JJF 1812-2020Calibration Specification for Apparatus to Measure Water Permeability of Concrete (English PDF)
混凝土抗渗仪校准规范
This is a limited preview
Buy now to download the full PDF (9 pages)
Need quotation for your management?
Get a formal quotation for JJF 1812-2020 in 30 minutes.
Issued by
State Administration for Market Regulation
Level / Type
National · Mandatory
Issue date
January 1, 2020
Implementation date
April 17, 2020
Scope
JJF 1812-2020 is the English-translated version of 混凝土抗渗仪校准规范.
JJF 1812-2020 is the Chinese standard "Calibration Specification for Apparatus to Measure Water Permeability of Concrete".
Found what you need?
Document preview — JJF 1812-2020
National Standard of the People's Republic of China
- ICS
- 17.100
Contents
- Introduction (II)
- 1 Scope (1)
- 2 Reference documents (1)
- 3 Overview (1)
- 4 Measurement characteristics (2)
- 4.1 Pressure indication error (2)
- 5 Calibration conditions (2)
- 5.1 Environmental conditions (2)
- 5.2 Measurement standards (2)
- 6 Calibration items and calibration methods (2)
- 6.1 Functional inspection (2)
- 6.2 System leak check (2)
- 6.3 Pressure indication error (3)
- 7 Calibration results (3)
- 7.1 Calibration data processing (3)
- 7.2 Calibration certificate (3)
- 7.3 Evaluation of calibration uncertainty (3)
- 8 Recalibration interval (3)
- Appendix A Impermeability Meter Calibration Record Format (4)
- Appendix B Internal page format of calibration certificate (5)
- Appendix C Uncertainty Evaluation Example of Pressure Indication Calibration (7)
1 Scope (1)
2 Reference documents (1)
3 Overview (1)
4 Measurement characteristics (2)
4.1 Pressure indication error (2)
5 Calibration conditions (2)
5.1 Environmental conditions (2)
5.2 Measurement standards (2)
6 Calibration items and calibration methods (2)
6.1 Functional inspection (2)
6.2 System leak check (2)
6.3 Pressure indication error (3)
7 Calibration results (3)
7.1 Calibration data processing (3)
7.2 Calibration certificate (3)
7.3 Evaluation of calibration uncertainty (3)
8 Recalibration interval (3)
Appendix A Impermeability Meter Calibration Record Format (4)
Appendix B Internal page format of calibration certificate (5)
Appendix C Uncertainty Evaluation Example of Pressure Indication Calibration (7)
Introduction
This specification is based on JJF 1071-2010 "Rules for the Compilation of National Metrology and Calibration Specifications", JJF 1001-2011 "General Measurement terms and definitions", JJF 1059.1-2012 "Evaluation and Expression of Measurement Uncertainty" and JJF 1094-2002 the "Quality Evaluation of Measuring Instruments" is a basic series of specifications.
This specification refers to JJG52-2013 "Elastic element type general pressure gauge, pressure vacuum gauge and vacuum gauge", JJG875-2005 "Digital Pressure Gauge" and JG/T 249-2009 "Concrete Impermeability Tester" are compiled.
This specification is released for the first time.
Calibration Specification for Concrete Impermeability Tester
1 Scope
This specification applies to the calibration of concrete impermeability instruments with a measuring range of (0~4) MPa, as well as the calibration of mortar impermeability instruments.
2 Reference documents this specification refers to the following documents.
JJG52-2013 elastic element type general pressure gauge, pressure vacuum gauge and vacuum gauge JJG875 Digital Pressure Gauge JG/T 249-2009 Concrete Impermeability Tester for dated reference documents, only the dated version applies to this specification; for undated reference documents, the latest version (including all amendments) is applicable to this specification.
3 overview
Concrete Impermeability Tester (hereinafter referred to as Impermeability Tester) is used to detect the waterproof performance of concrete after hardening, and to determine its impermeability marking measuring instrument. The impermeability meter applies Pascal's law, and the pressure at all points in the static sealed container is equal everywhere the principle of (the water level difference is negligible), the whole system is pressurized with a water pump, and the pressure gauge and electrical controller are used to the pressure is indicated and controlled to realize the penetration of water from bottom to top into the test piece press-fitted in the test mold to determine the test the anti-permeability performance of the piece and the instrument to calculate its anti-permeability mark.
The impermeability instrument is composed of pressure control instruments, pressure sensors and other main measuring instruments, valves, and connecting pipelines, and is equipped with There are 6 mold bases for installation and test molds. The impermeability instrument is pressurized by an electric plunger pump, and is connected to the water supply capacity through a hydraulic pipeline.
The pressure gauges and electrical control systems in the pipeline are connected to each pressure point of the anti-permeability instrument.
The adjustment control can make the impermeability tester perform a constant pressure test within the specified pressure range of (0.1~4) MPa. anti-the 6 test mold bases included in the osmometer, the middle hole of each mold bottom is used as a water pressure output port, and each output port can enter Line independent control and measurement. The structure of the impermeability instrument is shown in Figure 1.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 9 pages — is available in the English PDF.
How to Download JJF 1812-2020
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
JJF 2425-2026 — Calibration specification for carbon dioxide incubators
JJF 1183-2025 — Calibration Specification for Temperature Transmitters
JJF 1282-2025 — Calibration Specification for Time Delay Relays