National Standard of the People's Republic of China
- ICS
- 17.180
- Classification
- A 60
Issued by: General Administration of Quality Supervision, Inspection and Quarantine
Contents
- Introduction (II)
- 1 Scope (1)
- 2. References (1)
- 3 Overview (1)
- 4 Metrological characteristics (2)
- 4.1 Instrument indication error (2)
- 4.2 Repeatability measuring instrument (2)
- 4.3 Instrument stability (2)
- 4.4 Instrument 24h run stability (2)
- 4.5 Instrument phosphate anti-interference performance (2)
1 Scope (1)
2 References (1)
3 Overview (1)
4 Metrological characteristics (2)
4.1 Instrument indication error (2)
4.2 Repeatability measuring instrument (2)
4.3 Instrument stability (2)
4.4 Instrument 24h run stability (2)
4.5 Instrument phosphate anti-interference performance (2)
Calibration Condition 5 (2)
5.1 Calibration Environment (2)
5.2 Standard devices and other equipment (2)
6 Calibration items and calibration methods (3)
6.1 Calibration before the test (3)
6.2 Instrument indication error (3)
6.3 Instrument measurement repeatability (3)
6.4 Instrument stability (4)
6.5 Instrument 24h run stability (4)
6.6 Instrument phosphate anti-interference performance (4)
7 Calibration Results The expression (4)
8 Recalibration interval (5)
Appendix A calibration reagent and solution preparation method (6)
Appendix B recommended calibration record format (7)
Appendix C Recommended calibration certificate-page format (9)
Appendix D Evaluation of Uncertainty in Measurement Example (10)
Introduction
This specification based JJF 1071-2010 "National Calibration Specifications writing rules", JJF 1001 "universal metering Terms and definitions ", JJF 1059.1-2012" Evaluation and Expression of Uncertainty in Measurement ", and referring to (part) GB/T 12149-2007 "industrial circulating cooling water and boiler water Determination of Silicon", DL/T 677-2009 "hair Power line chemical instrumentation test procedures "requirement developed.
This specification calibrated main items are.
--- Instrument indication error;
--- Instrument measurement repeatability;
--- Instrument stability;
--- Instrument 24h operational stability;
--- Instrument phosphate anti-interference performance.
The specification for the first time to develop.
Silicate analyzer calibration specification
1 Scope
This instruction applies to silicate analyzer calibration.
2. References
This specification references the following documents.
JJF 1094 measuring instrument characteristics evaluation Laboratory use specifications and test methods GB/T 6682 Analysis GB/T 12149 Determination of industrial circulating cooling water and boiler water in silicon DL/T 677 meter inspection of chemical plants online for dated references, only the dated edition applies to this specification; undated reference documents Member, the latest edition (including any amendments) applies to this specification.
3 Overview
Silicate analyzer (hereinafter referred to as the instrument) is used to measure the water content of silicate analysis instrument that measures primary Water soluble silicon processing under certain conditions, the reagent reacts and color, according to Lambert - Beer law, when Bundle of parallel monochromatic light through the light-absorbing substance containing the colored solution, part of the light is absorbed when the absorbing layer thickness is not Varying proportional to the concentration and the absorbance of light-absorbing substance by photoelectric detection device converts the optical signal into an electrical signal by After the microprocessor, the display silicate content. Lambert - Beer law mathematical expression is.
A = -lg
lgT = klc
Where.
A --- absorbance substances;
I --- transmitted monochromatic light intensity;
Monochromatic light intensity I0 --- incident;
T --- substance transmittance;
Extinction coefficient k --- substances;
l --- analyte optical path substances;
c --- concentration of the substance.
Instruments mainly by the light source system, the spectrometer, the sample chamber, the detection system, etc., see Schematic figure 1.
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 12 pages — is available in the English PDF.
How to Download JJF 1539-2015
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
JJF 2425-2026 — Calibration specification for carbon dioxide incubators
JJF 1183-2025 — Calibration Specification for Temperature Transmitters
JJF 1282-2025 — Calibration Specification for Time Delay Relays
Secure payment via Stripe
Payments accepted
JJF 1539-2015
$235.00