Valid

YY 1727-2020Human immunodeficiency virus antibody detection kit for oral mucosal transudate (colloidal gold immunochromatography) (English PDF)

口腔黏膜渗出液人类免疫缺陷病毒抗体检测试剂盒(胶体金免疫层析法)

Preview PDF

This is a limited preview

Buy now to download the full PDF (17 pages)

Need quotation for your management?

Get a formal quotation for YY 1727-2020 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Mandatory

Issue date

June 30, 2020

Implementation date

June 1, 2022

Scope

YY 1727-2020 is the English-translated version of 口腔黏膜渗出液人类免疫缺陷病毒抗体检测试剂盒(胶体金免疫层析法).

YY 1727-2020 is the Chinese standard "Human immunodeficiency virus antibody detection kit for oral mucosal transudate (colloidal gold immunochromatography)".

Its scope clause reads: This standard specifies the requirements for the human immunodeficiency virus antibody detection kit (colloidal gold immunochromatography) for oral mucosal exudate, Test methods, labels and instructions for use, packaging, transportation and storage.

This standard is applicable to the detection of human immunodeficiency virus type 1 and human immunodeficiency virus in human oral mucosal exudate based on the principle of colloidal gold immunochromatography.

A human immunodeficiency virus antibody detection kit for qualitative detection of type 2 (HIV-1/2) antibodies (hereinafter referred to as "kit"). Its clauses include requirements.

Found what you need?

Document preview — YY 1727-2020

National Standard of the People's Republic of China

ICS
11.040.30
Classification
C30

Issued by: National Medical Products Administration

Contents

  • Foreword
  • 1 Scope
  • 2 Normative references
  • 3 requirements
  • 3.1 Appearance
  • 3.2 Physical inspection
  • 3.3 Compliance rate of antibody negative reference products

1 Scope

This standard specifies the requirements for the human immunodeficiency virus antibody detection kit (colloidal gold immunochromatography) for oral mucosal exudate,

Test methods, labels and instructions for use, packaging, transportation and storage.

This standard is applicable to the detection of human immunodeficiency virus type 1 and human immunodeficiency virus in human oral mucosal exudate based on the principle of colloidal gold immunochromatography.

A human immunodeficiency virus antibody detection kit for qualitative detection of type 2 (HIV-1/2) antibodies (hereinafter referred to as "kit").

2 Normative references

The following documents are indispensable for the application of this document. For dated reference documents, only the dated version applies to this article

Pieces. For undated references, the latest version (including all amendments) applies to this document.

GB/T 191 Packaging, storage and transportation pictorial signs

GB/T 29791.2 Information provided by manufacturers of in vitro diagnostic medical devices (labeling) Part 2.Professional in vitro diagnostic reagents

3 requirements

3.1 Appearance

The appearance should meet the following requirements.

a) Each component of the kit should be complete and complete, with no liquid leakage;

b) Chinese packaging labels should be clear and free from wear.

3.2 Physical inspection

The physical inspection should meet the following requirements.

a) The width of the film strip should be >=2.5mm;

b) The liquid moving speed should be >=10mm/min.

3.3 Compliance rate of antibody negative reference products

Use the national negative reference product for testing, the positive reaction should not be more than 2 copies, the negative reference product compliance rate (-/-) should be >=18/20; or use

After the standardized negative reference product is tested, the result should meet the corresponding requirements.

3.4 Compliance rate of antibody positive reference products

Use national positive reference materials (see Appendix A) for testing, the HIV-1 positive reference product conformance rate (/) should be 18/18, HIV-

The conformity rate of type 2 positive reference product (/) should be 2/2; or use a standardized positive reference product for testing, and the result should meet the corresponding requirements.

3.5 Lowest detection limit

Use the national minimum detection limit reference product for detection, the positive reaction shall not be less than 3 (>=3/5) and the diluted matrix (S1) shall be a negative reaction;

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 17 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download YY 1727-2020

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included