YY 0876-2013Linear cutter stapler and cartridge (English PDF)

直线型切割吻合器及组件

Preview PDF

This is a limited preview

Buy now to download the full PDF (26 pages)

Need quotation for your management?

Get a formal quotation for YY 0876-2013 in 30 minutes.

Issued by

NMPA

Level / Type

Industry · Mandatory

Issue date

October 21, 2013

Implementation date

October 1, 2014

Scope

YY 0876-2013 is the English-translated version of 直线型切割吻合器及组件.

YY 0876-2013 is the Chinese pharmaceutical industry standard for the linear cutter stapler and its cartridge, the instrument that in a single action staples the tissue in rows and cuts between them.

It specifies the structural types and the materials, the requirements, the test methods, the marks, the instructions for use and the rules for packaging, transport and storage, covering the formed staple and its dimensions, the firing and cutting force, the completeness of the staple rows and of the cut, the fit between stapler and cartridge, the corrosion resistance and the biological evaluation, and sterility. Issued on 21 October 2013, it has applied since 1 October 2014.

Found what you need?

Document preview — YY 0876-2013

National Standard of the People's Republic of China

Issued by: National Medical Products Administration of the People's Republic of China

Contents

  • Foreword3
  • 1 Scope4
  • 2 Normative references4
  • 3 Structural types and materials5
  • 4 Requirements8
  • 5 Test methods10
  • 6 Marks, instructions for use12
  • 7 Packaging, transportation, storage13
  • Annex A (normative) Cutting edge sharpness test15
  • Annex B (informative) Test material for anastomosis performance18
  • Annex C (normative) Pressure resistance performance test19
  • Annex D (informative) Sealing performance test21
  • Annex E (normative) Peel strength test23
  • Annex F (informative) Type inspection25
  • Bibliography26

1 Scope

This document specifies structural types and materials, requirements, test methods, marks, instructions for use and packaging, transportation and storage for linear cutter stapler and cartridge.

This Standard applies to linear cutter stapler and cartridge (hereinafter referred to as the stapler). The stapler is suitable for gastrointestinal reconstruction, anastomosis, amputation and resection of tissues and organs in organ resection.

This Standard does not apply to the stapler for endoscopic use.

2 Normative references

The following referenced documents are indispensable for the application of this document. For dated references, only the edition cited applies. For undated references, the latest edition of the referenced document (including any amendments) applies.

GB/T 228, Metallic materials - Tensile testing - Part 1: Method of test at room temperature

GB/T 1220, Stainless steel rod

GB/T 3280, Cold rolled stainless steel plate, sheet and strip

GB/T 10610, Geometrical product specifications (GPS) - Surface texture: Profile method - Rules and procedures for the assessment of surface texture

GB/T 12672, Acrylonitrile-butadiene-styrene (ABS) resin

GB/T 13810, Wrought titanium and titanium alloy for surgical implants

GB/T 16886.1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process

GB/T 16886.5, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity

GB/T 16886.7, Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals

GB/T 16886.10, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization

GB/T 19633, Packaging for terminally sterilized medical devices

YY 0167, Non-absorbable surgical suture

YY/T 0171-2008, Surgical instruments - Packaging, Marking and Instructions

YY/T 0174, Surgical blades

YY/T 0294.1, Surgical instruments - Metallic materials - Part 1: Stainless steel

HG/T 2503, Specification for polycarbonate resins by melt method Pharmacopoeia of the People's Republic of China (2010 Edition) Part II

ISO 13782:1996, Implants for surgery - Metallic materials - Unalloyed tantalum for surgical implant applications

3 Structural types and materials

3.1 According to the assembly position of the cutting knife, the stapler is divided into two types: instrument with knife (type I) and cartridge with knife (type II).

3.2 The stapler consists of a body and cartridge:

a) Type I stapler body consists of anvil seat, locking rod, cutting assembly frame and cutting knife. The cartridge is composed of staple cartridge, push piece and staple and so on;

b) Type II stapler body consists of anvil seat, locking rod, cutting assembly frame.

The cartridge consists of a staple cartridge, a pusher, a cutting knife, and staples.

3.3 See Figure 1, Figure 2 and Table 1 for the structural types and basic dimensions of the stapler and cartridge.

4 Requirements

4.3 Assembly

4.3.1 The replacement of stapler components shall be convenient and reliable.

4.3.2 After the assembly is shaken, the staples shall not be exposed on the surface of the staple cartridge.

4.4 Sharpness

The cutting edge shall be sharp. The cutting force shall not be greater than 0.80N.

4.5 Anastomotic and cutting performance

4.5.1 The stapler shall have good anastomotic and cutting performance, replaceable components. Make no less than 5 cutting anastomosis. The staples after each anastomosis shall be shaped like a "B".

4.5.2 At the same time, the cutting edge after each anastomosis shall be neat. There shall be no burrs.

4.5.3 The length of each staple line is at least 1.5 times longer than the length of the cutting line.

4.6 Pressure resistance

The anastomotic stoma after anastomosis shall be able to withstand a pressure of not less than 3.6 x 10^3Pa. Leak no more than 10 drops within 15s.

4.7 Safety device

The stapler shall have an empty cartridge safety device. Maintain its reliability.

NOTE: Empty cartridges refer to fired components.

4.8 Surface roughness

The roughness Ra of the metal outer surface of the stapler shall not be greater than 0.8µm.

4.9 Package seal

4.9.1 The stapler and cartridge shall be packaged in good condition. The adhesive seal area shall be free of channels or perforations.

4.9.2 The packaging and sealing peel strength of the stapler and cartridge shall not be less than 0.10N/mm. After peeling off, the two contact surfaces shall be smooth and continuous. There is no delamination or tearing.

4.10 Appearance

4.10.1 The stapler has a smooth shape, clear outline, no burrs, scratches and other defects.

4.10.2 The handwriting and signs on the outer surface of the stapler shall be clear. There shall be no defects such as dislocation and skew.

4.10.3 There shall be no burrs, dents and other defects on the surface of the staples.

4.11 Size

The size of the stapler and cartridge shall meet the requirements of Table 1.

4.12 Sterilization

4.12.1 The stapler and cartridge shall be sterilized by a validated sterilization process. The product shall be sterile.

4.12.2 If ethylene oxide is used for sterilization, the residual amount of ethylene oxide in the stapler and cartridge shall not exceed 10µg/g.

4.13 Biological evaluation

4.13.1 Staples, when they use pure titanium, titanium alloy materials and pure tantalum materials whose surface is modified, or other materials not specified in Table 2, shall be biologically evaluated according to the provisions of GB/T 16886.1. The evaluation results shall be free of biocompatibility hazards.

4.13.2 The cytotoxicity score of the anti-staple seat and the staple cartridge shall not be greater than 1.

4.13.3 The level of delayed type hypersensitivity reaction of the anti-staple seat and the staple cartridge shall not be greater than 1.

4.13.4 The score of the intradermal reaction of the anti-staple seat and the staple cartridge shall not be greater than 1.

5 Test methods

5.1 Material inspection

The chemical composition inspection of pure titanium and titanium alloy materials made of staples is carried out according to the method specified in GB/T 13810.

The chemical composition test of the tantalum wire material made of staples is carried out according to the method specified in ISO 13782.

The tensile strength of staple material is carried out according to the method specified Carry out the test according to the method specified in Annex D, which shall meet the requirements of 4.9.1.

5.9.2 Peel strength test

Carry out the test according to the method of Annex E, which shall meet the requirements of 4.9.2.

5.10 Appearance inspection

Visually inspect and touch with hands, which shall meet the requirements of 4.10.

5.11 Size inspection

Measure with a general-purpose measuring tool or a special measuring tool, which shall comply with the provisions of 4.11.

5.12 Sterility and ethylene oxide residue inspection

5.12.1 Carry out according to the method in Pharmacopoeia of the People's Republic of China (2010 Edition) Part II, which shall comply with the provisions of 4.12.1.

5.12.2 The determination of the residual amount of ethylene oxide shall be carried out according to the method specified in GB/T 16886.7, which shall meet the requirements of 4.12.2.

5.13 Biological evaluation

5.13.1 Staples made of pure titanium, titanium alloy materials and pure tantalum materials with modified surfaces or other materials not specified in Table 2 shall be biologically evaluated according to the items specified in GB/T 16886.1. It shall comply with the provisions of 4.13.1.

5.13.2 Carry out according to the method specified in 8.2 (extract test) of GB/T 16886.5-2003, which shall meet the requirements of 4.13.2.

5.13.3 Carry out according to the method specified in 7.4 (maximum dose test) in GB/T 16886.10-2005, which shall meet the requirements of 4.13.3.

5.13.4 Intradermal reaction test

Carry out according to the method specified in GB/T 16886.10-2005. (intradermal reaction test in Annex B.2), which shall meet the requirements of 4.13.4.

6 Marks, instructions for use

6.1 Marks, instructions for use

6.1.1 The body of each stapler shall have the manufacturer's code or trademark.

6.1.2 Packaging marks for the stapler and cartridge:

a) The single package mark shall comply with the provisions of 4.1.3 in YY/T 0171-2008;

b) The outer packaging mark shall comply with the provisions of 4.2.3 in YY/T 0171-2008.

6.1.3 The mark on the certificate shall comply with the provisions of 4.3 in YY/T 0171-2008.

6.2 Instructions for use

The preparation of the instructions for use shall comply with the provisions of Chapter 5 in YY/T 0171-2008.

7 Packaging, transportation, storage

7.1 Packaging

7.1.1 Stapler packaging

7.1.1.1 The stapler shall be packaged in a single unit. Sterility shall be maintained during the storage period.

7.1.1.2 The single package shall leave traces of opening after opening.

7.1.1.3 Put the single package into the packaging box. The instructions for use and certificate of conformity shall be attached.

7.1.2 Cartridge packaging

7.1.2.1 The cartridge shall be packaged in a single unit. Sterility shall be maintained during the storage period.

7.1.2.2 The single package shall leave traces of opening after opening.

7.1.2.3 Put several single packages into the packaging box. The instructions for use and certificate of conformity shall be attached.

7.2 Transportation and storage

7.2.1 The outer packaging of the stapler and cartridge shall ensure that they will not be damaged under normal transportation and storage conditions.

7.2.2 The words or signs on the outer packaging shall be guaranteed not to be blurred for a long time.

Annex C (normative) Pressure resistance performance test

C.1 Definition

C.1.1 Anastomosis (suture) pressure resistance: It refers to the pressure that the anastomosis (suture) can withstand after the tissue or organ is anastomosed (sutured) with a stapler.

C.1.2 Anastomosis: The incision of the tissue or organ is butted and completely anastomosed and unobstructed. C.1.3 Suture: Closure of the stump incision.

C.2 Instrument requirements

C.2.1 The instrument shall be manufactured according to the drawings and documents approved by the prescribed procedures.

C.2.2 The power supply voltage required by the instrument: 220V+22V. The power supply frequency: 50Hz±0.5Hz. C.2.3 Measurement unit of the instrument: kPa (kilopa). C.2.4 The instrument's indication range: 0~6kPa. Indication error: ±0.12kPa. C.2.5 The instrument has water inlet, drain and stop actions. C.2.6 The instrument's pressure gauge is checked every other year.

C.2.7 The water inlet and drainage pipes of the instrument shall be unobstructed. The control device shall be reliable.

C.3 Test material

C.3.1 The test material is fresh pig intestines. In vitro time is less than 24h. Store in the range of 5°C~15°C. No ice crystals are formed.

C.3.2 The intestinal cavity shall be cleaned before the test. Remove the intestinal fat and mesentery at the commissure to be anastomosed (sutured).

C.4 Test methods

C.4.1 Choose two bowels for side-to-side anastomosis.

Annex D (informative) Sealing performance test

D.1 Overview

This appendix uses the dye penetration test method to test the sealing performance of packaging seals.

D.2 Instrument and reagent preparation

D.2.1 1L beaker. D.2.2 25mL graduated cylinder. D.2.3 0.0005g balance.

D.3 Permeate formula

D.3.1 Add 5g of surfactant TritonX-100 to 100mL of grade three test water in compliance with GB/T 6682. Stir well.

D.3.2 Add grade three test water to 1000mL in C.3.1. Stir well again.

D.3.3 Take 0.5g of toluidine blue and add it to C.3.2. Stir well.

D.4 Test steps

D.4.1 During the test, the test personnel shall check that the test environment temperature is within the range of 23°C ± 2°C, and the relative humidity is within the range of 50% ± 2%.

D.4.2 Aspirate enough permeate with a syringe. Inject it into the sterilized primary package at the center of the package. Make the liquid cover the height of the sealing edge to 5mm.

Attention: Avoid splashing liquids on the skin. If necessary, wash hands with soap immediately to avoid damage to the skin.

D.4.3 The time for the permeation to cover the sealing edge shall be controlled within the range of 5s~20s. […]

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 26 pages — is available in the English PDF.

Referenced standards

Similar standards

How to Download YY 0876-2013

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
26 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

YY 0876-2013

$285.00