JJF 1905-2021Calibration Specification for Magnetic Flux Meters (English PDF)
磁通计校准规范
This is a limited preview
Buy now to download the full PDF (15 pages)
Need quotation for your management?
Get a formal quotation for JJF 1905-2021 in 30 minutes.
Issued by
State Administration for Market Regulation
Level / Type
National · Mandatory
Issue date
February 23, 2021
Implementation date
August 23, 2021
Scope
JJF 1905-2021 is the English-translated version of 磁通计校准规范.
JJF 1905-2021 is the Chinese standard "Calibration Specification for Magnetic Flux Meters". Issued on 2021-02-23, in force from 2021-08-23.
Found what you need?
Document preview — JJF 1905-2021
Contents
- Introduction (II)
- 1 Scope (1)
- 2 Reference documents (1)
- 3 Overview (1)
- 4 Measurement characteristics (1)
- 4.1 Zero drift (1)
- 4.2 Magnetic flux indication error (1)
- 5 Calibration conditions (1)
- 5.1 Environmental conditions (1)
- 5.2 Measurement standards and other equipment (1)
- 6 Calibration items and calibration methods (3)
- 6.1 Calibration items (3)
- 6.2 Calibration method (3)
- 7 Expression of calibration results (6)
- 8 Recalibration interval (7)
- Appendix A Measurement Uncertainty Evaluation Example of Calibration Error of Indication (8)
- Appendix B Calibration Original Record Format (13)
- Appendix C Internal Page Format of Calibration Certificate (15)
- Appendix D Correction method for the influence of external loop resistance on measurement results (17)
1 Scope (1)
2 Reference documents (1)
3 Overview (1)
4 Measurement characteristics (1)
4.1 Zero drift (1)
4.2 Magnetic flux indication error (1)
5 Calibration conditions (1)
5.1 Environmental conditions (1)
5.2 Measurement standards and other equipment (1)
6 Calibration items and calibration methods (3)
6.1 Calibration items (3)
6.2 Calibration method (3)
7 Expression of calibration results (6)
8 Recalibration interval (7)
Appendix A Measurement Uncertainty Evaluation Example of Calibration Error of Indication (8)
Appendix B Calibration Original Record Format (13)
Appendix C Internal Page Format of Calibration Certificate (15)
Appendix D Correction method for the influence of external loop resistance on measurement results (17)
Introduction
JJF 1071 "Rules for the Compilation of National Metrology and Calibration Specifications", JJF 1001 "General Metrology Terms and Definitions" and JJF 1059.1 "Evaluation and Expression of Measurement Uncertainty" together constitutes a basic series that supports the formulation of this specification.
This specification replaces JJG317-1983 "Trial Verification Regulations for Magnetic Flux Meters". Compared with JJG317-1983, the main the technical changes are as follows.
---Added reference files;
---Metering characteristics increase the zero point drift;
---The calibration method adds the volt-second method;
The previous releases of this specification are as follows.
---JJG317-1983.
Fluxmeter calibration specification
1 Scope
This specification applies to the measurement range of 0.1mWb~10Wb, and the maximum allowable error is ±(0.1%~5.0%) Calibration of the fluxmeter.
2 Reference documents the following documents are cited in this specification.
JJF 1013 Commonly used terms and definitions of magnetic measurement for dated reference documents, only the dated version applies to this specification; for all undated references the latest version (including all amendments) is applicable to this specification.
3 overview
Fluxmeter is used for the measurement of magnetic flux. According to the display mode, it is divided into pointer type and digital type. The digital type is divided into digital Integral and analog integral.
When the magnetic flux linked in the measuring coil of the fluxmeter changes, an induced electromotive force will be generated at both ends of the measuring coil.
The flux meter measures the magnetic flux by integrating the induced electromotive force.
4 Metrological characteristics
4.1 Zero drift
Under the condition that the input terminal of the flux meter coil is short-circuited, the maximum absolute value of the magnetic flux indication within 60s.
4.2 Magnetic flux indication error the magnetic flux indication error of the fluxmeter does not exceed its nominal maximum allowable error, and the maximum allowable error range is generally ±(0.1%~5.0%).
Note. The above indicators are not used for eligibility judgment and are for reference only.
5 Calibration conditions
5.1 Environmental conditions the environmental conditions and requirements are as follows.
a) Ambient temperature. (20±5)°C;
b) Relative humidity. <=80%;
c) Power supply. (220±22)V, (50±1)Hz;
d) There is no vibration, shock or other electromagnetic interference that obviously affects the calibration.
5.2 Measurement standards and other equipment
5.2.1 Measurement standards and other equipment used by each calibration method the measurement standards and other equipment used by each calibration method are shown in Table 1.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 15 pages — is available in the English PDF.
How to Download JJF 1905-2021
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
JJF 2425-2026 — Calibration specification for carbon dioxide incubators
JJF 1183-2025 — Calibration Specification for Temperature Transmitters
JJF 1282-2025 — Calibration Specification for Time Delay Relays
JJF 1905-2021
$300.00