JJF 1126-2004Calibration Specification for Ultrasonic Thickness Instruments (English PDF)
Also coversJJF1126-2004
This is a limited preview
Buy now to download the full PDF (14 pages)
Need quotation for your management?
Get a formal quotation for JJF 1126-2004 in 30 minutes.
Issued by
General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China Thi
Level / Type
National · Mandatory
Issue date
September 21, 2004
Implementation date
March 21, 2005
Scope
JJF 1126-2004 (Calibration Specification for Ultrasonic Thickness Instruments) is available as an English-translated PDF.
JJF 1126-2004 is the Chinese standard "Calibration Specification for Ultrasonic Thickness Instruments". Its scope clause reads: This Specification is applicable to the calibration of ultrasonic thickness instruments of which the resolution is 0.1mm and 0.01mm.
Its clauses include overview; metrological characteristics; calibration conditions; calibration items and calibration methods; calibration result expression; recalibration time interval.
It was issued by the General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China Thi on 2004-09-21 and took effect on 2005-03-21.
Found what you need?
Document preview — JJF 1126-2004
National Standard of the People's Republic of China
Issued by: General Administration of Quality Supervision, Inspection and Quarantine of the People's Republic of China Thi
Contents
- 1 Scope...5
- 2 Normative references...5
- 3 Overview...5
- 4 Metrological characteristics...7
- 5 Calibration conditions...8
- 6 Calibration items and calibration methods...9
- 7 Calibration result expression...11
- 8 Recalibration time interval...11
- Annex A Calibration certificate content...12
- Annex B Technical requirements for standard thickness block...13
- Annex C Technical requirements for standard round tubes...14
1 Scope
This Specification is applicable to the calibration of ultrasonic thickness instruments of which the resolution is 0.1mm and 0.01mm.
2 Normative references
This Specification referred to the following referenced documents.
JJF 1001-1998, General Terms in Metrology and Their Definitions
JJF 1059-1999, Evaluation and Expression of Uncertainty in Measurement
GB/T 12064.1-1999, Nondestructive testing terms - Ultrasonic testing
GB/T 11344-1989, Measuring thickness by ultrasonic pulse-echo contact method
When using this Specification, it shall use the currently valid versions of the above references.
3 Overview
The ultrasonic thickness instruments are mainly used to measure the wall thickness of hulls, oil and gas pipelines, high-pressure vessels, boilers as well as the thickness of large area sheet. The material to be tested can be metal materials such as steel, or non-metal materials such as plastic and nylon. The principle is that the instrument emits ultrasonic waves through the probe and reflects back to the test head after reaching the bottom surface of the test piece.
Use a counter to accurately measure the time that ultrasonic waves travel
4 Metrological characteristics
4.1 Repeatability
See Table 1 for instrument resolution specifications and corresponding repeatability requirements.
4.2 Indication error
See Table 1 for instrument resolution specifications and corresponding maximum allowable indication error.
4.3 Indication error of curved wall thickness measurement
See Table 2 for the instrument resolution specification and corresponding maximum allowable indication error.
4.6 Indication stability
See Table 3 for the instrument resolution specifications and corresponding indication stability requirements (the automatic shutdown type instrument does not conduct this calibration).
5 Calibration conditions
5.1 The calibration room temperature shall be 20°C±5°C. The relative humidity shall be less than 65%RH.
5.2 The standard thickness blocks and standard round tubes used for
calibration instruments shall meet the technical requirements in Annex B and
Annex C.
6 Calibration items and calibration methods
Before calibration, it is necessary to confirm that there are no defects in appearance and electromechanical functions that affect the calibration metrological characteristics.
6.1 Repeatability
After the instrument is turned on, operate the sound speed adjustment key or the sound speed dial to adjust the sound speed to 5900m/s. Return to the state of the instrument measuring thickness.
6.2 Indication error
Conduct calibration according to the procedures specified by the instrument.
Measure the standard thickness blocks in Annex B one by one in the measuring range of the instrument.
6.3 Indication error of curved wall thickness measurement
Conduct calibration according to the procedures specified by the instrument (do not recalibrate if the probe is not replaced).
6.4 Fine adjustment range for thickness calibration
After the instrument calibration, keep the coupling state of the probe and thickness block unchanged. Continuously adjust the " " and " " keys of the instrument calibration (or rotate the fine adjustment knob for calibration thickness). The indication range of the instrument shall not be less than ± 1mm.
7 Calibration result expression
After the instrument is calibrated, issue a calibration certificate. The calibration certificate shall include at least the information contained in Annex A.
8 Recalibration time interval
The recommended recalibration time interval is 1 year.
......
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 14 pages — is available in the English PDF.
Referenced standards
Normative references
- JJF 1001-1998General Terms in Metrology and Their Definitions
- GB/T 11344-1989Non-destructive testing - Ultrasonic thickness measurement
JJF 1059-1999 · GB/T 12064.1-1999
How to Download JJF 1126-2004
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards