JJF 1740-2019Calibration Specification for Cable and Antenna Analyzers (English PDF)

天馈线测试仪校准规范

Preview PDF

This is a limited preview

Buy now to download the full PDF (13 pages)

Need quotation for your management?

Get a formal quotation for JJF 1740-2019 in 30 minutes.

Issued by

State Administration for Market Regulation

Level / Type

National · Mandatory

Issue date

January 1, 2019

Implementation date

December 27, 2019

Scope

JJF 1740-2019 is the English-translated version of 天馈线测试仪校准规范.

JJF 1740-2019 is the Chinese standard "Calibration Specification for Cable and Antenna Analyzers".

Found what you need?

Document preview — JJF 1740-2019

National Standard of the People's Republic of China

ICS
17.220

Contents

  • Introduction (II)
  • 1 Scope (1)
  • 2 Terminology (1)
  • 3 Overview (1)
  • 4 Measurement characteristics (1)
  • 4.1 Output signal frequency (1)
  • 4.2 Output power (1)
  • 4.3 Voltage standing wave ratio (1)
  • 4.4 Attenuation (1)
  • 4.5 Fault location distance (1)
  • 4.6 RF power (1)
  • 5 Calibration conditions (1)
  • 5.1 Environmental conditions (1)
  • 5.2 Calibration equipment (2)
  • 6 Calibration items and calibration methods (2)
  • 6.1 Appearance and work normality inspection (3)
  • 6.2 Output signal frequency (3)
  • 6.3 Output power (3)
  • 6.4 Voltage standing wave ratio (3)
  • 6.5 Attenuation (4)
  • 6.6 Fault location distance (4)
  • 6.7 RF power (4)
  • 7 Expression of calibration results (5)
  • 8 Recalibration interval (5)
  • Appendix A Original Record Format (6)
  • Appendix B. Internal page format of the calibration certificate (8)
  • Appendix C Examples of Uncertainty Evaluation of Calibration Results of Main Items (10)

1 Scope (1)

2 Terminology (1)

3 Overview (1)

4 Measurement characteristics (1)

4.1 Output signal frequency (1)

4.2 Output power (1)

4.3 Voltage standing wave ratio (1)

4.4 Attenuation (1)

4.5 Fault location distance (1)

4.6 RF power (1)

5 Calibration conditions (1)

5.1 Environmental conditions (1)

5.2 Calibration equipment (2)

6 Calibration items and calibration methods (2)

6.1 Appearance and work normality inspection (3)

6.2 Output signal frequency (3)

6.3 Output power (3)

6.4 Voltage standing wave ratio (3)

6.5 Attenuation (4)

6.6 Fault location distance (4)

6.7 RF power (4)

7 Expression of calibration results (5)

8 Recalibration interval (5)

Appendix A Original Record Format (6)

Appendix B. Internal page format of the calibration certificate (8)

Appendix C Examples of Uncertainty Evaluation of Calibration Results of Main Items (10)

Introduction

This specification is based on JJF 1071-2010 "Rules for Compiling National Measurement and Calibration Specifications" and JJF 1059.1-2012 Preparation of "Measurement Uncertainty Evaluation and Representation".

This specification specifies the frequency accuracy, output power, voltage standing wave ratio, and attenuation of the output signal of the antenna feeder tester the calibration method of the quantity, fault location distance, RF power, and the measurement uncertainty of the calibration items are given in the appendix Evaluation example.

This specification is the first release.

Calibration specification for antenna feeder tester

1 Scope

This specification applies to the calibration of antenna feeder testers.

2 Terminology

Distance to fault

The antenna feeder tester determines the fault location according to the reflection coefficient or the maximum value of the voltage standing wave ratio.

The distance of the location is the fault location distance.

3 Overview

The antenna feeder tester can quickly evaluate the condition of the transmission line and antenna system, and measure the reflection coefficient or voltage standing wave Ratio, cable loss and fault location distance. Antenna feeder tester is a commonly used measuring instrument in communication, widely used in Wireless system installation, configuration, maintenance and failure analysis.

Antenna feeder tester usually consists of signal source unit, signal separation unit, receiver unit, digital control and display unit Composition. The port impedance is generally 50ohm.

4 Metering characteristics

4.1 Output signal frequency Range. 2MHz ~ 18GHz, maximum allowable error. ± (1 × 10-6 ~ 7.5 × 10-5).

4.2 Output power

Range. -30dBm ~ 20dBm, the maximum allowable error. ± 2dB.

4.3 Voltage standing wave ratio Range. VSWR > 1, maximum allowable error. ± 20%.

4.4 Attenuation

Range. 0dB ~ 30dB, maximum allowable error. ± 1.5dB.

4.5 Fault location distance Range. 0m ~ 1500m, maximum allowable error. ± (3% × physical length 0.1m).

4.6 RF power

Range. -50dBm ~ 20dBm, maximum allowable error. ± 2dB.

Note. The above technical indicators are not used to determine eligibility, only for reference.

5 Calibration conditions

5.1 Environmental conditions

a) Ambient temperature. (23 ± 5) °C;

b) Relative humidity. <=80%;

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 13 pages — is available in the English PDF.

How to Download JJF 1740-2019

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Fast delivery

    After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
13 pages
Fast delivery
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

JJF 1740-2019

$365.00