JJF 1609-2017Calibration Specification for Residual Chlorine Meters (English PDF)
This is a limited preview
Buy now to download the full PDF (9 pages)
Need quotation for your management?
Get a formal quotation for JJF 1609-2017 in 30 minutes.
Issued by
General Administration of Quality Supervision, Inspection and Quarantine
Level / Type
National · Mandatory
Issue date
February 28, 2017
Implementation date
May 28, 2017
Scope
JJF 1609-2017 (Calibration Specification for Residual Chlorine Meters) is available as an English-translated PDF.
JJF 1609-2017 — This standard applies to the colorimetric principle of chlorine analyzer calibration, including the determination of residual chlorine in water and (and) the total residual chlorine analyzer.
Found what you need?
Document preview — JJF 1609-2017
National Standard of the People's Republic of China
- ICS
- 17.020
- Classification
- A 61
Issued by: General Administration of Quality Supervision, Inspection and Quarantine
Contents
- Introduction (II)
- 1 Scope (1)
- 2 Terms and units of measurement (1)
- 2.1 Free residual chlorine (1)
- 2.2 Compounds Residual Chlorine (1)
- 2.3 Residual chlorine (total residual chlorine) (1)
- 3 Overview (1)
- 4 Metering characteristics (1)
- 5 Calibration conditions (2)
- 5.1 Environmental conditions (2)
- 5.2 Reference materials and other equipment (2)
- 6 Calibration items and calibration methods (2)
- 6.1 Indication error (2)
- 6.2 Repeatability (2)
- 7 Expression of calibration results (3)
- 8 Re-school time interval (3)
- Appendix A Examples of Uncertainty Evaluation of Indication Error (4)
- Appendix B Residual Chlorine Detector Recording Format (for reference) (8)
- Appendix C Calibration Certificate (End page) Format (for reference) (9)
1 Scope (1)
2 Terms and units of measurement (1)
2.1 Free residual chlorine (1)
2.2 Compounds Residual Chlorine (1)
2.3 Residual chlorine (total residual chlorine) (1)
3 Overview (1)
4 Metering characteristics (1)
5 Calibration conditions (2)
5.1 Environmental conditions (2)
5.2 Reference materials and other equipment (2)
6 Calibration items and calibration methods (2)
6.1 Indication error (2)
6.2 Repeatability (2)
7 Expression of calibration results (3)
8 Re-school time interval (3)
Appendix A Examples of Uncertainty Evaluation of Indication Error (4)
Appendix B Residual Chlorine Detector Recording Format (for reference) (8)
Appendix C Calibration Certificate (End page) Format (for reference) (9)
Introduction
This specification is based on JJF 1071-2010 "national measurement calibration standard preparation rules", JJF 1059.1-2012 "test Assessment and representation of uncertainty, "JJF 1001-2011" General measurement terms and definitions "provisions of the preparation, reference ASTM D1253-2014 "Standard Test Method for Residual Chlorine in Water" (StandardTestMethodforRe-sidualChlorinein Water), GB/T 14424-2008 "Determination of residual chlorine in industrial circulating cooling water" HJ586-2010 "Determination of free chlorine and total chlorine in water, N, N-diethyl-1,4-phenylenediamine spectrophotometry" Sub-content.
This specification is the first release.
Calibration standard for chlorine analyzer
1 Scope
The calibration standard applies to the colorimetric principle of the residual chlorine analyzer calibration, including the determination of residual chlorine in water or (and) Total residual chlorine analyzer.
2 terms and units of measurement 2.1 free residual chlorine Refers to chlorine present in the form of hypochlorous acid, hypochlorite ion and dissolved monochlorium chloride.
2.2 combined residual chlorine residualresidualchlorine Refers to chlorine in the form of chloramines and organochlorines.
2.3 residual chlorine (total residual chlorine) Refers to chlorine in the form of "free residual chlorine" or "chemical residual chlorine", or both.
Residual chlorine units. mg/L.
3 Overview
Residual chlorine refers to the water after disinfection of residual chlorine in water, is an important indicator of water quality evaluation, it is mainly used for self Water production, medical wastewater treatment and other areas of water quality monitoring. Residual chlorine analyzer is the use of colorimetric method of work, the root Quantitative analysis of the absorbance of the colored material produced at the specific wavelength after the reaction of residual chlorine in the water with the developer is carried out.
The measurement principle is based on the Lambert-Beer light absorption law.
A = -lg
I0 = -lgT = kcL
Where.
A --- absorbance;
I0 --- incident light intensity;
I - transmitted light intensity;
T --- transmittance;
k - absorption coefficient;
c - the amount of residual chlorine in the sample;
L - Light path through the sample.
The instrument is mainly composed of light source, sample room, detector and data processing unit and so on, as shown in Fig.
Light source -> sample chamber -> detector -> data processing unit
Figure 1 Schematic diagram of the structure of the chlorine detector
4 measurement characteristics 4.1 indication error. no more than ± 10%.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 9 pages — is available in the English PDF.
How to Download JJF 1609-2017
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
JJF 2425-2026 — Calibration specification for carbon dioxide incubators
JJF 1183-2025 — Calibration Specification for Temperature Transmitters
JJF 1282-2025 — Calibration Specification for Time Delay Relays
Secure payment via Stripe
Payments accepted
JJF 1609-2017
$289.00