JJF 1674-2017Calibration Specification for Alarmer Detectors of Benzene (English PDF)

Preview PDF

This is a limited preview

Buy now to download the full PDF (11 pages)

Need quotation for your management?

Get a formal quotation for JJF 1674-2017 in 30 minutes.

Issued by

General Administration of Quality Supervision, Inspection and Quarantine

Level / Type

National · Mandatory

Issue date

November 20, 2017

Implementation date

February 20, 2018

Scope

JJF 1674-2017 (Calibration Specification for Alarmer Detectors of Benzene) is available as an English-translated PDF.

JJF 1674-2017 — This standard applies to the calibration of benzene gas detection alarms (hereinafter referred to as instruments) whose measurement upper limit does not exceed 100 µmol/mol.

Found what you need?

Document preview — JJF 1674-2017

National Standard of the People's Republic of China

ICS
17.020
Classification
A 61

Issued by: General Administration of Quality Supervision, Inspection and Quarantine

Contents

  • Introduction (II)
  • 1 Scope (1)
  • 2 Overview (1)
  • 3 Measurement Characteristics (1)
  • 3.1 Indication error (1)
  • 3.2 Repeatability (1)
  • 3.3 Response Time (1)
  • 3.4 Alarm function (1)
  • 3.5 Drift (1)
  • 4 Calibration conditions (1)
  • 4.1 Environmental conditions (1)
  • 4.2 Calibration of measuring instruments and ancillary equipment (2)
  • 5 Calibration Items and Calibration Methods (2)
  • 5.1 Indication error (2)
  • 5.2 Repeatability (2)
  • 5.3 Response Time (3)
  • 5.4 Alarm function (3)
  • 5.5 Drift (3)
  • 6 Calibration Results Expression (3)
  • 7 Re-school interval (4)
  • Appendix A Benzene gas detection alarm calibration records (5)
  • Appendix B Certificate page format (6)
  • Appendix C Calibration error evaluation results of the uncertainty rating (7)
  • Appendix D Relative indication error calibration results of the uncertainty assessment (10)

1 Scope (1)

2 Overview (1)

3 Measurement Characteristics (1)

3.1 Indication error (1)

3.2 Repeatability (1)

3.3 Response Time (1)

3.4 Alarm function (1)

3.5 Drift (1)

4 Calibration conditions (1)

4.1 Environmental conditions (1)

4.2 Calibration of measuring instruments and ancillary equipment (2)

5 Calibration Items and Calibration Methods (2)

5.1 Indication error (2)

5.2 Repeatability (2)

5.3 Response Time (3)

5.4 Alarm function (3)

5.5 Drift (3)

6 Calibration Results Expression (3)

7 Re-school interval (4)

Appendix A benzene gas detection alarm calibration records (5)

Appendix B certificate page format (6)

Appendix C Calibration error evaluation results of the uncertainty rating (7)

Appendix D Relative indication error calibration results of the uncertainty assessment (10)

Introduction

The specification measurement characteristics of the reference to the GB 12358-2006 "workplace environmental gas detection alarm General technical requirements ", GB 50493-2009" petrochemical flammable gas and toxic gas detection and alarm design specifications " and other technical regulations.

This specification is first released.

Benzene gas detection alarm calibration specification

1 Scope

This standard applies to the measurement limit of not more than 100µmol/mol benzene gas detection alarm (hereinafter referred to as instrument Device) calibration.

2 Overview

Instrument is mainly used for workplace benzene gas detection and other environments, usually photoionization (PID) detection principle and semiconductor detection theory. Instrument mainly by the detection components, amplification circuits, alarm systems, monitors and other components.

Photon ionization (PID) detection principle refers to the gas under test in the ionization of ultraviolet light generated negative charge Charged electrons and positively charged ions form a weak current under the action of an electric field. The magnitude of this current is positively correlated with the concentration of gas.

Now the detection of gas concentrations. Semiconductor detection principle is the use of semiconductor gas sensors contact with the gas, its conductivity and so on the physical properties change to detect the concentration of the characteristic gas.

Instrument types are fixed and portable, sampling methods are diffused and inhaled.

3 measurement characteristics 3.1 indication error Indication error does not exceed the provisions of Table 1.

Table 1 shows the value of error

Measuring range/(µmol/mol) Indication error 0 ~ 10 ± 1.0µmol/mol > 10 ~ 100 ± 10%

3.2 Repeatability

Repeatability of not more than 3%.

3.3 Response time

Response time is not greater than 60s.

3.4 alarm function

Has the alarm function of the instrument, in its measurement range should have the alarm set value, when the instrument reaches the alarm set value Set value, should be able to automatically alarm.

3.5 Drift

3.5.1 Zero drift ± 3% FS

3.5.2 Range Drift. ± 5% FS Note. The above indicators are not for the qualification, for reference only.

4 calibration conditions

4.1 Environmental conditions

4.1.1 Ambient temperature (0 ~ 40) °C.

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 11 pages — is available in the English PDF.

How to Download JJF 1674-2017

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
11 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

JJF 1674-2017

$235.00