JJF 1925-2021Calibration Specification for Low-frequency Volt-meters (English PDF)
低频电压表校准规范
This is a limited preview
Buy now to download the full PDF (17 pages)
Need quotation for your management?
Get a formal quotation for JJF 1925-2021 in 30 minutes.
Issued by
State Administration for Market Regulation
Level / Type
National · Mandatory
Issue date
October 18, 2021
Implementation date
April 18, 2022
Scope
JJF 1925-2021 is the English-translated version of 低频电压表校准规范.
JJF 1925-2021 is the Chinese standard "Calibration Specification for Low-frequency Volt-meters". Issued on 2021-10-18, in force from 2022-04-18.
Found what you need?
Document preview — JJF 1925-2021
Contents
- 1 Scope(1)
- 2 Overview(1)
- 3 Metrology Characteristics(1)
- 3.1 Voltage(1)
- 3.2 Isolation between two channels(1)
- 4 Calibration Conditions(1)
- 4.1 Environmental conditions (1)
- 4.2 Equipment for calibration(1)
- 5 Calibration items and calibration methods (2)
- 5.1 Calibration items(2)
- 5.2 Calibration method(2)
- 6 Expression of calibration results (7)
- 7 Recalibration interval(8)
- Introduction
1 Scope(1)
2 Overview(1)
3 Metrology Characteristics(1)
3.1 Voltage(1)
3.2 Isolation between two channels(1)
4 Calibration Conditions(1)
4.1 Environmental conditions (1)
4.2 Equipment for calibration(1)
5 Calibration items and calibration methods (2)
5.1 Calibration items(2)
5.2 Calibration method(2)
6 Expression of calibration results (7)
7 Recalibration interval(8)
Appendix A Raw Record Format(9)
Appendix B Calibration Certificate Inner Page Format(11)
Appendix C Example of Calibration Uncertainty Evaluation of Main Items (13)
Introduction
This specification is based on JJF 1071-2010 "Rules for Compilation of National Metrology Calibration Specifications" and JJF 1059.1-2012 "Measurement Uncertainty Evaluation and Representation" written. Take JJG782-1992 "Low-frequency Electronic Voltmeter Verification Regulations" as the basis is revised.
Compared with JJG782-1992 "Low-frequency Electronic Voltmeter Verification Regulations", the main technical changes except editorial revisions.
is as follows.
---The name of the procedure is changed from "Verification Regulations for Low-Frequency Electronic Voltmeters" to the normative name "Regulations for Calibration of Low-Frequency Voltmeters";
--- The lower limit of the frequency range is extended from 2Hz to 1Hz, and the upper limit of the frequency range is extended from 1MHz to 30MHz (see Section Chapter 1);
--- Added "Overview" section (see Chapter 2);
---The voltage measurement range is expanded from 100µV~300V to. 100µV~1000V; (see 3.1);
--- Improve the measurement performance of the basic error and frequency additional error (see 3.1.2 and 3.1.3);
--- Added the technical requirements for isolation between the two channels of a dual-channel low-frequency voltmeter (see 3.2);
--- Increased the measurement content of the digital low-frequency voltmeter (see 5.2.2);
--- Increased the measurement method of the isolation between the two channels of the dual-channel low-frequency voltmeter (see 5.2.4).
The previous versions of this specification have been released as follows.
---JJG782-1992 "Low Frequency Electronic Voltmeter Verification Regulations".
Low Frequency Voltmeter Calibration Specification
1 Scope
This specification applies to the calibration of low-frequency voltmeters with a frequency range of 1Hz to 30MHz.
2 Overview
The low-frequency voltmeter is composed of a voltage divider, an amplifier, a detector and an indicating circuit. The composition principle is shown in Figure 1.low frequency electricity Pressure gauges are mainly used for calibration, testing and maintenance of electronic circuits and electronic equipment.
Figure 1 Schematic diagram of low frequency voltmeter
3 Metrology characteristics
3.1 Voltage
3.1.1 Voltage measurement range. 10mV~10V (1Hz~10Hz), 100µV~200V (10Hz~ 1MHz),.200V~1000V (15Hz~30kHz), 1mV~3.5V (1MHz~30MHz).
3.1.2 Basic error. ± (0.05~3)%.
3.1.3 Frequency additional error. ± (0.05~10)%.
3.2 Isolation between two channels. >=80dB.
Note. The above technical indicators are not used for qualification judgment, but only for reference.
4 Calibration Conditions
4.1 Environmental conditions
4.1.1 Ambient temperature. (23±5)°C.
4.1.2 Relative humidity. <=80%.
4.1.3 Power requirements. power supply voltage (220±11)V, power supply frequency (50±1)Hz.
4.1.4 There is no electromagnetic interference and mechanical vibration that affect the normal operation of the instrument.
4.2 Equipment for calibration
4.2.1 Standard source method Low frequency voltage standard source.
1) Frequency range. 1Hz~30MHz;
2) Voltage range. 10mV~10V (1Hz~10Hz), 100µV~200V (10Hz~ 1MHz),.200V~1000V(15Hz~30kHz), 1mV~3.5V (1MHz~30MHz);
3) Maximum allowable error of output voltage. ± (0.015~1)%;
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 17 pages — is available in the English PDF.
How to Download JJF 1925-2021
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
JJF 2425-2026 — Calibration specification for carbon dioxide incubators
JJF 1183-2025 — Calibration Specification for Temperature Transmitters
JJF 1282-2025 — Calibration Specification for Time Delay Relays
JJF 1925-2021
$390.00