HJ/T 97-2003The technical requirement for water quality automatic analyzer of electroconductivity (English PDF)

Preview PDF

Also coversHJT97-2003

This is a limited preview

Buy now to download the full PDF (12 pages)

Need quotation for your management?

Get a formal quotation for HJ/T 97-2003 in 30 minutes.

Issued by

State Environmental Protection Administration

Level / Type

National · Recommended

Issue date

March 28, 2003

Implementation date

July 1, 2003

Scope

HJ/T 97-2003 (The technical requirement for water quality automatic analyzer of electroconductivity) is available as an English-translated PDF.

HJ/T 97-2003 — This technical specification specifies the technical performance requirements and performance test methods of water quality automatic analyzer of electroconductivity in respect of surface water, industrial wastewater and municipal wastewater.

It is applicable to the design, manufacturing, and performance test of this kind of instruments.

Found what you need?

Document preview — HJ/T 97-2003

National Standard of the People's Republic of China

Issued by: State Environmental Protection Administration

Contents

  • Foreword...4
  • 1 Scope...5
  • 2 Terms and Definitions...5
  • 3 Determination Principle and Measuring Range...6
  • 4 Working Voltage and Frequency...6
  • 5 Performance Requirements...6
  • 6 Instrument Structure...6
  • 7 Test Method...8
  • 8 Mark...11
  • 9 Operating Instructions...12
  • 10 Calibration...12

Foreword

In order to carry out the "Environmental Protection Law of the People's Republic of China"

and the "Water Pollution Prevention and Control Law of the People's Republic of China",

improve the ability of water environment monitoring work, realize automation and

modernization of water quality monitoring, achieve early-warning monitoring of surface

water quality, and monitoring and controlling the total quantity of pollutant sources, this

Standard is formulated.

This Standard specifies main technical requirements in areas such as development and

production, performance test, type section and use, routine check of water quality

automatic analyzer of electroconductivity.

This Standard was proposed by Department of Science, Technology and Standards of

Administration of Environmental Protection of the People's Republic of China.

This Standard is responsibly drafted by China National Environmental Monitoring Center.

This Standard shall be responsibly interpreted by the Administration of Environmental

Protection of the People's Republic of China.

Technical Requirement for Water Quality

Automatic Analyzer of Electroconductivity

1 Scope

This technical specification specifies the technical performance requirements and

performance test methods of water quality automatic analyzer of electroconductivity in

respect of surface water, industrial wastewater and municipal wastewater. It is applicable

to the design, manufacturing, and performance test of this kind of instruments.

2 Terms and Definitions

For the purpose of this document, the following terms and definitions shall apply.

2.1 Sample

The surface water, industrial wastewater and municipal wastewater that is led into

automatic analyzer.

2.2 Zero drift

The percentage of the change of an automatic analyzer's indication value over a given

period TO its measurement-range, when the zero correction solution specified in the

technical requirement is employed to give continuous test to sample.

2.3 Span drift

The percentage of the change of an automatic analyzer's indication value over a given

period TO its measurement-range, when the measurement-range correction solution

specified in the technical requirement is employed to give continuous test to sample.

2.4 Mean Time between Failures

The ratio of total hours (h) to failures (number) of an automatic analyzer during its

inspection period, expressed as "MTBF", in h/times.

2.5 Response Time (T90)

The time required for the indication value of an electrode system to reach 90% of the

standard electroconductivity value of measurement-range correction solution, when the

electrode system is moved from zero correction solution to measurement-range correction

solution.

7.1.6 Electroconductivity standard solution temperature

(25.0 ± 0.5)°C.

7.2 Reagent

7.2.1 Pure Water

Let distilled water pass through an ion exchange column. The electroconductivity shall be

less than 0.1mS/m.

7.2.2 Zero-point Correction Solution

Adopt the pure water as specified in 8.2.1.

7.2.3 Measurement-Range Correction Solution

Adopt the solution of 80% measurement-range value.

7.2.4 0.0100mol/L KCl Standard Solution

Weigh 0.7456g of potassium chloride (KCl) that has been dried at 105°C for 2h and

cooled down. Dissolve it in pure water and dilute to 1000ml at 25°C. The

electroconductivity of this solution (0.0100mol/L KCl) is 141.3mS/m at 25°C.

7.3 Test Preparation and Correction

7.3.1 Test Preparation

Connect the electrode to the signal converter. Switch on the power. Preheat the automatic

analyzer for more than 30min before starting the test, so as to make the function of all

parts and the displaying records unit to be stable. In case the electrode is polluted, it shall

be washed with detergents or 0.01 mol/L hydrochloric acid or the like. Then clean up with

running water, and dry-clean it.

7.3.2 Correction
7.3.2.1 Dip the electrode into the zero correction solution. Adjust the indication value to be

zero.

7.3.2.2 Dip the electrode into the measurement-range correction solution. Adjust the

indication value to be the electroconductivity value of 0.0300mol/L KCl standard solution.

7.3.2.3 Adjustment

Alternate the operation of 7.3.2.1 and 7.3.2.2.Adjust the analyzer until the difference

between the measured value and the standard value of the standard solution falls within

±1%.

8.5 Power supply category and capacity.

8.6 Manufacturer name.

8.7 Manufacturing date and batch number.

8.8 Signal output class (if necessary).

9 Operating Instructions

The operating instructions must at least provide the relevant items as follows.

9.1 Selection of installation site.

9.2 Sample flow rate.

9.3 Piping and wiring.

9.4 Preheating time.

9.5 Use methods.

9.5.1 Measurement preparation and correction method.
9.5.2 Preparation method of correction solution.
9.5.3 Operating method of determination.
9.5.4 Handling when measurement is stopped.

9.6 Maintenance inspection.

9.6.1 Daily routine inspection methods.
9.6.2 Regular inspection methods.
9.6.3 Electrode system cleaning.
9.6.4 Countermeasures for failures.

9.7 Other considerations for use.

10 Calibration

10.1 Routine calibration

For reproducibility, drift and response time, the calibrating period is at least once a month,

either by automatic or manual calibration.

10.2 Calibration under Supervision

The installed continuous monitoring system shall be calibrated regularly. And the regular

calibration results shall be reported to relevant administrative department of

environmental protection. Regular calibration shall be undertaken by qualified monitoring

agencies.

According to environmental monitoring technical specification, regular calibration mainly

includes on-site contrast test, review and examination of operation data and daily

operation records etc.

......

......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 12 pages — is available in the English PDF.

How to Download HJ/T 97-2003

  1. 1

    Add to cart

    Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.

  2. 2

    Checkout

    Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).

  3. 3

    Instant delivery (0–9 sec)

    Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.

  4. 4

    Invoice included