YY 0950-2015Extracorporeal pressure wave therapy devices by compressed air (English PDF)
气压弹道式体外压力波治疗设备
This is a limited preview
Buy now to download the full PDF (21 pages)
Need quotation for your management?
Get a formal quotation for YY 0950-2015 in 30 minutes.
Issued by
NMPA
Level / Type
Industry · Mandatory
Issue date
March 2, 2015
Implementation date
January 1, 2017
Scope
YY 0950-2015 is the English-translated version of 气压弹道式体外压力波治疗设备.
YY 0950-2015 is the Chinese pharmaceutical industry standard for the extracorporeal pressure wave therapy device driven by compressed air, the machine that treats by sending a ballistic pressure wave into the tissue through a handpiece.
It specifies the terms and definitions, the composition, the requirements, the test methods, the inspection rules, the marks, the instructions for use and the rules for packaging, transport and storage, covering the pressure and the pulse rate delivered and their accuracy, the applicators, the air supply and the compressor, the noise, and the electrical safety of the equipment. Issued on 2 March 2015, it has applied since 1 January 2017.
Found what you need?
Document preview — YY 0950-2015
National Standard of the People's Republic of China
Issued by: National Medical Products Administration of the People's Republic of China
Contents
- Foreword3
- 1 Scope4
- 2 Normative references4
- 3 Terms and definitions5
- 4 Composition6
- 5 Requirements6
- 6 Test methods9
- 7 Inspection rules15
- 8 Marks and instructions for use16
- 9 Packaging, transportation and storage17
Foreword
This Standard was drafted in accordance with the provisions given in the GB/T 1.1-2009.
The safety requirements of this Standard have fully implemented the contents of GB 9706.1-2007 Medical electrical equipment - Part 1: General requirements for safety.
Please pay attention that some contents in this document might involve patents.
The issuing authority of this document does not undertake the responsibility for the identification of these patents.
1 Scope
This Standard specifies the terms and definitions, composition, requirements, test methods, inspection rules, marks, instructions for use, packaging, transportation and storage of the extracorporeal pressure wave therapy devices by compressed air (hereinafter referred to as equipment).
This Standard applies to the equipment specified in the Article 3.1.
This Standard does not apply to the rock crushing equipment by compressed air.
2 Normative references
The following documents are essential to the application of this document. For dated references, only the versions with the dates indicated are applicable to this document. For undated references, only the latest versions (including all the amendments) are applicable to this document.
GB/T 191-2008 Packaging - Pictorial marking for handling of goods
GB 9706.1-2007 Medical electrical equipment - Part 1: General requirements for safety
GB/T 14710-2009 Environmental requirement and test methods for medical electrical equipment
GB/T 16886.1 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
3 Terms and definitions
The terms and definitions defined in the GB 9706.1-2007 and the following ones apply to this document.
3.1 Extracorporeal pressure wave therapy devices by compressed air
The extracorporeal pressure wave therapy devices by compressed air refer to the therapy devices that use the projectiles in the energy-driven therapy handles generated by compressed air to conduct impulse-type impact to the therapy heads, and use the pressure waves generated by the elastic collision between projectiles and therapy heads to act on the locations of pain by means of percutaneous conduction.
3.2 Control handle
The operators hold the devices in hands to control the pressure waves generated by the collision between projectiles and therapy heads driven by compressed air.
4 Composition
The devices generally include host machines, control handles (containing projectiles), therapy heads, and air compressors (generally included in the host computers).
5 Requirements
5.1 Operating conditions
The operating conditions shall conform to the provisions formulated by manufacturers. If not specified, the operating conditions shall meet the requirements specified in the Chapter 10 of GB 9706.1-2007.
5.4 Energy density
The manufacturers shall declare the maximum energy density of each therapy head in the accompanying files. The error shall not be greater than ±20%.
5.5 Depth of penetration
The manufacturers shall declare the maximum depth of penetration of devices in the accompanying files. The error shall not be greater than ±20%.
5.6 Collision frequency
The collision frequency of projectiles and therapy heads shall be adjustable. The error of the collision frequency shall not exceed ±10%.
5.7 Timer or counter
The devices shall have at leave one of the following functions for single therapy counting.
5.8 Pulse width
The manufacturers shall declare the pulse width of output pressure waves in the accompanying files. The error shall not be greater than ±10%.
5.11 Fatigue performance of therapy heads
The manufacturers shall declare the life cycle of devices (calculated by number of times) in the accompanying files. Furthermore, the output energy density of devices throughout the entire cycle shall meet the requirements specified in the Article 5.4.
5.12 Corrosion resistance
The outer surfaces of therapy heads shall have good corrosion resistance. After being conducted with the corrosion resistance test, the therapy heads shall have no corrosion marks, or shall only have slight marks that can be removed after wiping and simple cleaning.
5.14 Functions
5.14.1 The devices shall have the function of displaying, setting or adjusting the operating pressure, collision frequency, time of therapy, or number of collisions.
5.14.2 The devices shall have the function of prompting the insufficient air pressure.
5.14.3 Besides the counting function required in the Article 5.7, each device shall also have a separate counter (or counting function) to record the total number of collisions of projectiles and therapy heads. Furthermore, the above-mentioned counters can only be reset by manufacturers.
5.15 Appearance
The appearance of devices shall be clean and without any mechanical damages, scratches or other defects. The marks shall be clear and distinct. The operating and regulating mechanisms shall be flexible and reliable. The fasteners shall be tight
5.16 Safety requirements
The safety requirements of devices shall meet the requirements specified in GB 9706.1-2007.
6 Test methods
6.6 to calculate the count value. The results shall meet the requirements specified in the Article 5.7.
6.1 Test devices
The following devices or the test devices whose accuracies are higher than those of the following devices are recommended to be used during the tests.
6.2 Operating pressure
CONNECT the pressure gauge to the input end of compressed air on the control handle in series.
6.3 Energy stability
MEASURE the output energy of devices by converting the energy of output pressure waves into gravitational potential energy. Closely CONNECT the therapy head with the device shown in Figure 1.The mass block shall be completely in contact with the probe.
6.8 Pulse width
The duration of pulse width refers to the corresponding time interval value at 50% of the pulse peak read from the pressure waves displaying on the oscilloscope during the test described in the Article 6.5. The results shall meet the requirements specified in the Article 5.8.
6.13 Biocompatibility
6.13.1 USE the therapy heads made of the proven materials conforming to the biocompatibility. The manufacturers shall declare the materials of therapy heads and the material standards used as basis, and provide relevant evidences as well. The material composition can be verified if necessary.
6.15 Appearance
The verification shall be performed by visual inspection and actual operation.
The results shall meet the requirements specified in the Article 5.15.
6.16 Safety requirements
The test shall be conducted according to the methods specified in the GB 9706.1-2007.
7 Inspection rules
7.1 Inspection category
The quality inspection of devices is divided into exit-factory inspection and type inspection.
7.2 Exit-factory inspection
The quality inspection department of the manufacturer shall inspect devices one by one before they leave the factory.
7.2.2 Determination of exit-factory inspection. The exit-factory inspection can only be determined as qualified after all the inspection items are qualified. Otherwise, the exit-factory inspection will be determined as unqualified.
7.3 Type inspection
7.3.1 The type inspection shall be conducted in the case of one of the following conditions.
7.3.2 The samples for type inspection shall be taken from the qualified exit-factory products. SEE Table 1 for the inspection items.
8 Marks and instructions for use
8.1 Marks
8.1.1 Nameplate
The nameplates shall be installed on the proper locations of devices. The nameplates shall include the following contents at least.
8.1.3 Qualification certificate of product inspection
The qualification certificate of product inspection shall include the following contents at least.
8.1.4 Labels, marks, and the symbols providing information
They shall meet the requirements specified in the YY/T 0466.1-2009.
8.2 Instructions for use
The instructions for use shall include the following contents at least.
9 Packaging, transportation and storage
9.1 Packaging
9.1.3 Solidly FIX the devices in the boxes to prevent the looseness and scratches during the transportation.
9.1.4 The nude packing of devices shall be performed according to the provisions specified in the purchase contact. It is necessary to take protective measures for nude packing during the transportation. The specific requirements shall conform to the provisions specified in the purchase contact.
9.1.5 The devices shall have the following accompanying files.
9.2 Transportation
The transportation shall conform to the provisions specified in the purchase contact.
9.3 Storage
The devices shall be stored in the well-ventilated rooms without any corrosive gases at -40°C~55°C. The relative humidity shall not be greater than 90%. […]
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 21 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 191-2008Graphical symbols marking for handling and storage of packages
- GB 9706.1-2007Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- GB/T 14710-2009The environmental requirement and test methods for medical electrical equipment
- GB/T 16886.1Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process
Similar standards
How to Download YY 0950-2015
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 14710-2009 — The environmental requirement and test methods for medical electrical equipment
GB/T 16886.1-2022 — Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process