YY/T 0456.2-2014Reagents for hematology analyzer - Part 2: Hemolysin (English PDF)
血液分析仪用试剂 第2部分:溶血剂
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for YY/T 0456.2-2014 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Recommended
Issue date
June 17, 2014
Implementation date
July 1, 2015
Scope
YY/T 0456.2-2014 is the English-translated version of 血液分析仪用试剂 第2部分:溶血剂.
YY/T 0456.2-2014 is the Chinese pharmaceutical industry standard for the hemolysin used with a haematology analyzer, the reagent that lyses the red cells so that the white cells and the haemoglobin can be measured.
It is the second part of the YY/T 0456 series and specifies the requirements, the test methods, the marking, the labelling, the instructions for use and the rules for packaging, transport and storage, covering the appearance, the lysing performance, the accuracy and repeatability of the results obtained with it and the stability. Issued on 17 June 2014, it has applied since 1 July 2015.
Found what you need?
Document preview — YY/T 0456.2-2014
National Standard of the People's Republic of China
- ICS
- 11.100
- Classification
- C 44
- Replacing
- YY/T 0456.2-2003
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword
- 1 Scope
- 2 Normative references
- 3 Requirements
Foreword
YY/T 0456 "hematology analyzer reagent" is divided into five parts.
--- Part 1. cleaning solution;
--- Part 2. hemolytic agent;
--- Part 3. dilution;
--- Part 4. nucleated red blood cell detection reagents;
--- Part 5. reticulocyte detection reagent. This section YY/T Section 20456 of. This section drafted in accordance with GB/T 1.1-2009 given rules.
This section instead of YY/T 0456.2-2003 "hematology analyzer reagent applications," compared with the YY/T 0456.2-2003, the main technical Surgery changes are as follows.
--- Standard name "Blood Cell Analyzer agent" to "blood analyzer reagent";
--- Range "This section applies to the principle of electrical impedance of blood cell analyzer hemolytic agent" to "This section applies to blood points Analyzer hemolytic agent ";
--- Remove the "naming and classification";
--- Appearance requirements changed to "should be a clear liquid, without precipitation, particles or floc";
--- Removal requests "external signs", "Manual" contents;
--- "Net weight" to "loading capacity", its content was changed to "shall not be less than the nominal value";
--- Delete "absorbance values" content;
--- Blank WBC count requirements "<=0.3 x 10^9/L" changed to "<=0.5 x 10^9/L";
--- PH requirements "pH hemolytic agent should not exceed the product of the nominal value ± 0.20" changed to "pH should be in the range of the nominal value ± 0.50 Inside";
--- Accuracy requirements combined original reagents and reagent alternative content, content revised to "white blood cell count. relative deviation ± 7.5% Within the scope "," hemoglobin. the relative deviation within ± 3.5% range ";
--- WBC histogram modification of WBC histogram or scatter plot, scatter plot content added on request;
--- Batch difference between the requirements of "the difference between the pH of the batch should be consistent. deltapH<=0.20" was changed to "the difference between the pH of the batch should be consistent. deltapH<=1.0".
1 Scope
YY/T 0456 provisions of this part of the blood analyzer with a hemolytic agent requirements, test methods, marking, labeling and instructions for use, package Loading, transport and storage and so on. This section applies to a blood analyzer hemolytic agent.
2 Normative references
The following documents for the application of this document is essential. For dated references, only the dated version suitable for use herein Member. For undated references, the latest edition (including any amendments) applies to this document.
GB/T 191 Packaging - Pictorial signs
3 Requirements
3.1 Appearance
It should be a clear liquid, without precipitation, particles or floc.
3.2 PACKING
It should not be less than the nominal value.
3.3 pH
The pH should be within ± 0.50 in nominal value range.
3.4 Absorption peak wavelength
After hemolysis hemolytic agent, a characteristic absorption peak wavelength lambdamax should be within ± 10nm nominal value range.
3.5 Blank Count
Blank count shall conform to the requirements of Table 1.
3.6 Accuracy
Deviation of the test results shall comply with the requirements of Table 2.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
Referenced standards
Normative references
Similar standards
Editions of YY/T 0456.2
| Edition | Title | Revision | Status |
|---|---|---|---|
| YY/T 0456.2-2014 | Reagents for hematology analyzer - Part 2: Hemolysin | current edition | Current |
| YY/T 0456.2-2003 | Reagents for hematology analyzer - Part 2: Hemolysin | previous edition | Superseded |
This page sells the current edition, YY/T 0456.2-2014. Earlier editions are listed for reference only.
How to Download YY/T 0456.2-2014
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 191-2000 — Packaging-Pictorial marking for handling of goods
GB/T 191-2008 — Packaging - Pictorial marking for handling of goods
GB/T 191-2025 — Graphical symbols marking for handling and storage of packages
Secure payment via Stripe
Payments accepted
YY/T 0456.2-2014
$95.00