JJF 1531-2015Calibration specification for fourier transform mass spectrometers (English PDF)
This is a limited preview
Buy now to download the full PDF (12 pages)
Need quotation for your management?
Get a formal quotation for JJF 1531-2015 in 30 minutes.
Issued by
General Administration of Quality Supervision, Inspection and Quarantine
Level / Type
National · Mandatory
Issue date
June 15, 2015
Implementation date
September 15, 2015
Scope
JJF 1531-2015 (Calibration specification for fourier transform mass spectrometers) is available as an English-translated PDF.
JJF 1531-2015 — This Standard applies to the Fourier transform mass spectrometer calibration.
Found what you need?
Document preview — JJF 1531-2015
National Standard of the People's Republic of China
- ICS
- 17.02
- Classification
- A 61
Issued by: General Administration of Quality Supervision, Inspection and Quarantine
Contents
- Introduction (II)
- 1 Scope (1)
- 2. References (1)
- 3 Terminology and units of measurement (1)
- 3.1 The term (1)
- 3.2 Units of measurement (1)
- 4 Overview (1)
- 4.1 Basic Principle (1)
- 4.2-instrument (2)
- 5 Metrological characteristics (2)
1 Scope (1)
2 References (1)
3 Terminology and units of measurement (1)
3.1 The term (1)
3.2 Units of measurement (1)
4 Overview (1)
4.1 Basic Principle (1)
4.2-instrument (2)
5 Metrological characteristics (2)
Calibration Condition 6 (2)
6.1 Environmental conditions (2)
6.2 Standard materials and reagents (2)
7 Calibration items and calibration methods (3)
7.1 Indication error (3)
SNR 7.2 (3)
7.3 Resolving power (3)
7.4 Repeatability (4)
Drift 7.5 (4)
8 Calibration results expression (4)
9 Recalibration interval (5)
Appendix A preparation and use of reference material solution (6)
Appendix B Calibration record format (7)
Appendix C calibration certificate page format (10)
Uncertainty Evaluation of Indication Error Examples Appendix D (11)
Introduction
This specification based on the national measurement specifications JJF 1071-2010 "national measurement standards to prepare rules" JJF 1001-2011 "common measurement terms and definitions" JJF 1059.1-2012 "Evaluation and Expression of Uncertainty in Measurement" prepared by.
Development of this specification primarily with reference to GB/T 6041-2002 "General mass spectrometry method" and other documents.
This specification is the first release.
Fourier transform mass spectrometer calibration specification
1 Scope
This standard applies to the Fourier transform mass spectrometer calibration.
2. References
This specification references the following documents.
GB/T 6041-2002 General Mass Spectrometry for dated references, only the dated edition applies to this specification; undated reference documents Member, the latest edition (including any amendments) applies to this specification.
3 terminology and units of measurement
3.1 Terms
3.1.1 Sensitivity [GB/T 6041-20022.2] Under specified conditions, the response to the value of a mass peak, the instrument unit generated by the sample.
3.1.2 The ability to distinguish resolution in a given sample conditions, instrument two adjacent mass peaks discriminatory power. Two adjacent high-mass spectrometry when the peak, the peak is not more than 10% of the peak height, it is defined as can be distinguished.
Note. The ability to distinguish two peaks of representation when R peak at 10% peak height, the quality of the two peaks represented by the average value of quality the ratio of the amount of the difference. When describing the use of mass spectral peak mass to charge ratio, listed in the corresponding resolution of the data is appropriate, because Different instruments, while its resolving power varies with different mass to charge ratio.
3.1.3 SNR signaltonoiseratio [GB/T 6041-20022.9] in mass spectrometry, the signal strength and noise intensity ratio.
3.2 units of measurement Atomic mass units (u) atomicmassunit [JJF 1164-20063.6]
4 Overview
4.1 Fundamentals
Fourier transform mass spectrometer according to different mass analyzers are divided into two categories. Fourier transform ion cyclotron resonance mass spectrometry Fourier transform instrument and Orbitrap mass spectrometer.
Fourier transform ion cyclotron resonance mass spectrometer is a core component in a high magnetic field of the ion cyclotron resonance trap (ICRiontrap) --- analysis room, which can capture and store ions in a high magnetic field. The basic principle is to leave Ion source produces an ion beam is introduced ICR ion trap under the action of magnetic force, all the ions are beginning along well Center to do a small radius of cyclotron motion. If the magnetic field strength B is fixed, and only the cyclotron frequency of the ion mass to charge ratio have and independent of the customs and ion kinetic energy. When the frequency of the high frequency signal is applied and the frequency of a mass to charge ratio of ions in a cyclotron motion when delivered, the ions will absorb the energy of a high-frequency electrical signals, the cyclotron radius of motion increases. When stopping the high-frequency excitation
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 12 pages — is available in the English PDF.
How to Download JJF 1531-2015
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
JJF 2425-2026 — Calibration specification for carbon dioxide incubators
JJF 1183-2025 — Calibration Specification for Temperature Transmitters
JJF 1282-2025 — Calibration Specification for Time Delay Relays
Secure payment via Stripe
Payments accepted
JJF 1531-2015
$460.00