a) Ambient temperature. as specified in the flow cytometer manual;

b) Relative humidity. as specified in the flow cytometer manual;

c) Power supply voltage. AC 220V±22V; 50Hz±1Hz;

d) Atmospheric pressure. as specified in the flow cytometer manual;

e) Protect from direct expose to the sunlight and avoid heat sources.

4.2 Sensitivity of fluorescence

The sensitivity of fluorescence of flow cytometer shall meet the following requirements.

a) The sensitivity of fluorescence of flow cytometer against the fluorescein isothiocyanate (FITC) shall be no more than 200 MESF;

b) The sensitivity of fluorescence of flow cytometer against phycoerythrin (PE) shall be no more than 100 MESF;

c) The sensitivity of fluorescence of flow cytometer against the channel fluorescein (each laser shall choose at least one fluorescein) of other lasers (such as red laser, ultraviolet laser, green laser) shall meet the claimed requirements of the manufacturer.

4.3 Fluorescence linearity

The correlation coefficient (r) of fluorescence intensity linearity of the flow cytometer less than 10000 pieces of standard particles; perform the histogram analysis on the test results; obtain the average fluorescence intensity of each peak; according to the equivalent amount of MESF of each peak provided by the standard particle instructions, and average fluorescence intensity of each peak obtained by the analysis, take the common logarithm (Lg value) to perform the linear regression; the antilog of MESF corresponding to the average fluorescence intensity value at the blank particle position is the sensitivity of fluorescence.

5.3 Fluorescence linearity

The standard particles are thoroughly mixed, and tested on the machine; collect no less than 10000 pieces of standard particles; the flow cytometer performs the histogram analysis on the test results; obtain the average fluorescence intensity of each peak; according to the equivalent amount of MESF of each peak provided by the standard particle instructions, and average fluorescence intensity of each peak obtained by the analysis, calculate the relevant coefficient (r) by the linear regression between MESF amount (y) and average fluorescence intensity (x).

5.4 FSC sensitivity

The standard particles are thoroughly mixed and tested on the machine; check the peak signal and standard particle diameter of peak signal displayed on the histogram.

5.5 Instrument resolution

The standard particles are thoroughly mixed and tested on the machine; calculate the CV value of full peak width of the standard particles in each fluorescent channel; the results shall meet the requirements of 4.5.

5.6 Resolution of FSC and SSC

5.6.1 Add 5µL of sodium citrate anticoagulated whole blood to the test tube containing 1mL of sheath liquid; mix evenly; test on the machine; check whether the FSC and SSC spot maps can separate the platelets from the red blood cells.

5.6.2 Take 100µL of EDTA anticoagulated whole blood; after dissolving the red blood cells, check on the machine; check whether the FSC and SSC spot maps can separate the three groups (lymphocytes, monocytes, granulocytes) of white blood cells.

5.7 Polity analysis linearity

Use the fluorescently stained standard cell line or standard cell nucleus to test the average fluorescence intensity of G0/G1 phase and G2/M phase on the machine; then calculate the ratio of the two average fluorescence intensities.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 19 pages — is available in the English PDF.

Referenced standards

Editions of YY/T 0588

EditionTitleRevisionStatus
YY/T 0588-2017Flow cytometercurrent editionCurrent
YY/T 0588-2005Flow cytometerprevious editionSuperseded

This page sells the current edition, YY/T 0588-2017. Earlier editions are listed for reference only.

How to Download YY/T 0588-2017

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Fast delivery

    After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
19 pages
Fast delivery
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY/T 0588-2017

$125.00