Valid

JJF 1150-2006Calibration specification for relative spectral responsivity for photoelectric detectors (English PDF)

Preview PDF

This is a limited preview

Buy now to download the full PDF (11 pages)

Need quotation for your management?

Get a formal quotation for JJF 1150-2006 in 30 minutes.

Issued by

General Administration of Quality Supervision, Inspection and Quarantine

Level / Type

National · Mandatory

Issue date

May 23, 2006

Implementation date

November 23, 2006

Scope

JJF 1150-2006 (Calibration specification for relative spectral responsivity for photoelectric detectors) is available as an English-translated PDF.

JJF 1150-2006 — This standard applies to the photodetector, light meter probe luminance meter probes and other optical test equipment used by the photodetector relative spectral responsivity calibration.

Found what you need?

Document preview — JJF 1150-2006

National Standard of the People's Republic of China

ICS
17.180
Classification
A 60
Replacing
JJG 685-1990

Issued by: General Administration of Quality Supervision, Inspection and Quarantine

Contents

  • 1 Scope (1)
  • 2 Citations (1)
  • 3 Terms and definitions (1)
  • 4 Overview (1)
  • 5 Measurement characteristics (1)
  • 6 Calibration conditions (2)
  • 6.1 Environmental conditions (2)
  • 6.2 Calibration Equipment (2)
  • 7 Calibration items (2)
  • 7.1 Visual inspection (2)
  • 7.2 Relative spectral responsivity of photodetectors (2)
  • 8 Calibration method (2)
  • 8.1 Visual inspection (2)
  • 8.2 Photodetector relative spectral responsivity calibration (2)
  • 9 Calibration result expression (4)
  • 10 Re-study interval (4)
  • Appendix A Analysis of Uncertainty in Measurement of Relative Spectral Responsivity of Photodetectors (5)
  • Appendix B Another Method for Calibration of Relative Spectral Responsivity of Photodetectors (8)
  • Appendix C Photodetector Relative Spectral Responsive Calibration Record Format (10)
  • Appendix D Calibration Certificate (Report) Content (11)

1 Scope (1)

2 Citations (1)

3 Terms and definitions (1)

4 Overview (1)

5 Measurement characteristics (1)

6 Calibration conditions (2)

6.1 Environmental conditions (2)

6.2 Calibration Equipment (2)

7 Calibration items (2)

7.1 Visual inspection (2)

7.2 Relative spectral responsivity of photodetectors (2)

8 Calibration method (2)

8.1 Visual inspection (2)

8.2 Photodetector relative spectral responsivity calibration (2)

9 Calibration result expression (4)

10 Re-study interval (4)

Appendix A Analysis of Uncertainty in Measurement of Relative Spectral Responsivity of Photodetectors (5)

Appendix B Another Method for Calibration of Relative Spectral Responsivity of Photodetectors (8)

Appendix C Photodetector Relative Spectral Responsive Calibration Record Format (10)

Appendix D Calibration Certificate (Report) Content (11)

Photodetector relative spectral responsivity calibration specification

1 Scope

This specification applies to light used in photodetectors, illuminometer probes, luminance meter probes, and other optical test instruments.

Calibration of the relative spectral responsivity of the electrical detector. The wavelength range is (300~2500) nm, and other wavelength ranges can be used.

Calibrate according to this specification.

2 Citations

GB/T 6495.8-2002 "Photovoltaic devices - Part 8. Measurement of spectral response of photovoltaic devices" JJF 1001-1998 General Terms and Definitions Use of this specification should be done with the current valid version of the above cited documents.

3 Terms and definitions the spectral responsivity R(lambda) of a photodetector is defined as the output of the photodetector as a function of wavelength lambda dY(lambda) divided by the quotient of the photodetector monochromatic input dXe(lambda)=Xe(lambda)dlambda in the wavelength interval dlambda.

R(lambda)=

dY(lambda)

dXe(lambda)

The ratio of the spectral responsivity R(lambda) of the photodetector at wavelength lambda to the given reference value Sm.

Rr(lambda)=

R(lambda)

Rr(lambda) is the relative spectral responsivity of the detector. Sm can be the average, maximum or arbitrary of R(lambda) Value.

4 Overview

Photodetector is a sensor that uses a material with photoelectric effect to convert an optical radiation signal into an electrical signal.

Device. The photoelectric effect is divided into an internal photoelectric effect and an external photoelectric effect. Photodetectors fabricated by external photoelectric effect have photocells and photomultiplier tubes. Photodetectors fabricated using internal photoelectric effects are photoconductive and photovoltaic. Photoconductive detector Lead sulfide, mercury cadmium telluride, photoresistor, etc. Photovoltaic photodetectors are silicon photodiodes, silicon photocells, selenium optoelectronics Pool and so on. In general, whether it is a photodetector made by internal photoelectric effect or external photoelectric effect, their ringing it should only be in a certain spectral region and have strong spectral selectivity. That is, the spectral response of the photodetector Degrees are different at different wavelengths.

5 Measurement characteristics the relative spectral responsivity of the photodetector operating wavelength range and calibration uncertainty are.

Wavelength (300~400) nm. 2.6% (k=2.5);

Wavelength (400~1100) nm. 1.7% (k=2.5);

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 11 pages — is available in the English PDF.

How to Download JJF 1150-2006

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included

    A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.

Related Standards

English PDF
11 pages
Instant delivery (0–9 sec)
Invoice included
View Cart

Secure payment via Stripe

Payments accepted

VisaMastercardAmerican ExpressApple PayGoogle PayStripe

JJF 1150-2006

$424.00