JJF 1383-2012Calibration Specification for Portable Blood Glucose Meters (English PDF)
便携式血糖分析仪校准规范
This is a limited preview
Buy now to download the full PDF (11 pages)
Need quotation for your management?
Get a formal quotation for JJF 1383-2012 in 30 minutes.
Issued by
General Administration of Quality Supervision, Inspection and Quarantine
Level / Type
National · Mandatory
Issue date
December 21, 2012
Implementation date
March 21, 2013
Scope
JJF 1383-2012 is the English-translated version of 便携式血糖分析仪校准规范.
JJF 1383-2012 - This standard provides a portable blood glucose analyzer metrological characteristics, calibration conditions and calibration methods. This specification is for in vitro monitoring of blood sugar levels portable blood glucose analyzer.
Found what you need?
Document preview — JJF 1383-2012
Contents
- 1 Scope (1)
- 2. References (1)
- 3 Terms and Definitions (1)
- 3.1 Carryover (1)
- 4 Overview (1)
- 5 Metrological characteristics (1)
- 5.1 Indication error (1)
- 5.2 Measurement repeatability (2)
- 5.3 Linearity error (2)
- 5.4 Carryover (2)
1 Scope (1)
2 References (1)
3 Terms and Definitions (1)
3.1 Carryover (1)
4 Overview (1)
5 Metrological characteristics (1)
5.1 Indication error (1)
5.2 Measurement repeatability (2)
5.3 Linearity error (2)
5.4 Carryover (2)
Calibration Condition 6 (2)
6.1 Environmental conditions (2)
6.2 Standards and other measuring equipment (2)
7 Calibration items and calibration methods (2)
7.1 Appearance and working properly check (2)
7.2 Preparation before calibration (2)
7.3 Indication error (2)
7.4 Measurement repeatability (3)
7.5 Linearity error (3)
7.6 Carryover (3)
8 Calibration results expression (4)
9 Recalibration interval (4)
Appendix A portable blood glucose analyzer calibration record format (5)
Appendix B portable blood glucose analyzer calibration certificate (within the page) format (6)
Appendix C portable blood glucose analyzer Indication Error of Measurement Uncertainty (7)
Portable blood glucose analyzer calibration specification
1 Scope
This specification defines the metrological characteristics of a portable blood glucose analyzer calibration conditions and calibration methods. The specification for the body External monitoring of blood glucose content portable blood glucose analyzer.
2. References
JJF 1001-2011 common measurement terms and definitions JJF 1059-1999 Measurement Uncertainty and representation GB 6682-2008 analytical laboratory use specifications and test methods GB/T 19634-2005 vitro diagnostic test systems for self blood glucose monitoring system General technical requirements WS/T 226-2002 portable blood glucose meter blood glucose measurement guide for dated references, only the dated edition applies to this specification; undated reference documents Member, the latest edition (including any amendments) applies to this specification.
3 Terms and Definitions 3.1 carryover carry-overrate by the measurement system will carry a test sample reaction to detect the analyte another sample of the reaction discrete amount by this error affected the performance of another quantity of the test sample. Usually taken given the known nominal value of high and low sample Measured value of the sample cross many times, calculated using a specific formula.
4 Overview
Portable blood glucose analyzer according to the measure of sugar technology can be divided electrochemical test and the light reflective technology test two categories.
The former is an enzyme reacts with glucose to produce an electronic current through the counting device, read the number of electrons; microcontroller according to Correspondence between the current size and the blood glucose concentration in the blood glucose concentration is calculated and displayed. The latter is by Contact with the blood test reagent block blood glucose test strips, and color change agent blocks react portable blood glucose by irradiating the test wells analyzer reagent strip block strip on the platform, and measuring the reflected light, thereby obtaining a reagent strip Built-in microcontroller based on the color of the corresponding block and reagent strip blood glucose concentration; changes in color blocks Relationship between blood glucose concentration values calculated and displayed.
Portable blood glucose analyzer by the analysis system, microcontrollers, data storage, LCD display, data output interface Ports and other components.
Rapid measurement portable blood glucose analyzer is mainly used in fresh capillary whole blood glucose content can also be used Venous whole blood, rapid measurement of serum (plasma) of the glucose content.
5 Metrological characteristics
5.1 Indication error
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 11 pages — is available in the English PDF.
How to Download JJF 1383-2012
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Fast delivery
After payment you'll receive an email confirmation. The English PDF is sent to the same address, usually within minutes and at the latest within 1 business day.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
JJF 2425-2026 — Calibration specification for carbon dioxide incubators
JJF 1183-2025 — Calibration Specification for Temperature Transmitters
JJF 1282-2025 — Calibration Specification for Time Delay Relays
JJF 1383-2012
$255.00