YY 9706.250-2021Medical electrical equipment - Part 2-50: Particular requirements for the basic safety and essential performance of infant phototherapy equipment (English PDF)

医用电气设备 第2-50部分:婴儿光治疗设备的基本安全和基本性能专用要求

Open the YY 9706.250-2021 preview as PDF

Preview — first pages of YY 9706.250-2021 (full document: 38 pages)

This is a limited preview

Buy now to download the full PDF (38 pages)

Issued by

NMPA

Level / Type

Industry · Mandatory

Issue date

March 9, 2021

Implementation date

May 1, 2023

Scope

YY 9706.250-2021 is the English-translated version of 医用电气设备 第2-50部分:婴儿光治疗设备的基本安全和基本性能专用要求.

YY 9706.250 is the mandatory Chinese standard for infant phototherapy equipment, the blue light units used to treat neonatal jaundice. It sets the particular requirements for basic safety and essential performance on top of the general medical electrical safety standard, covering the irradiance delivered at the infant's skin and its uniformity across the treated area, the spectral band that actually breaks down bilirubin, the thermal effect on a newborn with limited temperature regulation, and the eye protection. Under-dosing prolongs jaundice towards exchange transfusion and kernicterus; over-heating harms a baby who cannot move away. Phototherapy units run in every Chinese maternity unit. Being a YY standard without the /T suffix, compliance is compulsory. It took effect in May 2023.

Document preview — YY 9706.250-2021

National Standard of the People's Republic of China

Classification
C40

Issued by: National Medical Products Administration

Contents

  • ForewordIII
  • IntroductionV
  • 201 1 Scope, purpose and related standards1
  • 201 2 Normative references2
  • 201 3 Terms and Definitions2
  • 201 4 General requirements4
  • 201 5 General requirements for ME equipment testing4
  • 201 6 Classification of ME equipment and ME systems5
  • 201 7 ME equipment identification, marking and documentation5
  • 201 8 Protection of ME equipment against electric shock hazard7
  • 201 9 Protection of ME equipment and ME system against mechanical hazards7
  • 201 10 Protection against unwanted or excessive radiation hazards (sources)8
  • 201 11 Protection against over-temperature and other hazards9
  • 201 12 Accuracy of controllers and instruments and protection against dangerous outputs9
  • 201 13 Dangerous situations and fault states11
  • 201 14 Programmable Medical Electrical System (PEMS)11
  • 201 15 Structure of ME equipment11
  • 201 16 ME System12
  • 201 17 Electromagnetic compatibility of ME equipment and ME system12
  • 202 Electromagnetic compatibility 12 Appendix 13 Appendix AA (informative appendix) Specific guidelines and explanations of principles14

Foreword

This document was issued on 9 March 2021 by the National Medical Products Administration and takes effect on 1 May 2023.

It is a YY standard without the /T suffix: compliance is mandatory in China.

It is classified under Chinese classification C40.

All technical content of this part is mandatory. The "Medical Electrical Equipment" series of standards is divided into two parts.

---Part 1: General and parallel requirements;

---Part 2: Special requirements. This part is part 2-50. This section was drafted in accordance with the rules given in GB/T 1.1-2009. This part replaces YY 0669-2008 "Medical Electrical Equipment Part 2: Special Requirements for the Safety of Infant Phototherapy Equipment", and Compared with YY 0669-2008, the main technical changes except for editorial changes are as follows.

--- Added the term and definition "baby" (see.201.3.202);

--- Deleted the term and definition of irradiance parameter" (see 2.12.101 of YY 0669-2008);

--- Deleted the term and definition "uniformity of total bilirubin irradiance" (see 2.12.103 of YY 0669-2008);

---Modified the "measurement location" (see.201.5.4.102, 4.6.102 of YY 0669-2008);

---Modified the "spatial layout" (see.201.5.4.104, 4.6.104 of YY 0669-2008);

---Modified "Infant light therapy equipment placed under the patient" (see.201.6.3.101, 5.3.101 of YY 0669-2008);

--- Added "Safety Marks for Patient Eye Masks" (see.201.7.2.101);

--- Modified "Warning and Safety Instructions" (see 6.8.2 of.201.7.9.2.2, YY 0669-2008);

--- Added "Accessories, additional equipment, materials used" (see.201.7.9.2.14);

--- Added "Accessory bracket and mounting frame" (see.201.9.8.101);

--- Added "ME equipment power supply/power supply network interruption" (see.201.11.8);

--- Modified "Bilirubin total irradiance after pre-aging" (see.201.12.1.105, 50.104 of YY 0669-2008);

--- Added "weight scale" (see.201.12.1.107);

---Modified the "Immunity Test Level" (see 202.8.9, 36.202 of YY 0669-2008);

---Modified the "UV radiation exposure limit and spectral weighting function" (see Table AA.1, YY 0669-2008 Table AA.1). This part uses the redrafting law to amend and adopt IEC 60601-2-50.2009 A1.2016 "Medical Electrical Equipment Part 2-50. Special requirements for basic safety and basic performance of infant phototherapy equipment. Compared with IEC 60601-2-50.2009 A1.2016, this part has technical differences. The vertical single line (|) of the outer margin position is marked. This part is technically poor with IEC 60601-2-50.2009 A1.2016 The differences and their reasons are as follows.

---Regarding normative reference documents, this section has made adjustments with technical differences to adapt to my country's technical conditions and adjustments. The situation is collectively reflected in the "201.2 Normative Reference Documents", and the specific adjustments are as follows. * Replace IEC 60601-1 with GB 9706.1 which is modified to adopt international standards; * Replace IEC 60601-1-2 with YY 9706.102 which is modified to adopt international standards;

---Modified the safety mark of the patient's eyewear (see.201.7.2.11) to make the requirements more clear;

--- Modified.201.11.1.2.2 (see Appendix AA), corresponding to.201.11.1.2.2. This section has made the following editorial changes.

---Deleted part of the notes in IEC 60601-2-50.2009 A1.2016. Please note that some of the contents of this document may involve patents. The issuing agency of this document is not responsible for identifying these patents. This part was proposed by the State Drug Administration. This part is organized by the National Optical and Photonics Standardization Technical Committee Medical Optics and Instrument Sub-Technical Committee (SAC/TC103/SC 1) Centralized. Drafting organizations of this section. Zhejiang Medical Device Inspection and Research Institute, Ningbo Dawei Medical Devices Co., Ltd. The main drafters of this section. Ye Yueshun, Du Kun, Guo Yongbing, Fang Chunzi, Lin Dingyu, Wu Yi. The previous releases of the standards replaced by this part are as follows.

---YY 0669-2008.

Introduction

This section deals with the safety of infant light therapy equipment. This part is a revision and supplement to GB 9706.1 (hereinafter referred to as the general standard) Charge. The requirements of this section take precedence over the requirements of general standards. The guidelines and basic principles required by this section are contained in Appendix AA. Understanding the reasons for the preparation of these requirements not only helps to apply this standard correctly, but also accelerates the changes due to clinical practice in a timely manner. Or the process of revising the standard as a result of technological development. Nevertheless, this appendix is not an integral part of the requirements of this section. Chapters and articles marked with an asterisk (*) provide explanatory notes in Appendix AA "Guidelines and Basic Principles for Specific Provisions" of this part. Medical Electrical Equipment Part 2-50. Basic safety and safety of infant light therapy equipment Special requirements for basic performance 201.1 Scope, purpose and related standards In addition to the following, Chapter 1 of the general standard applies. 201.1.1 *Scope replace. This section specifies the safety requirements for infant phototherapy equipment. However, if the manufacturer has indicated in its risk management Compared with the benefits of preparation treatment, the related hazards and risks can be maintained within an acceptable range, and the replacement that meets specific provisions and has the same safety Substitution method will not be regarded as a violation of relevant regulations. This section applies to the basic safety and basic performance of infant light therapy equipment, also known as ME equipment. If a clause or sub-clause is specifically applicable to ME equipment or ME system, the title and content of the clause or sub-clause will be clearly stated. Make it clear. If not clearly stated, this clause or sub-clause applies to the relevant ME equipment and ME system. Except for 7.2.13 and 8.4.1 in the general standards, the special requirements of this part do not include ME equipment or ME systems within the scope of this part The inherent hazards of expected physiological functions. Note. See 4.2 in the general standard. This section does not apply to.

--- Devices that provide heat through blankets, pads or mattresses in medical use, see YY 9706.235 for information;

---Infant incubator, see YY 0455 for information;

---Infant transfer incubator, see YY 9706.220 for information;

---Infant radiant warmer, see GB 11243 for information. 201.1.2 Purpose replace. The goal of this part is to formulate special requirements for the basic safety and basic performance of infant phototherapy equipment, so as to reduce such equipment as much as possible. Prepare the safety hazards to patients and operators, and specify tests to confirm compliance with these requirements. 201.1.3 Parallel standards supplement. These applicable parallel standards involved in this section have been listed in Chapter 2 of the General Standard and Chapter 2 of this Part. According to the modification described in 202, YY 9706.102 applies. GB 9706.103 and YY/T 9706.1101) are not applicable. According to the issued For information, all other published parallel standards in the GB 9706.1 series apply. 1) YY/T 9706.110 "Medical Electrical Equipment Part 1-10: General Requirements for Basic Safety and Basic Performance. Parallel Standard. Physiological Closed Loop Controller Development requirements 201.1.4 Specific standards replace.

1 Scope

YY 9706.250 is the mandatory Chinese standard for infant phototherapy equipment, the blue light units used to treat neonatal jaundice. It sets the particular requirements for basic safety and essential performance on top of the general medical electrical safety standard, covering the irradiance delivered at the infant's skin and its uniformity across the treated area, the spectral band that actually breaks down bilirubin, the thermal effect on a newborn with limited temperature regulation, and the eye protection. Under-dosing prolongs jaundice towards exchange transfusion and kernicterus; over-heating harms a baby who cannot move away. Phototherapy units run in every Chinese maternity unit. Being a YY standard without the /T suffix, compliance is compulsory. It took effect in May 2023.

Remaining clauses in the full document

  • 201 1 Scope, purpose and related standards
  • 201 2 Normative references
  • 201 3 Terms and Definitions
  • 201 4 General requirements
  • 201 5 General requirements for ME equipment testing
  • 201 6 Classification of ME equipment and ME systems
  • 201 7 ME equipment identification, marking and documentation
  • 201 8 Protection of ME equipment against electric shock hazard
  • 201 9 Protection of ME equipment and ME system against mechanical hazards
  • 201 10 Protection against unwanted or excessive radiation hazards (sources)
  • 201 11 Protection against over-temperature and other hazards
  • 201 12 Accuracy of controllers and instruments and protection against dangerous outputs
  • 201 13 Dangerous situations and fault states
  • 201 14 Programmable Medical Electrical System (PEMS)
  • 201 15 Structure of ME equipment
  • 201 16 ME System
  • 201 17 Electromagnetic compatibility of ME equipment and ME system
  • 202 Electromagnetic compatibility 12 Appendix 13 Appendix AA (informative appendix) Specific guidelines and explanations of principles

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 38 pages — is available in the English PDF.

How to Buy YY 9706.250-2021

  1. 1Add to cart. Click the "Buy YY 9706.250-2021" button on this page. You can add more standards before checkout.
  2. 2Checkout. Enter your email and billing details. Payment is processed securely by Stripe (cards, Apple Pay, Google Pay supported).
  3. 3Instant delivery (0–9 sec). Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
  4. 4Invoice included. A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
38 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY 9706.250-2021

$390.00

$330.00for partners