HJ 870-2017Stationary Source Emission - Determination of Carbon Dioxide - Non-Dispersive Infrared Absorption Method (English PDF)
Also coversHJ870-2017
This is a limited preview
Buy now to download the full PDF (11 pages)
Need quotation for your management?
Get a formal quotation for HJ 870-2017 in 30 minutes.
Issued by
Ministry of Environmental Protection of PRC
Level / Type
National · Mandatory
Issue date
November 28, 2017
Implementation date
January 1, 2018
Scope
HJ 870-2017 (Stationary Source Emission - Determination of Carbon Dioxide - Non-Dispersive Infrared Absorption Method) is available as an English-translated PDF.
HJ 870-2017 is the Chinese standard "Stationary Source Emission - Determination of Carbon Dioxide - Non-Dispersive Infrared Absorption Method". Its scope clause reads: This Standard specifies the non-dispersive infrared absorption method for determination of carbon dioxide in the stationary source emission.
This Standard is applicable to the determination of carbon dioxide in the stationary source emission. Under the condition that the instrument range is 20% (volume concentration), the detection limit of this Standard is 0.03% (0.6g/m3); while the lower limit of determination is 0.12% (2.4g/m3).
Its clauses include terms and definitions; reagents and materials; apparatus; results calculation and expression; quality assurance and quality control. It was issued by the Ministry of Environmental Protection of PRC on 2017-11-28 and took effect on 2018-01-01.
Found what you need?
Document preview — HJ 870-2017
National Standard of the People's Republic of China
Issued by: Ministry of Environmental Protection of PRC
Contents
- Foreword...3
- 1 Scope...4
- 2 Normative References...4
- 3 Terms and Definitions...4
- 4 Method Principle...5
- 5 Interference and Elimination...5
- 6 Reagents and Materials...6
- 7 Apparatus...6
- 8 Sampling and Determination...7
- 9 Results Calculation and Expression...8
- 10 Precision and Accuracy...9
- 11 Quality Assurance and Quality Control...10
- 12 Precautions...11
1 Scope
This Standard specifies the non-dispersive infrared absorption method for determination of carbon dioxide in the stationary source emission.
This Standard is applicable to the determination of carbon dioxide in the stationary source emission.
Under the condition that the instrument range is 20% (volume concentration), the detection limit of this Standard is 0.03% (0.6g/m3); while the lower limit of determination is 0.12% (2.4g/m3).
2 Normative References
The contents of this Standard refer to the following documents or their terms. For the undated references, only the valid version applies to this Standard.
GB/T 16157 The Determination of Particulates and Sampling Methods of Gaseous
Pollutants Emitted from Exhaust Gas of Stationary Source
HJ/T 373 Technical Specifications of Quality Assurance and Quality Control for
Monitoring of Stationary Pollution Source
HJ/T 397 Technical Specifications for Emission Monitoring of Stationary Source
3 Terms and Definitions
3.1 Zero gas
The gas with no measurement components or less than the specified value; the concentration of other component does not interfere with the measurement component results or the interference of the measurement component may be negligible.
3.2 Calibration span
6 Reagents and Materials
6.1 Standard gas of carbon dioxide
The commercial certificated standard gas with uncertainty <=2%.
6.2 Zero gas
Nitrogen gas with purity >=99.99% or clean air not interfering the determination.
7 Apparatus
7.1 Carbon dioxide detector by non-dispersive infrared absorption method
7.1.1 Composition
Instrument consists of: analyzer (including gas flow meter and flow control unit, air pump, detector, etc.), sampling tube (including dust filter device and heating and heat preservation device), air guiding pipe, dehumidifying device, portable printer, etc.
7.1.2 Performance requirements
a) Calibration error: not exceeding ±5%;
b) System bias: not exceeding ±5%;
c) Zero drift: not exceeding ±3%;
d) Span drift: not exceeding ±3%;
e) Has the function to eliminate the interference;
f) The dehumidifying device shall meet the requirements of GB/T 16157;
g) Has the function to show the sampling flow rate; the measurement range and accuracy of gas flow meter shall meet the requirements of the instrument;
h) The heating and heat preserving temperature of the sampling tube: it may be set or adjusted within 120°C~160°C.
7.2 Steel cylinder of standard gas
It is equipped with adjustable pressure reducing valve, adjustable rotor flowmeter and air guiding tube.
7.3 Gas collecting bag
8.5 Sample determination
8.5.1 Place the front end of the sampling tube of the instrument in the exhaust pipe;
block the sampling hole, so that it does not leak gas.
8.5.2 Start the air pump; conduct sampling determination according to the sampling
flow specified by the detector. After the detector is stable, save the measurement data in minutes; take the average value of the measure data for at least continuous 5 minutes as the one measurement value.
8.5.3 After the end of one measurement, use the zero gas to clean the instrument according to the instruction manual of the instrument.
8.5.4 After obtaining the measurement result, use zero gas to clean the instrument.
When the indication value returns to near zero point, turn off the power and finish the measurement.
9 Results Calculation and Expression
9.1 Results calculation
9.1.1 Calculation of exhaust flow rate
According to the provisions of GB/T 16157, calculate the dry exhaust flow rate Qsn (m3/h) under the standard state (273K, 101.325kPa).
9.1.2 Calculation of carbon dioxide concentration
The concentration (g/m3) of carbon dioxide in the stationary source emission shall be calculated as per the following formula:
Where:
rho – mass concentration of carbon dioxide in the dry exhaust under the standard state, in g/m3;
omega – volume concentration of carbon dioxide in the measured exhaust determined by the instrument, in %.
9.1.3 Calculation of carbon dioxide emission rate
The relative standard deviations in the laboratory are: 0.05% ~ 0.12%, 0.01% ~
0.20%, 0.04% ~ 0.13%;
The relative standard deviations among laboratories are: 3.4%, 2.1%, 3.2%;
The repeatability limits are: 0.04%, 0.06%, 0.02%;
The reproducibility limits are: 1.3%, 1.0%, 0.52%.
10.2 Accuracy
Determination of carbon dioxide standard gases at concentration levels of 5.01%,
15.1%, 20.0% in 6 laboratories:
Relative errors are: -1.9% ~ 2.0%, -1.4% ~ 1.7%, -0.08% ~ 0.06%;
Final values of relative errors are: -0.17%±2.6%, -0.10%±2.1%, 0.02%±0.10%.
11 Quality Assurance and Quality Control
11.1 Before monitoring, measure the zero gas and carbon dioxide standard gas;
calculate the calibration error and system bias. If the calibration error and/or system bias does not meet the requirements of Items a) and b) in 7.1.2; then find out the causes; and conduct the corresponding repair and maintenance till the requirements are met.
11.2 After monitoring, measure the zero gas and carbon dioxide standard gas again;
calculate the calibration error and system bias. If the calibration error and system bias meet the requirements of Items a) and b) in 7.1.2; the determination results of the sample are judged to be valid; otherwise, the determination results of the sample are judged to be invalid.
NOTE: The inspection of the calibration error of the whole system including sampling tube, air guiding pipe, dehumidifying device, etc. can be taken to replace the inspection of the calibration error and system bias of the analyzer [its evaluation shall implement the requirements of Item a) in 7.1.2].
11.3 The determination results of the sample shall be within 20%~100% of the calibration span; otherwise, the calibration span shall be re-selected.
11.4 At least once inspection for zero drift and span drift shall be carried out every half a year; the results shall meet the requirements of Items c) and d) in 7.1.2; otherwise, timely maintain and repair the instrument.
11.5 Perform linear calibration against the instrument by standard gas with low (< 20%
C.S.), medium (40%~60% C.S.), high (80%~100% C.S.) concentration at least once
......
......
This preview omits tables, figures, formulas and parts of the technical clauses. The complete document — 11 pages — is available in the English PDF.
Referenced standards
Normative references
- GB/T 16157Determination of particulates and sampling methods of gaseous pollutants emitted from exhaust gas of stationary source
- HJ/T 373Technical specifications of quality assurance and quality control for monitoring of stationary pollution source(on trial)
- HJ/T 397Technical specifications for emission monitoring of stationary source
How to Download HJ 870-2017
- 1
Add to cart
Click "Download PDF" on this page and choose "Buy this standard". You can add more standards before checkout.
- 2
Checkout
Enter your email and billing details. Payment is processed securely by Stripe (cards or bank transfer). Prefer bank details? Request a quotation (min. order $300).
- 3
Instant delivery (0–9 sec)
Delivery is automatic: within seconds of payment you'll receive an email with a secure download link. The link stays valid for 72 hours.
- 4
Invoice included
A tax invoice is attached to the confirmation email. Need a custom invoice? Contact us.
Related Standards
GB/T 16157-1996 — Determination of particulates and sampling methods of gaseous pollutants emitted from exhaust gas of stationary source
HJ/T 373-2007 — Technical specifications of quality assurance and quality control for monitoring of stationary pollution source(on trial)
HJ/T 397-2007 — Technical specifications for emission monitoring of stationary source