JJF 1946-2021Calibration Specification for Wave Measuring Devices (Acoustic method) (English PDF)
波浪测量仪(声学法)校准规范
This is a limited preview
Buy now to download the full PDF (12 pages)
Need quotation for your management?
Get a formal quotation for JJF 1946-2021 in 30 minutes.
Issued by
State Administration for Market Regulation
Level / Type
National · Mandatory
Issue date
December 28, 2021
Implementation date
June 28, 2022
Scope
JJF 1946-2021 is the English-translated version of 波浪测量仪(声学法)校准规范.
JJF 1946-2021 is the Chinese standard "Calibration Specification for Wave Measuring Devices (Acoustic method)". Issued on 2021-12-28, in force from 2022-06-28.
Found what you need?
Document preview — JJF 1946-2021
Contents
- Introduction (II)
- 1 Scope(1)
- 2 References (1)
- 3 Terms (1)
- 4 Overview(1)
- 5 Metrology Characteristics(1)
- 5.1 Wave height(1)
- 5.2 Cycle(2)
- 5.3 Center frequency of transmitted signal(2)
- 5.4 Transmit signal pulse width(2)
- 5.5 Sound transducer beam angle(2)
- 6 Calibration Conditions(2)
- 6.1 Environmental conditions (2)
- 6.2 Main standards and other equipment(2)
- 7 Calibration items and calibration methods (3)
- 7.1 Calibration items(3)
- 7.2 Calibration method(3)
- 8 Expression of calibration results (5)
- 8.1 Calibration records (5)
- 8.2 Calibration data processing(5)
- 8.3 Calibration certificate(5)
- 8.4 Measurement uncertainty of calibration results(5)
- 9 Recalibration interval(6)
- Appendix A Contents of Calibration Records(7)
- Appendix B Contents of Calibration Certificate(9)
- Appendix C Measurement Uncertainty Evaluation Example(11)
1 Scope(1)
2 References (1)
3 Terms (1)
4 Overview(1)
5 Metrology Characteristics(1)
5.1 Wave height(1)
5.2 Cycle(2)
5.3 Center frequency of transmitted signal(2)
5.4 Transmit signal pulse width(2)
5.5 Sound transducer beam angle(2)
6 Calibration Conditions(2)
6.1 Environmental conditions (2)
6.2 Main standards and other equipment(2)
7 Calibration items and calibration methods (3)
7.1 Calibration items(3)
7.2 Calibration method(3)
8 Expression of calibration results (5)
8.1 Calibration records (5)
8.2 Calibration data processing(5)
8.3 Calibration certificate(5)
8.4 Measurement uncertainty of calibration results(5)
9 Recalibration interval(6)
Appendix A Contents of Calibration Records(7)
Appendix B Contents of Calibration Certificate(9)
Appendix C Measurement Uncertainty Evaluation Example(11)
Introduction
JJF 1071-2010 "Rules for Compilation of National Metrology Calibration Specifications", JJF 1001 "General Metrology Terms and Definitions", JJF 1059.1-2012 "Assessment and Representation of Measurement Uncertainty" together constitute the basis for the preparation of this specification Series Specifications. This specification mainly refers to GB/T 14914.2-2019 "Ocean Observation Specification No. 2" in terms of technology.
Part. Seashore Observation Technical Requirements for Wave Observation.
This specification is published for the first time.
Specification for calibration of wave measuring instruments (acoustic method)
1 Scope
This specification applies to the ultrasonic ranging principle of wave measuring instrument (hereinafter referred to as "acoustic wave measuring instrument") calibration.
2 Reference documents this specification references the following documents.
JJF 1001-2011 General Metrology Terms and Definitions JJF 1034 Acoustic Metrology Terms and Definitions JJF 1059.1-2012 Evaluation and representation of measurement uncertainty GB 3102.5-1993 Electrical and Magnetic Quantities and Units GB 3102.7-1993 Quantities and Units of Acoustics GB/T 14914.2-2019 Ocean Observation Specification Part 2: Coastal Observation for dated references, only the dated version applies to this specification; for undated references the latest version (including all amendments) applies to this specification.
3 Terminology
Terms defined in JJF 1001-2011, JJF 1034 and GB/T 14914.2-2019 apply to this specification.
4 Overview
The acoustic wave measuring instrument usually consists of an acoustic transducer, a transmitter, a receiver, a preamplifier module, a filter module, control system, etc. The acoustic wave measuring instrument is based on the principle of sound wave ranging, and the transmitter transmits a wave to the water surface.
A beam of narrow pulsed sound waves, reflected by the water surface, returned to the receiving transducer at the same position as the transmitting transducer to receive, through Accurate wave height can be obtained from wave speed and time, and parameters such as wave period can be obtained from the sequence value of wave height changing with time. acoustic method the wave measuring instrument is mainly used for wave measurement of ocean stations, ports, docks, offshore platforms, rivers, lakes, reservoirs, etc.
High, period and waveform measurements.
5 Metrology characteristics
5.1 Wave height
a) The upper limit of the wave height measurement range generally does not exceed 20m;
b) The maximum allowable error of measurement. when the wave height is not greater than 1m, generally not more than ±0.10m; when the wave height is greater than 1m, generally not more than ± (0.1 × H 0.1m), where H is the measured value of wave height, the unit is meters.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 12 pages — is available in the English PDF.
How to Download JJF 1946-2021
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
JJF 2425-2026 — Calibration specification for carbon dioxide incubators
JJF 1183-2025 — Calibration Specification for Temperature Transmitters
JJF 1282-2025 — Calibration Specification for Time Delay Relays
JJF 1946-2021